Skip to main content

Dexcom Inc

FDA adverse event reports naming Dexcom Inc as the device manufacturer, with recall history from the live FDA recall database. Reporter-supplied name as filed: DEXCOM INC.

Total reports
1,705
Deaths
1
Injuries
30
Malfunctions
1,667

Reports per year

09
14
16
17
19
20
21
22
23
24
25
26

Source: FDA adverse event reports by year received, 2009–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.

Top device categories

Recent recalls

  • 2026-06-24Open, ClassifiedNonconforming Material/Component

    Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 US Retail Sensor/Transmitter 1-pack Model/Catalog Number: STP-AT-012 Software Version: N/A Product Description: Dexcom G7 US Retail Sensor/Transmitter 1-pack Component: Dexcom G7 Continuous Glucose Monitoring System

    Certain lots of Dexcom G7 sensors originally designated as scrap and intended for destruction were stolen during the destruction process. The sensors were later sold to pharmacies and distributors by Pharmsource, LLC. Lot 1725204004 were designated for destruction because the G7 sensors were not properly sterilized and have an increased risk of skin infection if used. Lot 1725069002 were designated for destruction due to an increased risk of having no sensor readings available or experiencing sensor inaccuracy that includes the potential for missed detection of a hyperglycemic event or incorrect treatment decisions that can result in patient injury, including medical intervention to prevent permanent injury or impairment. Additionally, due to the unknown storage and handling and storage conditions these sensors may have undergone during the distribution, these sensors may present additional risk.

  • 2026-05-11Open, ClassifiedSoftware design

    Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 CGM iOS App Model/Catalog Number: SW12300 Software Version: Versions 2.9.0 through 2.11.2 Product Description: Dexcom G7 CGM iOS App Component: Dexcom G7 Continuous Glucose Monitoring System, Dexcom G7 15 Day Continuous Glucose Monitoring System

    A software defect in which a low-priority file I/O check blocks the higher-priority thread responsible for processing estimated glucose values (EGVs) and trend arrows. When the app is running in the background, EGV notifications may queue up and then "replay" sequentially when the user brings the app to the foreground, causing the Glucose Compass to flicker through previous EGVs. In more severe instances, the app may perceive signal loss and deploy signal loss mitigations. This delay in EGV and alert processing creates a risk that users may miss detection of a hypoglycemic or hyperglycemic event, or make treatment decisions based on outdated glucose information.

  • 2026-05-11Open, ClassifiedSoftware design

    Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 CGM watchOS App Model/Catalog Number: SW13355 Software Version: Versions 2.9.0 through 2.11.2 Product Description: Dexcom G7 CGM watchOS App - The watchOS app is installed along with the iOS phone app and the versions used are the same for each user. The watchOS app cannot be installed independently from the iOS phone app. Component: Dexcom G7 Continuous Glucose Monitoring System, Dexcom G7 15 Day Continuous Glucose Monitoring System

    A software defect in which a low-priority file I/O check blocks the higher-priority thread responsible for processing estimated glucose values (EGVs) and trend arrows. When the app is running in the background, EGV notifications may queue up and then "replay" sequentially when the user brings the app to the foreground, causing the Glucose Compass to flicker through previous EGVs. In more severe instances, the app may perceive signal loss and deploy signal loss mitigations. This delay in EGV and alert processing creates a risk that users may miss detection of a hypoglycemic or hyperglycemic event, or make treatment decisions based on outdated glucose information.

  • 2026-05-11Open, ClassifiedSoftware design

    Brand Name: Dexcom ONE+ Continuous Glucose Monitoring System Product Name: Dexcom ONE+ iOS CGM App Model/Catalog Number: SW14244 Software Version: versions 1.5.0 and 1.6.0 Product Description: Dexcom ONE+ iOS CGM App Component: Dexcom ONE+ Continuous Glucose Monitoring System

    A software defect in which a low-priority file I/O check blocks the higher-priority thread responsible for processing estimated glucose values (EGVs) and trend arrows. When the app is running in the background, EGV notifications may queue up and then "replay" sequentially when the user brings the app to the foreground, causing the Glucose Compass to flicker through previous EGVs. In more severe instances, the app may perceive signal loss and deploy signal loss mitigations. This delay in EGV and alert processing creates a risk that users may miss detection of a hypoglycemic or hyperglycemic event, or make treatment decisions based on outdated glucose information.

  • 2025-10-30Open, ClassifiedSoftware Design Change

    Brand Name: Dexcom G6 Continuous Glucose Monitoring System Product Name: Dexcom G6 and G6 Pro Android US CGM App Model/Catalog Number: SW11678 Software Version: version 1.15.0 Product Description: Dexcom G6 and G6 Pro Android US CGM App SW11678 version 1.15.0 Component: Dexcom G6 Continuous Glucose Monitoring System

    A software defect in version v1.15.0 of the G6 Android app can cause the app to terminate unexpectedly, which may result in the user not receiving estimated glucose values, alarms, alerts or notifications. This could result in the missed detection of a hyperglycemic or hypoglycemic event, protentional resulting in severe hyperglycemia, diabetic ketoacidosis (DKA), or hyperosmolar hyperglycemic state (HHS).

