Skip to main content

Ethicon Endo-surgery, Inc.

FDA adverse event reports naming Ethicon Endo-surgery, Inc. as the device manufacturer, with recall history from the live FDA recall database. Reporter-supplied name as filed: ETHICON ENDO-SURGERY, INC..

Total reports
18,010
Deaths
87
Injuries
1,435
Malfunctions
16,273

Reports per year

12
13
14
15
16
17
18
19
20
21
22
23
24
25
26

Source: FDA adverse event reports by year received, 2012–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.

Top device categories

Recent recalls

  • 2025-10-14Open, ClassifiedUnder Investigation by firm

    ECHELON LINEAR Cutters Reload 80 mm Blue, intended for transection, resection, and/or creation of anastomoses, Product Number GLC80

    The devices were shipped unsterilized.

  • 2025-05-20Open, ClassifiedDevice Design

    Endopath Echelon Vascular White Reload for Advanced Placement Tip (35mm, 4 row); Product code VASECR35

    Firm has received an increase in reports regarding inadvertent instrument lockout during surgical procedures involving the device. If lockout occurs, the device will momentarily activate but will not cut or staple tissue, and additional steps will be required to open it and remove it from tissue. If user is unable to remove the device from the tissue, potential harms to patient include surgical delay, bleeding/hemorrhage, life threatening hemorrhage/hemorrhagic shock, and conversion to open therapy. Two adverse events received, including one patient death.

  • 2024-01-26Open, ClassifiedUnder Investigation by firm

    ENDOPATH ECHELON 45mm RELOADS WITH GRIPPING SURFACE TECHNOLOGY (GST), ECHELON Endopath Stapler, Endoscopic Linear Cutter Reload (+Gripping Surface technology), 45MM WHITE VASCULAR/THIN, Product No. GST45W

    There is the potential for incomplete staple line formation during the firing sequence, without most or all staples, which may lead to insufficient tissue apposition that could require surgical intervention. This may result in bleeding, prolonged surgery, or surgical delay.

  • 2019-10-29TerminatedNonconforming Material/Component

    ECHELON FLEX ENDOPATH 60mm Stapler - 340mm shaft, Product Code EC60A

    The staplers may contain an out of specification anvil component within the jaw of the device. This condition may lead to malformed staples, which can compromise staple line integrity. If the staple line is compromised, there is a potential risk of prolonged surgery, postoperative anastomotic leak, hemorrhage, hemorrhagic shock, additional surgical intervention, or death.

  • 2019-10-29TerminatedNonconforming Material/Component

    ECHELON FLEX Powered Plus ENDOPATH 60mm Stapler - 280mm shaft, Product Code PCEE60A

    The staplers may contain an out of specification anvil component within the jaw of the device. This condition may lead to malformed staples, which can compromise staple line integrity. If the staple line is compromised, there is a potential risk of prolonged surgery, postoperative anastomotic leak, hemorrhage, hemorrhagic shock, additional surgical intervention, or death.

  • 2019-10-29TerminatedNonconforming Material/Component

    ECHELON FLEX Powered Plus ENDOPATH 60mm Stapler - 440mm shaft, Product Code PLEE60A

    The staplers may contain an out of specification anvil component within the jaw of the device. This condition may lead to malformed staples, which can compromise staple line integrity. If the staple line is compromised, there is a potential risk of prolonged surgery, postoperative anastomotic leak, hemorrhage, hemorrhagic shock, additional surgical intervention, or death.

  • 2019-10-29TerminatedNonconforming Material/Component

    ECHELON FLEX Powered Plus ENDOPATH 60mm Stapler - 340mm shaft, Product Code PSEE60A

    The staplers may contain an out of specification anvil component within the jaw of the device. This condition may lead to malformed staples, which can compromise staple line integrity. If the staple line is compromised, there is a potential risk of prolonged surgery, postoperative anastomotic leak, hemorrhage, hemorrhagic shock, additional surgical intervention, or death.

  • 2016-10-20TerminatedDevice Design

    Endoscopic Articulating Linear Cutter, ATB35. Standard 35 MM. The ENDOPATH ETS-Flex Endoscopic Articulating Linear Cutter has application in general, gynecologic, urologic, and thoracic surgery for transection, resection, and/or creation of anastomoses

    Quality control identified a component issue where the pinion gear in the device could fail under extreme use cases. If this condition occurs, staples will be formed past the cut line and the device can be opened and removed from the patient. The firing stroke may be interrupted and the knife may not fully return to the home position potentially exposing the healthcare professional to a sharps injury.

  • 2016-10-20TerminatedDevice Design

    Endoscopic Articulating Linear Cutter, 38ATW35. VASCULAR/THIN 35MM The ENDOPATH ETS-Flex Endoscopic Articulating Linear Cutter has application in general, gynecologic, urologic, and thoracic surgery for transection, resection, and/or creation of anastomoses

    Quality control identified a component issue where the pinion gear in the device could fail under extreme use cases. If this condition occurs, staples will be formed past the cut line and the device can be opened and removed from the patient. The firing stroke may be interrupted and the knife may not fully return to the home position potentially exposing the healthcare professional to a sharps injury.

  • 2015-11-16TerminatedDevice Design

    HARMONIC ACE+ 7 Laparoscopic Shears with Advanced Hemostasis (23cm, 36cm, 45cm length), Ethicon Endo-Surgery. Indicated for soft tissue incisions when bleeding control and minimal thermal injury are desired.

    Complaints of holes in Tyvek which can compromise the sterile barrier provided by the packaging.

Monitor Ethicon Endo-surgery, Inc. automatically

DeviceWatch watches FDA adverse events and recalls for the products you care about, summarizes clinical narratives with AI and alerts you when a safety signal emerges.