Skip to main content

Bard

FDA adverse event reports naming Bard as the device manufacturer, with recall history from the live FDA recall database. Reporter-supplied name as filed: BARD.

Total reports
1,304
Deaths
22
Injuries
679
Malfunctions
494

Reports per year

10
11
12
13
14
15
16
17
18
19
20
21
22
23
24

Source: FDA adverse event reports by year received, 2010–2024. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.

Top device categories

Recent recalls

  • 2026-09-25Open, ClassifiedSoftware design

    EnCor EnCompass Breast Biopsy and Tissue Removal System Console (model number: ENCPCONSOLE) and Driver (model number: ENCPDRIVER) with software

    Breast biopsy and tissue removal system may experience the following errors: 1) Probe calibration and /or handheld driver failure (code HH01-00020009) , 2) Pressure sensor (code E01-0054000E), which prompts the user to restart and re-calibrate; errors may lead to interruption or prolongation of the biopsy procedure, or the need for additional tissue sampling or repeat biopsy.

  • 2026-09-11Open, ClassifiedMaterial/Component Contamination

    Power-Trialysis Short-Term Straight Dialysis Catheter, REF: 5605150; Power-Trialysis Short-Term Straight Dialysis Catheter, REF: 5605200; Power-Trialysis Short-Term Straight Dialysis Catheter, REF: 5605240; Power-Trialysis Short-Term Straight Dialysis Catheter, REF: 5605300; Power-Trialysis Short-Term Curved Extension Dialysis Catheter, REF: 5615120; Power-Trialysis Short-Term Curved Extension Dialysis Catheter, REF: 5615150; Power-Trialysis Short-Term Curved Extension Dialysis Catheter, REF: 5615200; Power-Trialysis Short-Term Curved Extension Dialysis Catheter, REF: 5615240; Power-Trialysis Short-Term Alphacurve Dialysis Catheter, REF: 5655120; Power-Trialysis Short-Term Alphacurve Dialysis Catheter, REF: 5655150; Power-Trialysis Short-Term Alphacurve Dialysis Catheter, REF: 5655200; Power-Trialysis Short-Term Alphacurve Dialysis Catheter, REF: 5655240; Power-Trialysis Short-Term Straight Dialysis Catheter Full Procedure Tray, REF: 5605200G; Power-Trialysis Slim-Cath Straight Short-Term Dialysis Catheter, REF: 5855200G; Power-Trialysis Slim-Cath Precurved Extension Short-Term Dialysis Catheter, REF: 5865150G; Power-Trialysis Slim-Cath Straight Short-Term Dialysis Catheter, REF: 5855150; Power-Trialysis Slim-Cath Straight Short-Term Dialysis Catheter, REF: 5855200; Power-Trialysis Slim-Cath Straight Short-Term Dialysis Catheter, REF: 5855240; Power-Trialysis Slim-Cath Straight Short-Term Dialysis Catheter, REF: 5855300; Power-Trialysis Slim-Cath Precurved Extension Short-Term Dialysis Catheter, REF: 5865120; Power-Trialysis Slim-Cath Precurved Extension Short-Term Dialysis Catheter, REF: 5865150; Power-Trialysis Slim-Cath Precurved Extension Short-Term Dialysis Catheter, REF: 5865200; Power-Trialysis Slim-Cath Precurved Extension Short-Term Dialysis Catheter, REF: 5865240

    this is a downstream recall of class I recall RES 98777

  • 2026-09-11Open, ClassifiedMaterial/Component Contamination

    DuoGlide Short-Term Straight Dialysis Catheter, REF: 5625150; DuoGlide Short-Term Straight Dialysis Catheter, REF: 5625200; DuoGlide Short-Term Straight Dialysis Catheter, REF: 5625240; DuoGlide Short-Term Straight Dialysis Catheter, REF: 5625300; DuoGlide Short-Term Curved Extension Dialysis Catheter, REF: 5626120; DuoGlide Short-Term Curved Extension Dialysis Catheter, REF: 5626150; DuoGlide Short-Term Curved Extension Dialysis Catheter, REF: 5626200; DuoGlide Short-Term Curved Extension Dialysis Catheter, REF: 5626240; Niagara Dual-Lumen Short-Term Curved Extension Dialysis Catheter, REF: 5766200

