Stent, Iliac Vein
FDA adverse event reports and recall history for stent, iliac vein (product code QAN), from the public MAUDE corpus and the live FDA recall database.
- Total reports
- 1,324
- Deaths
- 8
- Injuries
- 557
- Malfunctions
- 759
Reports per year
Source: FDA adverse event reports by year received, 2019–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.
Recent recalls
- 2026-01-15Open, ClassifiedUnder Investigation by firm
Abre Venous Self-Expanding Stent System, Model numbers AB9G14120090 AB9G12100090 AB9G14080090 AB9G12080090 AB9U12060090 AB9U14100090 AB9U14080090 AB9U12080090 AB9U14060090 AB9G14100090 AB9G12060090
Stents may be impacted by a Nitinol anomaly which causes a reduced stent diameter when deployed and corresponding reduced radial force. Issue may result in stent migration.
- 2024-03-19Open, ClassifiedProcess control
Abre Venous Self-expanding Stent System, Product Numbers: a) AB9U14080090; b) AB9U18100090; c) AB9G14100090
There is a potential for sterile package breach.
- TerminatedOther
Instructions for Use for all model numbers of the Medtronic Abre Venous Self-Expanding Stent System.
There have been reports of stent migration.
- TerminatedProcess control
Medtronic Abre Venous Self-expanding Sent System, Catalog number REF AB9U16080090 - Product Usage: intended for use in the iliofemoral veins for the treatment of symptomatic venous outflow obstruction.
Incorrect size printed on the device; packaging is labeled correctly.
- Open, ClassifiedUnder Investigation by firm
VICI VENOUS STENT System Product Usage: is intended for the treatment of obstructions and occlusions in the venous vasculature.
The firm has received reports of stent migration after implantation
- Open, ClassifiedUnder Investigation by firm
VICI RDS VENOUS STENT System Product Usage: is indicated for improving luminal diameter in the iliofemoral veins for the treatment of symptomatic venous outflow obstruction.
The firm has received reports of stent migration after implantation
- Open, ClassifiedUnder Investigation by firm
VENOVO Venous Stent System 10F Product Description(REF)/dimension: VENEL16040 Venovo 10F 16/40/1200mm OUS VENEL16060 Venovo 10F 16/60/1200mm OUS VENEL16080 Venovo 10F 16/80/1200mm OUS VENEL16100 Venovo 10F 16/100/1200mm OUS VENEL16120 Venovo 10F 16/120/1200mm OUS VENEL16140 Venovo 10F 16/140/1200mm OUS VENEL16160 Venovo 10F 16/160/1200mm OUS VENEL18040 Venovo 10F 18/40/1200mm OUS VENEL18060 Venovo 10F 18/60/1200mm OUS VENEL18080 Venovo 10F 18/80/1200mm OUS VENEL18100 Venovo 10F 18/100/1200mm OUS VENEL18120 Venovo 10F 18/120/1200mm OUS VENEL18140 Venovo 10F 18/140/1200mm OUS VENEL18160 Venovo 10F 18/160/1200mm OUS VENEL20040 Venovo 10F 20/40/1200mm OUS VENEL20060 Venovo 10F 20/60/1200mm OUS VENEL20080 Venovo 10F 20/80/1200mm OUS VENEL20100 Venovo 10F 20/100/1200mm OUS VENEL20120 Venovo 10F 20/120/1200mm OUS VENEL20140 Venovo 10F 20/140/1200mm OUS VENEL20160 Venovo 10F 20/160/1200mm OUS VENEM16040 Venovo 10F 16/40/800mm OUS VENEM16060 Venovo 10F 16/60/800mm OUS VENEM16080 Venovo 10F 16/80/800mm OUS VENEM16100 Venovo 10F 16/100/800mm OUS VENEM16120 