Micro Therapeutics, Inc. DBA EV3

FDA adverse event reports naming Micro Therapeutics, Inc. DBA EV3 as the device manufacturer, with recall history from the live FDA recall database. Reporter-supplied name as filed: MICRO THERAPEUTICS, INC. DBA EV3.

Total reports
15,559
Deaths
1,172
Injuries
4,209
Malfunctions
10,178

Reports per year

12
16
17
18
19
20
21
22
23
24
25
26

Source: FDA adverse event reports by year received, 2012–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.

Top device categories

Recent recalls

  • 2020-03-09TerminatedProcess change control

    ev3 Pipeline Flex Embolization Device. For neurological endovascular use.

    Embolization device delivery system may fracture at the distal section during device implantation. A weakened bond may result in unintended separation, where the distal portion of the device delivery system remains in the patient. If a Pipeline Flex embolization device has already been implanted successfully, there is no increased risk to patients due to this issue.

  • 2020-03-09TerminatedProcess change control

    ev3 Pipeline Flex Embolization Device with Shield Technology. Sold OUS. For neurological endovascular use.

    Embolization device delivery system may fracture at the distal section during device implantation. A weakened bond may result in unintended separation, where the distal portion of the device delivery system remains in the patient. If a Pipeline Flex embolization device has already been implanted successfully, there is no increased risk to patients due to this issue.

  • 2018-05-16TerminatedProcess design

    MindFrame Capture LP Revascularization Devices: Product Numbers: (a) REF 300010 (b) REF 300011 (c) REF 300012 (d) REF 300013 (e) REF 300014 (f) REF 300015 (g) REF 300016 (h) REF 300017 (i) REF 300018 Intended to restore blood flow by removing thrombus from a large intracranial vessel experiencing ischemic stroke within 8 hours of symptom onset.

    There is a potential issue for all MindFrame Capture devices to partially detach or separate from the delivery wire.

  • 2016-11-09TerminatedDevice Design

    Pipeline Embolization Device (Pipeline Classic) The device is indicated for the endovascular treatment of adults (22 year of age or older) with large or giant widenecked intracranial aneurysms (lAs) in the internal carotid artery from the petrous to the superior hypophyseal segements. Catalog No. FA-71300-25V09 FA-71350-25V16 FA-71400-25V10 FA-71425-35V13 FA-71475-30V16 FA-71300-25V12 FA-71350-25V17 FA-71400-25V11 FA-71425-35V15 FA-71475-30V18 FA-71300-25V13 FA-71350-25V18 FA-71400-25V12 FA-71425-35V18 FA-71475-35V09 FA-71300-25V15 FA-71350-25V19 FA-71400-25V13 FA-71450-25V09 FA-71475-35V10 FA-71300-25V16 FA-71350-30V09 FA-71400-25V15 FA-71450-25V10 FA-71475-35V11 FA-71300-25V17 FA-71350-30V10 FA-71400-25V16 FA-71450-25V11 FA-71475-35V12 FA-71300-25V18 FA-71350-30V12 FA-71400-25V18 FA-71450-25V12 FA-71475-35V13 FA-71300-25V19 FA-71350-30V13 FA-71400-25V19 FA-71450-25V13 FA-71475-35V15 FA-71300-30V09 FA-71350-30V15 FA-71400-30V09 FA-71450-25V15 FA-71475-35V18 FA-71300-30V12 FA-71350-30V16 FA-71400-30V10 FA-71450-25V16 FA-71475-35V19 FA-71300-30V13 FA-71350-30V18 FA-71400-30V11 FA-71450-25V18 FA-71500-25V09 FA-71300-30V15 FA-71350-35V09 FA-71400-30V12 FA-71450-25V19 FA-71500-25V10 FA-71300-30V16 FA-71350-35V10 FA-71400-30V13 FA-71450-30V09 FA-71500-25V11 FA-71300-30V19 FA-71350-35V12 FA-71400-30V15 FA-71450-30V10 FA-71500-25V12 FA-71300-35V09 FA-71350-35V13 FA-71400-30V16 FA-71450-30V11 FA-71500-25V13 FA-71300-35V12 FA-71350-35V15 FA-71400-30V18 FA-71450-30V12 FA-71500-25V15 FA-71300-35V13 FA-71350-35V18 FA-71400-35V09 FA-71450-30V13 FA-71500-25V16 FA-71300-35V15 FA-71375-25V09 FA-71400-35V13 FA-71450-30V15 FA-71500-25V18 FA-71325-25V09 FA-71375-25V10 FA-71400-35V15 FA-71450-30V16 FA-71500-25V19 FA-71325-25V10 FA-71375-25V11 FA-71400-35V18 FA-71450-30V18 FA-71500-30V09 FA-71325-25V12 FA-71375-25V12 FA-71425-25V09 FA-71450-30V19 FA-71500-30V10 FA-71325-25V13 FA-71375-25V13 FA-71425-25V10 FA-71450-35V09 FA-71500-30V11 FA-71325-25V15 FA-71375-25V15 FA-71425-25V11 FA-71450-35V10 FA-71500-30V12 FA-71325-25V18 FA-71375-25V16 FA-71425-25V12 FA-71450-35V12 FA-71500-30V13 FA-71325-25V19 FA-71375-25V17 FA-71425-25V13 FA-71450-35V13 FA-71500-30V15 FA-71325-30V09 FA-71375-25V18 FA-71425-25V15 FA-71450-35V15 FA-71500-30V16 FA-71325-30V10 FA-71375-25V19 FA-71425-25V16 FA-71450-35V16 FA-71500-30V17 FA-71325-30V12 FA-71375-30V09 FA-71425-25V18 FA-71475-25V09 FA-71500-30V18 FA-71325-30V13 FA-71375-30V10 FA-71425-25V19 FA-71475-25V10 FA-71500-35V09 FA-71325-30V15 FA-71375-30V12 FA-71425-30V09 FA-71475-25V11 FA-71500-35V10 FA-71325-30V16 FA-71375-30V13 FA-71425-30V10 FA-71475-25V12 FA-71500-35V11 FA-71325-30V18 FA-71375-30V15 FA-71425-30V11 FA-71475-25V13 FA-71500-35V12 FA-71325-30V19 FA-71375-30V16 FA-71425-30V12 FA-71475-25V15 FA-71500-35V13 FA-71325-35V09 FA-71375-30V18 FA-71425-30V13 FA-71475-25V16 FA-71500-35V15 FA-71325-35V12 FA-71325-35V13 FA-71325-35V15 FA-71350-25V09 FA-71350-25V10 FA-71350-25V13 FA-71350-25V15 FA-77250-10V13 FA-77250-10V19 FA-77250-12V11 FA-77250-12V12 FA-77250-12V13 FA-77250-12V18 FA-77250-14V12 FA-77250-14V13 FA-77250-14V15 FA-77250-14V18 FA-77250-14V19 FA-77250-16V11 FA-77250-16V12 FA-77250-16V13 FA-77250-16V15 FA-77250-16V16 FA-77250-16V19 FA-77250-18V09 FA-77250-18V10 FA-77250-18V11 FA-77250-18V12 FA-77250-18V13 FA-77250-18V16 FA-77250-18V18 FA-77250-18V19 FA-77250-20V10 FA-77250-20V11 FA-71375-35V09 FA-71375-35V12 FA-71375-35V13 FA-71375-35V15 FA-71375-35V16 FA-71375-35V18 FA-71400-25V09 FA-77275-14V11 FA-77275-14V12 FA-77275-14V13 FA-77275-14V17 FA-77275-14V18 FA-77275-14V19 FA-77275-16V12 FA-77275-16V13 FA-77275-16V16 FA-77275-16V18 FA-77275-16V19 FA-772