    this is a downstream recall of class I recall RES 98777

  • 2026-09-04Open, ClassifiedUnder Investigation by firm

    Products that contain the Pericare Wipes: Bardex Lubricath Foley Catheter Tray REF: 899914 899916 899918 UDI codes: 00801741024801 10801741024808 00801741024818 10801741024815 00801741024825 10801741024822 SureStep" Foley Tray System with Bardex Lubricath Temp Sensing Foley Catheter REF: A119416M UDI code: 00801741073700 10801741073707 SureStep" Foley Tray System Bard Lubricath Foley REF: A800361 UDI code: 00801741073939 10801741073936 SureStep" Foley Tray System Bard Lubricath Foley Catheter Tray REF: A899616 UDI code: 00801741104534 10801741104531 SureStep" Foley Tray System Bardex Lubricath Foley Catheter REF: A899914 UDI code: 00801741074004 10801741074001 SureStep" Foley Tray System Bardex Lubricath Foley Catheter REF: A899916 UDI code: 00801741074011 10801741074018 SureStep" Foley Tray System Bardex Lubricath Temperature Sensing Catheter REF: A900216 UDI code: 00801741120978 10801741120975 SureStep" Foley Tray System BardLubricath Foley Catheter REF: A902914 A902916 A902918 UDI code: 0080174107406610801741074063 00801741074073 10801741074070 0080174107408010801741074087 Bard Urine Meter Foley Catheter Tray With Bardex Lubricath Foley Catheter REF: 304916 UDI code: 00801741019736 01801741019739 Bard Urine Meter Foley Catheter Tray With Bardex Lubricath Foley Catheter REF: 902916 UDI codes: 00801741024863 10801741024860 (416A) SureStep Foley Tray System Bardex IC Complete Care (714A/716A/718A)SureStep Foley Tray System Bardex IC Complete REF: A304914 A304916 A304918 UDI codes: 00801741104350 10801741104357 00801741104367 10801741104364 00801741100499 10801741100496 SureStep" Foley Tray System Bard Lubricath Foley REF: A800365 UDI codes: 00801741073960 10801741073967 SureStep" Foley Tray System Bardex Lubricath Foley Catheter REF: A899918 UDI codes: 00801741074028 10801741074025 SureStep" Foley Tray System BardLubricath Foley Catheter REF: A902618 UDI codes: 00801741104640 10801741104647 SureStep" Foley Tray System BardLubricath Foley Catheter REF: A902616 UDI codes: 00801741104633 10801741104630 SureStep Intermittent Catheter Tray REF: INTL16C UDI codes: 00801741183195 10801741183192 For urological use only, Intended for use in the drainage and/or collection and/or measurement of urine.

    Due to complaints received indicating presence of contamination on the surface of pericare wipes provided with various Foley and Intermittent Catheter Trays.