Venovo 10F 16/120/800mm OUS VENEM16140 Venovo 10F 16/140/800mm OUS VENEM16160 Venovo 10F 16/160/800mm OUS VENEM18040 Venovo 10F 18/40/800mm OUS VENEM18060 Venovo 10F 18/60/800mm OUS VENEM18080 Venovo 10F 18/80/800mm OUS VENEM18100 Venovo 10F 18/100/800mm OUS VENEM18120 Venovo 10F 18/120/800mm OUS VENEM18140 Venovo 10F 18/140/800mm OUS VENEM18160 Venovo 10F 18/160/800mm OUS VENEM20040 Venovo 10F 20/40/800mm OUS VENEM20060 Venovo 10F 20/60/800mm OUS VENEM20080 Venovo 10F 20/80/800mm OUS VENEM20100 Venovo 10F 20/100/800mm OUS VENEM20120 Venovo 10F 20/120/800mm OUS VENEM20140 Venovo 10F 20/140/800mm OUS VENEM20160 Venovo 10F 20/160/800mm OUS VENUL16040 Venovo 10F 16/40/1200mm US VENUL16060 Venovo 10F 16/60/1200mm US VENUL16080 Venovo 10F 16/80/1200mm US VENUL16100 Venovo 10F 16/100/1200mm US VENUL16120 Venovo 10F 16/120/1200mm US VENUL16140 Venovo 10F 16/140/1200mm US VENUL16160 Venovo 10F 16/160/1200mm US VENUL18040 Venovo 10F 18/40/1200mm US VENUL18060 Venovo 10F 18/60/1200mm US VENUL18080 Venovo 10F 18/80/1200mm US VENUL18100 Venovo 10F 18/100/1200mm US VENUL18120 Venovo 10F 18/120/1200mm US VENUL18140 Venovo 10F 18/140/1200mm US VENUL18160 Venovo 10F 18/160/1200mm US VENUL20040 Venovo 10F 20/40/1200mm US VENUL20060 Venovo 10F 20/60/1200mm US VENUL20080 Venovo 10F 20/80/1200mm US VENUL20100 Venovo 10F 20/100/1200mm US VENUL20120 Venovo 10F 20/120/1200mm US VENUL20140 Venovo 10F 20/140/1200mm US VENUL20160 Venovo 10F 20/160/1200mm US VENUM16040 Venovo 10F 16/40/800mm US VENUM16060 Venovo 10F 16/60/800mm US VENUM16080 Venovo 10F 16/80/800mm US VENUM16100 Venovo 10F 16/100/800mm US VENUM16120 Venovo 10F 16/120/800mm US VENUM16140 Venovo 10F 16/140/800mm US VENUM16160 Venovo 10F 16/160/800mm US VENUM18040 Venovo 10F 18/40/800mm US VENUM18060 Venovo 10F 18/60/800mm US VENUM18080 Venovo 10F 18/80/800mm US VENUM18100 Venovo 10F 18/100/800mm US VENUM18120 Venovo 10F 18/120/800mm US VENUM18140 Venovo 10F 18/140/800mm US VENUM18160 Venovo 10F 18/160/800mm US VENUM20040 Venovo 10F 20/40/800mm US VENUM20060 Venovo 10F 20/60/800mm US VENUM20080 Venovo 10F 20/80/800mm US VENUM20100 Venovo 10F 20/100/800mm US VENUM20120 Venovo 10F 20/120/800mm US VENUM20140 Venovo 10F 20/140/800mm US VENUM20160 Venovo 10F 20/160/800mm US
Reports of the proximal end of the stent not immediately expanding upon deployment and remaining connected to the delivery system.
- Open, ClassifiedUnder Investigation by firm
VENOVO Venous Stent System 8F Product Description(REF)/dimension: VENEL10040 /Venovo 8F 10/40/1200mm OUS; VENEL10060 /Venovo 8F 10/60/1200mm OUS; VENEL10080 /Venovo 8F 10/80/1200mm OUS; VENEL10100 /Venovo 8F 10/100/1200mm OUS; VENEL10120 /Venovo 8F 10/120/1200mm OUS; VENEL10140 /Venovo 8F 10/140/1200mm OUS; VENEL10160 /Venovo 8F10/160/1200mm OUS; VENEL12040 /Venovo 8F 12/40/1200mm OUS; VENEL12060 /Venovo 8F 12/60/1200mm OUS; VENEL12080 /Venovo 8F 12/80/1200mm OUS; VENEL12100 /Venovo 8F 12/100/1200mm OUS; VENEL12120 /Venovo 8F 12/120/1200mm OUS; VENEL12140 /Venovo 8F 12/140/1200mm OUS; VENEL12160 /Venovo 8F 12/160/1200mm OUS; VENEM10040 /Venovo 8F 10/40/800mm OUS; VENEM10060 /Venovo 