Skip to main content

Ethicon Endo-surgery

FDA adverse event reports naming Ethicon Endo-surgery as the device manufacturer, with recall history from the live FDA recall database. Reporter-supplied name as filed: ETHICON ENDO-SURGERY.

Total reports
6,754
Deaths
27
Injuries
823
Malfunctions
5,813

Reports per year

07
08
09
10
11
12
13
14
15
16
17
18
19
20
21

Source: FDA adverse event reports by year received, 2007–2021. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.

Top device categories

Recent recalls

  • 2025-10-14Open, ClassifiedUnder Investigation by firm

    ECHELON LINEAR Cutters Reload 80 mm Blue, intended for transection, resection, and/or creation of anastomoses, Product Number GLC80

    The devices were shipped unsterilized.

  • 2025-05-20Open, ClassifiedDevice Design

    Endopath Echelon Vascular White Reload for Advanced Placement Tip (35mm, 4 row); Product code VASECR35

    Firm has received an increase in reports regarding inadvertent instrument lockout during surgical procedures involving the device. If lockout occurs, the device will momentarily activate but will not cut or staple tissue, and additional steps will be required to open it and remove it from tissue. If user is unable to remove the device from the tissue, potential harms to patient include surgical delay, bleeding/hemorrhage, life threatening hemorrhage/hemorrhagic shock, and conversion to open therapy. Two adverse events received, including one patient death.

  • 2025-04-03Open, ClassifiedProcess control

    PDSPLUSVIO30IN(75CM)USP6-0(M0.7)D/AC-1MP

    Identified curing issues with the silicone during the needles manufacturing process.

  • 2025-04-03Open, ClassifiedProcess control

    PDSIIVIO30IN(75CM)USP6-0(M0.7)D/AC-1

    Identified curing issues with the silicone during the needles manufacturing process.

  • 2025-04-03Open, ClassifiedProcess control

    PROLENE BLU 18IN(45CM) USP6-0(M0.7) S/A P-1 PRM

    Identified curing issues with the silicone during the needles manufacturing process.

  • 2025-04-03Open, ClassifiedProcess control

    ETHIBOND EXCEL GRN 36IN(90CM) USP2-0(M3) D/A MH

    Identified curing issues with the silicone during the needles manufacturing process.

  • 2024-01-26Open, ClassifiedUnder Investigation by firm

    ENDOPATH ECHELON 45mm RELOADS WITH GRIPPING SURFACE TECHNOLOGY (GST), ECHELON Endopath Stapler, Endoscopic Linear Cutter Reload (+Gripping Surface technology), 45MM WHITE VASCULAR/THIN, Product No. GST45W

    There is the potential for incomplete staple line formation during the firing sequence, without most or all staples, which may lead to insufficient tissue apposition that could require surgical intervention. This may result in bleeding, prolonged surgery, or surgical delay.

  • 2019-10-29TerminatedNonconforming Material/Component

    ECHELON FLEX ENDOPATH 60mm Stapler - 340mm shaft, Product Code EC60A

    The staplers may contain an out of specification anvil component within the jaw of the device. This condition may lead to malformed staples, which can compromise staple line integrity. If the staple line is compromised, there is a potential risk of prolonged surgery, postoperative anastomotic leak, hemorrhage, hemorrhagic shock, additional surgical intervention, or death.

  • 2019-10-29TerminatedNonconforming Material/Component

    ECHELON FLEX Powered Plus ENDOPATH 60mm Stapler - 280mm shaft, Product Code PCEE60A

    The staplers may contain an out of specification anvil component within the jaw of the device. This condition may lead to malformed staples, which can compromise staple line integrity. If the staple line is compromised, there is a potential risk of prolonged surgery, postoperative anastomotic leak, hemorrhage, hemorrhagic shock, additional surgical intervention, or death.

  • 2019-10-29TerminatedNonconforming Material/Component

    ECHELON FLEX Powered Plus ENDOPATH 60mm Stapler - 440mm shaft, Product Code PLEE60A

    The staplers may contain an out of specification anvil component within the jaw of the device. This condition may lead to malformed staples, which can compromise staple line integrity. If the staple line is compromised, there is a potential risk of prolonged surgery, postoperative anastomotic leak, hemorrhage, hemorrhagic shock, additional surgical intervention, or death.

Monitor Ethicon Endo-surgery automatically

DeviceWatch watches FDA adverse events and recalls for the products you care about, summarizes clinical narratives with AI and alerts you when a safety signal emerges.