  • 2026-09-04Open, ClassifiedUnder Investigation by firm

    Products that contain the Pericare Wipes: Bard Foley Tray With Lubri-Sil All-Silicone Foley Catheter REF: 947314 947316 947318 UDI codes: 00801741024924 10801741024921 00801741024931 10801741024938 00801741024948 10801741024945 Lubri-Sil I.C. Complete Care All-Silicone Foley Catheter Tray REF: 300416A UDI code: 00801741024580 10801741024587 SureStep" Foley Tray System Lubri-Sil IC Complete Care REF: A300416A UDI code: 00801741073762 10801741073769 SureStep" Foley Tray System Lubri-Sil IC Complete Care REF: A303416A UDI code: 00801741073823 10801741073820 SureStep" Foley Tray System Lubri-Sil IC Complete Care REF: A307416A A307418A UDI codes: 00801741104381 10801741104388 00801741100505 10801741100502 SureStep" Foley Tray System Bard Silicone Foley Catheter Tray REF: A897516 UDI code: 00801741073984 10801741073981 SureStep" Foley Tray System Bardex All- Silicone Foley Catheter Tray REF: A902416 UDI code: 00801741074042 10801741074049 SureStep" Foley Tray System Bardex All- Silicone Foley Catheter Tray REF: A902418 UDI code: 00801741074059 10801741074056 SureStep" Foley Tray System Bard Lubri-Sil Foley Catheter REF: A907316 UDI code: 00801741104749 10801741104746 SureStep" Foley Tray System with Bard Lubri-Sil Catheter REF: A942210 UDI code: 00801741100567 10801741100564 SureStep" Foley Tray System with Bard Lubri-Sil Catheter REF: A942212 UDI code: 00801741100536 10801741100533 SureStep" Foley Tray System Bard Lubri-Sil Foley Catheter Tray REF: A942214 UDI code: 00801741074103 10801741074100 SureStep" Foley Tray System Bard Lubri-Sil Foley Catheter Tray REF: A942216 UDI code: 00801741074110 10801741074117 SureStep" Foley Tray System Bard Lubri-Sil Foley Catheter REF: A947316 UDI code: 00801741074141 10801741074148 SureStep" Foley Tray System with Bard Lubri-Sil Catheter REF: A942206 UDI code: 00801741100543 10801741100540 **** Lubri-Sil I.C. Complete Care Temp-Sensing Foley Catheter Tray and Urine Meter REF: 319516AM UDI code: 00801741024696 10801741024693 SureStep" Foley Tray System with Bard Lubri-Sil Temp Sensing Foley Catheter REF: A119214M UDI code: 00801741073670 10801741073677 REF: A119216M UDI code: 00801741073687 10801741073684 SureStep" Foley Tray Lubri-Sil I.C. Complete Care Temp Sensing Catheter REF: A300916A UDI code: 00801741120862 10801741120869 SureStep" Foley Tray System Lubri-Sil IC Complete Care REF: A319516A UDI code: 00801741100338 10801741100335 SureStep" Foley Tray System Lubri-Sil IC Complete Care REF: A319516AM UDI code: 00801741073892 10801741073899 SureStep" Foley Tray Lubri-Sil I.C. Complete Care Temp Sensing Catheter REF: A900916A UDI code: 00801741121029 10801741121026 SureStep" Foley Tray System Lubri-Sil Temperature Sensing Foley Catheter REF: A909216M UDI code: 00801741104763 10801741104760 ***** Bard Urine Meter Foley Catheter Tray With Bard Lubri-Sil Catheter REF 942208 UDI code: 00801741029370 10801741029377 Bard Urine Meter Foley Tray With Lubri-Sil All-Silicone Foley Catheter REF: 942216 UDI code: 00801741024979 10801741024976 Lubri-Sil I.C. Complete Care All-Silicone Foley Catheter Tray and Urine Meter REF: 303416A UDI code: 00801741024726 10801741024723 SureStep" Foley Tray System with Bard Lubri-Sil Temp Sensing Foley Catheter REF: A119218M UDI code: 00801741073694 10801741073691 SureStep" Foley Tray System Lubri-Sil IC Complete Care REF: A300418A UDI code: 00801741073779 10801741073776 SureStep" Foley Tray System Lubri-Sil IC Complete Care REF: A303418A UDI code: 00801741073830 10801741073837 SureStep" Foley Tray System Lubri-Sil IC Complete Care REF: A319514AM UDI code: 00801741073885 10801741073882 SureStep" Foley Tray System with Bard Lubri-Sil Catheter REF: A942208 UDI code: 00801741100550 10801741100557 SureStep" Foley Tray System Bard Lubri-Sil Foley Catheter Tray REF: A942218 UDI code: 00801741074127 10801741074124 SureStep" Foley Tray System Bard Lubri-Sil Foley

    Due to complaints received indicating presence of contamination on the surface of pericare wipes provided with various Foley and Intermittent Catheter Trays.

