Skip to main content

GE Medical Systems (china) Co., Ltd

FDA adverse event reports naming GE Medical Systems (china) Co., Ltd as the device manufacturer, with recall history from the live FDA recall database. Reporter-supplied name as filed: GE MEDICAL SYSTEMS (CHINA) CO., LTD.

Total reports
2,015
Deaths
0
Injuries
3
Malfunctions
2,012

Reports per year

12
15
25
26

Source: FDA adverse event reports by year received, 20122026. Reporting paused during the 2026 MAUDE-to-AEMS transition and has since resumed.

Top device categories

Recent recalls

  • 2026-06-03Open, ClassifiedSoftware Manufacturing/Software Deployment

    MAC 5 A4 Resting ECG Analysis System 855001-001; MAC 5 A5 Resting ECG Analysis System 8855002-001. Electrocardiograph

    Software in development inadvertently made available for customer use. Use of this software version can result in clinical information being associated with the incorrect patient. This could impact clinical decision-making.

  • 2025-12-18Open, ClassifiedComponent change control

    GE Healthcare Carestation anesthesia system, product number and REF numbers: Carestation 610 A1, REF 1012-9620-222; Carestation 620 A1 REF 1012-9620-000; Carestation 620 A1, REF 1012-9620-200; Carestation 620 A1, REF 1012-9620-202; Carestation 620 SE A1, REF 1012-9620-212; Carestation 630 A1, REF 1012-9650-222; Carestation 650 A1, REF 1012-9650-000; Carestation 650 A1, REF 1012-9650-200; Carestation 650 A1, REF 1012-9650-202; Carestation 650 SE A1, REF 1012-9650-212; Carestation 650c A1, REF 1012-9655-202; Anesthesia Gas Machine Products for which affected PMB can be used as a spare part: Carestation 620 A2, REF 1012-9620-002 00840682124546 Carestation 620 SE A2, REF 1012-9620-012 00195278569677 Carestation 650 A2, REF 1012-9650-002 00840682124560 Carestation 650 SE A2, REF 1012-9650-012 00195278569684 Carestation 650c A1, REF 1012-9655-000 00840682103954 Carestation 650c A2, REF 1012-9655-002 00840682124539 Carestation 650c A1, REF 1012-9655-200 00195278439543

    Potential unexpected shutdown of Carestation 600 and 700 Series Anesthesia Systems containing certain power management boards, if the AC mains power is unplugged or in the event of an AC mains power failure. Anesthesia systems only operate on battery power in a rare event that AC mains power is lost and there is no continuous backup emergency power. If AC power is interrupted, the system will not automatically switch over to the battery supply mode and will reboot. If this issue occurs, a temporary disruption of mechanical ventilation, manual ventilation, and volatile agent delivery may occur. Following the reboot, the system will not return to the previous ventilation settings.

  • 2025-12-18Open, ClassifiedComponent change control

    GE Healthcare Carestation anesthesia system, product number and REF numbers: Carestation 750 A1, REF 1012-9750-000; Carestation 750 A2, REF 1012-9750-002; Carestation 750c A1, REF 1012-9755-000; Products for which affected PMB can be used as a spare part: Carestation 750c A2, REF 1012-9755-002

    Potential unexpected shutdown of Carestation 600 and 700 Series Anesthesia Systems containing certain power management boards, if the AC mains power is unplugged or in the event of an AC mains power failure. Anesthesia systems only operate on battery power in a rare event that AC mains power is lost and there is no continuous backup emergency power. If AC power is interrupted, the system will not automatically switch over to the battery supply mode and will reboot. If this issue occurs, a temporary disruption of mechanical ventilation, manual ventilation, and volatile agent delivery may occur. Following the reboot, the system will not return to the previous ventilation settings.

  • 2025-12-18Open, ClassifiedComponent change control

    GE Healthcare Carestation anesthesia system Power Management Board (PMB) Field Replaceable Unit (FRU), 2076139-001-S

    Potential unexpected shutdown of Carestation 600 and 700 Series Anesthesia Systems containing certain power management boards, if the AC mains power is unplugged or in the event of an AC mains power failure. Anesthesia systems only operate on battery power in a rare event that AC mains power is lost and there is no continuous backup emergency power. If AC power is interrupted, the system will not automatically switch over to the battery supply mode and will reboot. If this issue occurs, a temporary disruption of mechanical ventilation, manual ventilation, and volatile agent delivery may occur. Following the reboot, the system will not return to the previous ventilation settings.

  • 2025-06-26Open, ClassifiedSoftware design (manufacturing process)

    GE Healthcare LOGIQ F R3 Ultrasound System, Model/Catalog Number 5943263, diagnostic ultrasound system

    GE HealthCare has become aware that the Estimated Fetal Weight (EFW) measurement data feature on the Versana Premier R3 and LOGIQ F R3 series ultrasound systems can display previous obstetric patient data in the Whizz report. This could potentially lead to an incorrect clinical decision due to inaccuracy in the fetal size and growth calculation.

  • 2025-06-26Open, ClassifiedSoftware design (manufacturing process)

    GE Healthcare Versana Ultrasound Systems, sold under the following names and Model/Catalog Numbers: 1) Versana Premier R3 VS, Model/Catalog Number 5938558; 2) Versana Premier R3 VA, Model/Catalog Number 5938559; 3) Versana Premier Lotus 4PP, Model/Catalog Number 5938560; 4) Versana Premier Lotus 5PP, Model/Catalog Number 5938561; 5) Versana Premier R3 VA Vet, Model/Catalog Number 5946387; 6) Versana Premier R3 VS Vet, Model/Catalog Number 5946481. 7) Versana Premier R3 Expert, Model/Catalog Number 5946950; 8) Versana Premier R3, Model/Catalog Number 5946951; 9) Versana Premier R3 Pro, Model/Catalog Number 5946952; 10) Versana Premier R3 Elite, Model/Catalog Number 5946953; 11) Versana Premier R3 Plus, Model/Catalog Number 5946954; 12) Versana Premier R3 Ultra, Model/Catalog Number 5946955; 13) Versana Premier R3 Max, Model/Catalog Number 5946956; 14) Versana Premier R3 Super, Model/Catalog Number 5946957; 15) Versana Premier R3 VS India, Model/Catalog Number 5948398; 16) Versana Premier R3 VA India, Model/Catalog Number 5948399; diagnostic ultrasound system

    GE HealthCare has become aware that the Estimated Fetal Weight (EFW) measurement data feature on the Versana Premier R3 and LOGIQ F R3 series ultrasound systems can display previous obstetric patient data in the Whizz report. This could potentially lead to an incorrect clinical decision due to inaccuracy in the fetal size and growth calculation.

  • 2025-04-29Open, ClassifiedDevice Design

    GE Healthcare CARESTATION 620 A1, Model/REF Number 1012-9620-000. anesthesia systems

    Certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.

  • 2025-04-29Open, ClassifiedDevice Design

    GE Healthcare CARESTATION 620 A2, Model/REF Number 1012-9620-002. Anesthesia system.

    Certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.

  • 2025-04-29Open, ClassifiedDevice Design

    GE Healthcare Carestation 620 SE, Model/REF Number 1012-9620-012. Anesthesia system.

    Certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.

  • 2025-04-29Open, ClassifiedDevice Design

    GE Healthcare Carestation 620 A1, Model/REF Number 1012-9620-200. Anesthesia system.

    Certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.

Monitor GE Medical Systems (china) Co., Ltd automatically

DeviceWatch watches FDA adverse events and recalls for the products you care about, summarizes clinical narratives with AI and alerts you when a safety signal emerges.