8F 10/60/800mm OUS; VENEM10080 /Venovo 8F 10/80/800mm OUS; VENEM10100 /Venovo 8F 10/100/800mm OUS; VENEM10120 /Venovo 8F 10/120/800mm OUS; VENEM10140 /Venovo 8F 10/140/800mm OUS; VENEM10160 /Venovo 8F 10/160/800mm OUS; VENEM12040 /Venovo 8F 12/40/800mm OUS; VENEM12060 /Venovo 8F 12/60/800mm OUS; VENEM12080 /Venovo 8F 12/80/800mm OUS; VENEM12100 /Venovo 8F 12/100/800mm OUS; VENEM12120 /Venovo 8F 12/120/800mm OUS; VENEM12140 /Venovo 8F 12/140/800mm OUS; VENEM12160 /Venovo 8F 12/160/800mm OUS; VENUL10040 /Venovo 8F 10/40/1200mm US; VENUL10060 /Venovo 8F 10/60/1200mm US; VENUL10080 /Venovo 8F 10/80/1200mm US; VENUL10100 /Venovo 8F 10/100/1200mm US; VENUL10120 /Venovo 8F 10/120/1200mm US; VENUL10140 /Venovo 8F 10/140/1200mm US; VENUL10160 /Venovo 8F 10/160/1200mm US; VENUL12040 /Venovo 8F 12/40/1200mm US; VENUL12060 /Venovo 8F 12/60/1200mm US; VENUL12080 /Venovo 8F 12/80/1200mm US; VENUL12100 /Venovo 8F 12/100/1200mm US; VENUL12120 /Venovo 8F 12/120/1200mm US; VENUL12140 /Venovo 8F 12/140/1200mm US; VENUL12160 /Venovo 8F 12/160/1200mm US; VENUM10040 /Venovo 8F 10/40/800mm US; VENUM10060 /Venovo 8F 10/60/800mm US; VENUM10080 /Venovo 8F 10/80/800mm US; VENUM10100 /Venovo 8F 10/100/800mm US; VENUM10120 /Venovo 8F 10/120/800mm US; VENUM10140 /Venovo 8F 10/140/800mm US; VENUM10160 /Venovo 8F 10/160/800mm US; VENUM12040 /Venovo 8F 12/40/800mm US; VENUM12060 /Venovo 8F 12/60/800mm US; VENUM12080 /Venovo 8F 12/80/800mm US; VENUM12100 /Venovo 8F 12/100/800mm US; VENUM12120 /Venovo 8F 12/120/800mm US; VENUM12140 /Venovo 8F 12/140/800mm US; VENUM12160 /Venovo 8F 12/160/800mm US;
Reports of the proximal end of the stent not immediately expanding upon deployment and remaining connected to the delivery system.
- Open, ClassifiedUnder Investigation by firm
VENOVO Venous Stent System 9F Product Description(REF)/dimension: VENEL14040 /Venovo 9F 14/40/1200mm OUS; VENEL14060 /Venovo 9F 14/60/1200mm OUS; VENEL14080 /Venovo 9F 14/80/1200mm OUS; VENEL14100 /Venovo 9F 14/100/1200mm OUS; VENEL14120 /Venovo 9F 14/120/1200mm OUS; VENEL14140 /Venovo 9F 14/140/1200mm OUS; VENEL14160 /Venovo 9F 14/160/1200mm OUS; VENEM14040 /Venovo 9F 14/40/800mm OUS; VENEM14060 /Venovo 9F 14/60/800mm OUS; VENEM14080 /Venovo 9F 14/80/800mm OUS; VENEM14100 /Venovo 9F 14/100/800mm OUS; VENEM14120 /Venovo 9F 14/120/800mm OUS; VENEM14140 /Venovo 9F14/140/800mm OUS; VENEM14160 /Venovo 9F 14/160/800mm OUS; VENUL14040 /Venovo 9F 14/40/1200mm US; VENUL14060 /Venovo 9F 14/60/1200mm US; VENUL14080 /Venovo 9F 14/80/1200mm US; VENUL14100 /Venovo 9F 14/100/1200mm US; VENUL14120 /Venovo 9F 14/120/1200mm US; VENUL14140 /Venovo 9F 14/140/1200mm US; VENUL14160 /Venovo 9F 14/160/1200mm US; VENUM14040 /Venovo 9F 14/40/800mm US; VENUM14060 /Venovo 9F 14/60/800mm US; VENUM14080 /Venovo 9F 14/80/800mm US; VENUM14100 /Venovo 9F 14/100/800mm US; VENUM14120 /Venovo 9F 14/120/800mm US; VENUM14140 /Venovo 9F 14/140/800mm US; VENUM14160 /Venovo 9F 14/160/800mm US;
Reports of the proximal end of the stent not immediately expanding upon deployment and remaining connected to the delivery system.
Monitor stent, iliac vein automatically