  • 2026-09-04Open, ClassifiedUnder Investigation by firm

    Products that contain the Pericare Wipes: *Bardex I.C. Complete Care Foley Catheter Tray REF: 300316A UDI codes: 00801741024344 10801741024341 *Bardex I.C. Complete Care Foley Catheter Tray and Urine Meter REF: 303314A 303316A 303318A UDI code: 00801741024252 10801741024259 00801741024269 10801741024266 00801741024276 10801741024273 *SureStep" Foley Tray System Bardex IC Complete Care REF: A300314A UDI codes: 00801741073724 10801741073721 REF: A300314AC (Canada Only) UDI-DI code: 00801741143243 *SureStep" Foley Tray System Bardex IC Complete Care REF: A300316A UDI code: 00801741073731 10801741073738 *SureStep" Foley Tray System Bardex IC Complete Care REF: A300318A UDI code: 00801741073748 10801741073745 *SureStep" Foley Tray System Bardex IC Complete Care REF: A303314A A303316A A303318A UDI code: 00801741073786 10801741073783 00801741073793 10801741073790 00801741073809 10801741073806 *SureStep" Foley Tray System Lubri-Sil IC Complete Care REF: A303414A UDI code: 00801741073816 10801741073813 *(416A)SureStep" Foley Tray System Bardex IC Complete Care (714A/716A/718A)SureStep" Foley Tray System Bardex IC Complete Care (914)SureStep" Foley Tray System REF: A304716A A304714A A304716A A304718A UDI code: 00801741104206 10801741104203 00801741104268 10801741104265 00801741104299 10801741104296 00801741100376 10801741100373 *SureStep" Foley Tray System Bardex IC Complete Care REF: A319416A UDI code: 00801741100307 10801741100304 *SureStep" Foley Tray System Bardex I.C. Complete Care Foley Catheter REF: A319416AM UDI code: 00801741073861 10801741073868 *SureStep" Foley Tray System Bardex IC Complete Care REF: A319418A UDI code: 00801741100314 10801741100311 SureStep" Foley Tray System Bardex IC Complete Care REF: A903316A UDI code: 00801741096600 10801741096607 ***** SureStep" Foley Tray System Lubri-Sil IC Complete Care REF: A300414A UDI code: 00801741073755 10801741073752 SureStep" Foley Tray System Bardex I.C. Complete Care Foley Catheter REF: A319414AM UDI code: 00801741073854 10801741073851 SureStep" Foley Tray System Bardex I.C. Complete Care Foley Catheter REF: A319418AM UDI code: 00801741073878 10801741073875 SureStep" Foley Tray System Bardex IC Complete Care REF: A903318A UDI code: 00801741104657 10801741104654 For urological use only, Intended for use in the drainage and/or collection and/or measurement of urine.

    Due to complaints received indicating presence of contamination on the surface of pericare wipes provided with various Foley and Intermittent Catheter Trays.

  • 2026-09-04Open, ClassifiedUnder Investigation by firm

    Products that contain the Pericare Wipes: SureStep Intermittent Catheter Tray REF: INTP16 INTS14 INTS14C INTS16 INTS16C PEDINTS06 PEDINTS08 PEDINTS10 UDI code: 00801741183300 10801741183307 00801741183218 10801741183215 00801741183232 10801741183239 00801741183225 10801741183222 00801741183249 10801741183246 00801741183263 10801741183260 00801741183270 10801741183277 00801741183287 10801741183284 Preamendment REF: INTL14 INTL15 INTL16 INTP14 UDI code: 00801741183164 10801741183161 00801741183171 10801741183178 00801741183188 10801741183185 00801741183294 10801741183291 For urological use only, Intended for use in the drainage and/or collection and/or measurement of urine.

    Due to complaints received indicating presence of contamination on the surface of pericare wipes provided with various Foley and Intermittent Catheter Trays.

