Skip to main content

Ferno-washington, Inc.

FDA adverse event reports naming Ferno-washington, Inc. as the device manufacturer, with recall history from the live FDA recall database. Reporter-supplied name as filed: FERNO-WASHINGTON, INC..

Total reports
1,329
Deaths
12
Injuries
276
Malfunctions
959

Reports per year

12
13
14
15
16
17
18
19
20
21
22
23
24
25
26

Source: FDA adverse event reports by year received, 20122026. Reporting paused during the 2026 MAUDE-to-AEMS transition and has since resumed.

Top device categories

Recent recalls

  • 2025-10-10Open, ClassifiedDevice Design

    FERNO POWER X2 POWERED AMBULANCE COT, Model/Catalog Number: 0015816; FOR USE IN GROUND TRANSPORTATION AMBULANCES.

    The fastening post hardware could loosen allowing the fastening post to detach from the cot

  • 2022-11-04Open, ClassifiedComponent design/selection

    FERNO POWER X1 AMBULANCE COT- Item #0015807/UDI-DI 00190790002183 with 36V Li-Ion Batteries Model Number #0822484/0822483 Note: Transcend Stair Chair Powertraxx (Item #0731403/UDI-DI 00190790004064) not included in this action).

    Batteries could become unstable when not properly maintained and/or subjected to repeated drops, power washing directly into battery contacts or abuse, may result in a fire and/or injury

  • 2017-04-15TerminatedMixed-up of materials/components

    Ferno PROFlexx Model 28Z Chair Cot

    The recall was initiated as a result of the detection of an incorrect nut used in manufacturing.

  • 2015-10-01TerminatedDevice Design

    Ferno iNX Wheeled Stretcher, Integrated Patient Transport & Loading System Product Usage: Stretcher for pre-hospital patient transport.

    Hole elongation in the telescoping load frame channel. A hole elongation could affect the ability of the retention pin to stay seated which is necessary to preclude movement of the telescoping load frame.

  • 2015-08-25TerminatedProcess control

    Wheel caster set screw for Ferno iNX Wheeled Stretcher. Ferno iNX Wheeled Stretcher for pre-hospital patient transport.

    The firm discovered an improperly installed set screw in one caster arm assembly on a unit in production. The screw assists in preventing the bearing cup on the caster assembly from backing out of the caster arm during use.

  • 2015-01-16TerminatedMixed-up of materials/components

    Ferno Model 35X PROFlexx Stretchers, one unit per package

    The wheel castor assemblies may be loosening on the stretchers.

  • 2008-09-30TerminatedDevice Design

    Ferno Model 35X PROFlexX Ambulance Stretcher, Model #'s: 35X; 35XST; 35XIT. The Ferno Model 35-X PROFlexx ambulance stretcher is used to transport a patient during rescue or emergency situations as well as a means to transport patients for general medical purposes.

    The Ferno Model 35-X PROFlexx legs could experience metal fatigue and possibly fracture after one to two years of use based on the usage and method of operation.

  • 2008-09-11TerminatedMixed-up of materials/components

    Model 35A Series Mobile Transporters is an ambulance stretcher for patient transport.

    A component defect was found in corner castings of the firm's wheeled ambulance stretchers. Some corner castings on the stretchers were found to be over-sized. Using an over-sized cone-shaped hole corner casting with an improperly seated shank pin could put undue stress on the top screw, possibly causing the screw securing a wheel to become loose or break and the wheel to detach from the stretcher

  • 2005-11-26TerminatedOther

    PROFlex Transporter, Model numbers 35PST. The catalog number for the model 35PST is PT3565.

    The firm received complaints of the stretchers folding. A fold is when the stretcher lowers on its own to either the next position, or any of the other seven postions of the stretcher.

  • 2005-11-26TerminatedOther

    PROFlex Transporter, Model number 35P. The catalog number for the model 35P is PT3550.

    The firm received complaints of the stretchers folding. A fold is when the stretcher lowers on its own to either the next position, or any of the other seven postions of the stretcher.