  • 2025-09-04Open, ClassifiedSoftware design

    Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 CGM Android App Model/Catalog Number: SW12299 Software Version: version 2.8.0 and earlier Product Description: Dexcom G7 CGM Android App SW12299 Component: Dexcom G7 Continuous Glucose Monitoring System

    The affected devices are the Dexcom G7 Continuous Glucose Monitoring System (CGM System) using iOS, watchOS and Android App software (versions 2.8.0 or earlier) and Dexcom ONE+ CGM System using iOS and Android App software (versions 1.4.0 or earlier), henceforth referred to as the App. The affected App does not provide an expected "Sensor Failed" alert when the transmitter sends a "transmitterFailed" error message to the App when the CGM experiences a hardware or firmware failure. Instead, the App ends the CGM sensor session, stops reporting glucose values and displays the Start Sensor screen or No active sensor message without alerting the user, which can lead to missed detection of a hyperglycemic or hypoglycemic event and a delay in treatment.

  • 2025-09-04Open, ClassifiedSoftware design

    Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 CGM iOS App Model/Catalog Number: SW12300 Software Version: version 2.8.0 and earlier Product Description: Dexcom G7 CGM iOS App Component: Dexcom G7 Continuous Glucose Monitoring System

    The affected devices are the Dexcom G7 Continuous Glucose Monitoring System (CGM System) using iOS, watchOS and Android App software (versions 2.8.0 or earlier) and Dexcom ONE+ CGM System using iOS and Android App software (versions 1.4.0 or earlier), henceforth referred to as the App. The affected App does not provide an expected "Sensor Failed" alert when the transmitter sends a "transmitterFailed" error message to the App when the CGM experiences a hardware or firmware failure. Instead, the App ends the CGM sensor session, stops reporting glucose values and displays the Start Sensor screen or No active sensor message without alerting the user, which can lead to missed detection of a hyperglycemic or hypoglycemic event and a delay in treatment.

  • 2025-09-04Open, ClassifiedSoftware design

    Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 CGM watchOS App Model/Catalog Number: SW13355 Software Version: versions 2.8.0 and earlier Product Description: Dexcom G7 CGM watchOS App - The watchOS app is installed along with the iOS phone app and the versions used of each are the same for each user. The watchOS app cannot be installed independently from the iOS phone app Component: Dexcom G7 CGM iOS App

    The affected devices are the Dexcom G7 Continuous Glucose Monitoring System (CGM System) using iOS, watchOS and Android App software (versions 2.8.0 or earlier) and Dexcom ONE+ CGM System using iOS and Android App software (versions 1.4.0 or earlier), henceforth referred to as the App. The affected App does not provide an expected "Sensor Failed" alert when the transmitter sends a "transmitterFailed" error message to the App when the CGM experiences a hardware or firmware failure. Instead, the App ends the CGM sensor session, stops reporting glucose values and displays the Start Sensor screen or No active sensor message without alerting the user, which can lead to missed detection of a hyperglycemic or hypoglycemic event and a delay in treatment.

  • 2025-09-04Open, ClassifiedSoftware design

    Brand Name: Dexcom ONE+ Continuous Glucose Monitoring System Product Name: Dexcom ONE+ Android CGM App Model/Catalog Number: SW14245 Software Version: versions 1.4.0 and earlier Product Description: Dexcom ONE+ Androind CGM App Component: Dexcom ONE+ Continuous Glucose Monitoring System

    The affected devices are the Dexcom G7 Continuous Glucose Monitoring System (CGM System) using iOS, watchOS and Android App software (versions 2.8.0 or earlier) and Dexcom ONE+ CGM System using iOS and Android App software (versions 1.4.0 or earlier), henceforth referred to as the App. The affected App does not provide an expected "Sensor Failed" alert when the transmitter sends a "transmitterFailed" error message to the App when the CGM experiences a hardware or firmware failure. Instead, the App ends the CGM sensor session, stops reporting glucose values and displays the Start Sensor screen or No active sensor message without alerting the user, which can lead to missed detection of a hyperglycemic or hypoglycemic event and a delay in treatment.

  • 2025-09-04Open, ClassifiedSoftware design

    Brand Name: Dexcom ONE+ Continuous Glucose Monitoring System Product Name: Dexcom ONE+ iOS CGM App Model/Catalog Number: SW14244 Software Version: versions 1.4.0 and earlier Product Description: Dexcom ONE+ iOS CGM App Component: Dexcom ONE+ Continuous Glucose Monitoring System

    The affected devices are the Dexcom G7 Continuous Glucose Monitoring System (CGM System) using iOS, watchOS and Android App software (versions 2.8.0 or earlier) and Dexcom ONE+ CGM System using iOS and Android App software (versions 1.4.0 or earlier), henceforth referred to as the App. The affected App does not provide an expected "Sensor Failed" alert when the transmitter sends a "transmitterFailed" error message to the App when the CGM experiences a hardware or firmware failure. Instead, the App ends the CGM sensor session, stops reporting glucose values and displays the Start Sensor screen or No active sensor message without alerting the user, which can lead to missed detection of a hyperglycemic or hypoglycemic event and a delay in treatment.

Monitor Dexcom Inc automatically

DeviceWatch watches FDA adverse events and recalls for the products you care about, summarizes clinical narratives with AI and alerts you when a safety signal emerges.