    Delamination and detachment of the PTFE (polytetrafluoroethylene) coating material may lead to PTFE coating in the blood stream.

  • 2016-11-09TerminatedDevice Design

    Alligator Retrieval Device The device is intended for use in the peripheral and neurovasculature for foreign body removal. Catalog No. FA-88810-20V04 FA-88810-20V06 FA-88810-30V04 FA-88810-30V06 FA-88810-40V04 FA-88810-40V06 FA-88810-50V04 FA-88810-50V06

    Delamination and detachment of the PTFE (polytetrafluoroethylene) coating material may lead to PTFE coating in the blood stream.

  • 2016-11-09TerminatedDevice Design

    Marathon Flow Directed Micro Catheter, UltraFlow HPC Flow Directed Micro Catheter Marathon Flow: The device is intended to access periphera and neurovasculature for the controlled selective infusion of physician-specified therapeutics agents such as embolization materials and of diagnostic materials such as contrast media. UltraFlow HPC Flow: The device is intended for the controlled selective infusion of physician-specified pharmacologic agents or contrast media into the distal vasculature of the peripheral and neuroanatomy. It is not intended for use in the coronary vasculature. Catalog No. 105-5055V02 105-5055V02 105-5055V05 105-5065V02 105-5065V05 105-5066V02 105-5066V05

    Delamination and detachment of the PTFE (polytetrafluoroethylene) coating material may lead to PTFE coating in the blood stream.

  • 2016-11-09TerminatedDevice Design

    X-Celerator Hydrophilic Exchange Guidewire Catalog No. 103-0601-300 103-0601-350V10 103-0602-350V06 103-0601-300V01 103-0602-300 103-0602-350V09 103-0601-300V02 103-0602-300V01 103-0602-350V10 103-0601-300V03 103-0602-300V02 103-0601-300V06 103-0602-300v06 103-0601-300V10 103-0602-300V09 103-0601-300V11 103-0602-300V10 103-0601-350 103-0602-350 103-0601-350V01 103-0602-350V01 103-0601-350V06 103-0602-350V02

    Delamination and detachment of the PTFE (polytetrafluoroethylene) coating material may lead to PTFE coating in the blood stream.

  • 2015-10-14TerminatedProcess control

    -Pipeline Embolization Device -Pipeline Flex Embolization Device The Pipeline and Pipeline Flex Embolization Devices consist of a permanent implant combined with a guidewire based delivery system. The Pipeline and Pipeline Flex embolization device implant is a braided, multi-alloy, mesh cylinder woven from platinum/tungsten and cobalt-chromium-nickel alloy wires. The Pipeline and Pipeline Flex Embolization Devices are indicated for the endovascular treatment of adults (22 years of age or older) with large or giant wide-necked intracranial aneurysms (IAs) in the internal carotid artery from the petrous to the superior hypophyseal segments.

    The firm is recalling Pipeline and Pipeline Flex Embolization Devices from U.S. since the devices were shipped with an EU version of the Instructions for Use.

  • 2014-04-16TerminatedProcess change control

    Pipeline Embolization Device (PED). Used endovascular treatment of adults with intracranial aneurysms in the internal carotid artery.

    Potential for the PTFE (polytetrafluoroethylene) coating to delaminate and detach from the delivery wire.

  • 2014-04-16TerminatedProcess change control

    Alligator Retrieval Device (ARD). Used in the peripheral and neurovasculature for foreign body removal.

    Potential for the PTFE (polytetrafluoroethylene) coating to delaminate and detach from the delivery wire.

Monitor Micro Therapeutics, Inc. DBA EV3 automatically