  • 2026-09-04Open, ClassifiedUnder Investigation by firm

    Products that contain the Pericare Wipes: SureStep" Foley Tray System Add-A-Foley Catheter REF: A153200 UDI code: 00801741121067 10801741121064 SureStep" Foley Tray System Complete Care Add-A-Foley Catheter REF: A304400A UDI code: 00801741073847 10801741073844 SureStep" Foley Tray System Add-A-Foley Catheter REF: A399400 UDI code: 00801741100369 10801741100366 SureStep" Foley Tray System Complete Care Add-A-Foley Catheter REF: A399400A UDI code: 00801741073915 10801741073912 SureStep" Foley Tray System Complete Care Add-A-Foley Catheter REF: A904400A UDI code: 00801741074097 10801741074094 SureStep Intermittent Catheter Tray REF: INTAAI UDI code: 00801741183317 10801741183314 For urological use only, Intended for use in the drainage and/or collection and/or measurement of urine.

    Due to complaints received indicating presence of contamination on the surface of pericare wipes provided with various Foley and Intermittent Catheter Trays.

  • 2026-09-04Open, ClassifiedUnder Investigation by firm

    Products that contain the Pericare Wipes: (Canada Only) SureStep" Foley Tray System Bardex IC Complete Care REF: A300314AC UDI-DI code: 00801741143243 SureStep" Foley Tray System Bardex IC Complete Care REF: A300316AC UDI-DI code: 00801741143250 SureStep" Foley Tray System Lubri-Sil IC Complete Care REF: A319516AMC UDI-DI code: 00801741143472 SureStep" Foley Tray System Bard Lubricath Foley REF: A800366C UDI-DI code: 00801741143571 SureStep" Foley Tray System Bard Silicone Foley Catheter Tray REF: A897514C UDI-DI code: 00801741074011 SureStep" Foley Tray System Bard Silicone Foley Catheter Tray REF: A897516C UDI-DI code: 00801741143595 SureStep" Foley Tray System Bard Silicone Foley Catheter Tray REF: A902114C UDI-DI code: 00801741144325 ***** SureStep" Foley Tray System Lubri-Sil IC Complete Care REF: A300416AC UDI code: 00801741143281 10801741143288 SureStep" Foley Tray System Bardex IC Complete Care REF: A303314AC UDI code: 00801741143304 10801741143301 SureStep" Foley Tray System Lubri-Sil IC Complete Care REF: A319514AMC UDI code: 00801741143465 10801741143462 SureStep" Foley Tray System Bard Lubricath Foley Catheter Tray REF: A800061C UDI code: 00801741144172 10801741144179 SureStep" Foley Tray System Bard Lubricath Foley Catheter Tray REF: A800065C UDI code: 00801741144202 10801741144209 SureStep" Foley Tray System Bard Lubricath Foley Catheter Tray REF: A800360C UDI code: 00801741143519 10801741143516 SureStep" Foley Tray System Bard Lubricath Foley Catheter Tray REF:A800361C UDI code: 00801741143526 10801741143523 SureStep" Foley Tray System Bard Lubricath Foley Catheter Tray REF: A800362C UDI code: 00801741143540 10801741143547 SureStep" Foley Tray System Bard Lubricath Foley REF: A800365C UDI code: 00801741143564 10801741143561 For urological use only, Intended for use in the drainage and/or collection and/or measurement of urine.

    Due to complaints received indicating presence of contamination on the surface of pericare wipes provided with various Foley and Intermittent Catheter Trays.

  • 2026-08-14Open, ClassifiedComponent change control

    BD Needle Kit for Powered Driver 15mm x 15Ga, REF: D015151NK, BD Needle Kit for Powered Driver 25mm x 15Ga, REF: D015251NK, BD Needle Kit for Powered Driver 35mm x 15Ga, REF: D015351NK, BD Needle Kit for Powered Driver 45mm x 15Ga, REF: D015451NK, BD Needle Kit for Powered Driver 55mm x 15Ga, REF: D015551NK

    After placement of intraosseous (IO) needle sets, used when intravenous access is difficult or impossible to obtain in emergent, urgent, or medically necessary cases, users may experience difficulty/inability to remove obturator, due to it locking to the needle hub, due to out-of-tolerance dimensions and IFU unclarity, which may result in new needle placement, which may result in delayed therapy

Monitor Bard automatically