Catheter, Percutaneous, Cutting/Scoring
FDA adverse event reports and recall history for catheter, percutaneous, cutting/scoring (product code PNO), from the public MAUDE corpus and the live FDA recall database.
- Total reports
- 2,083
- Deaths
- 2
- Injuries
- 199
- Malfunctions
- 1,882
Reports per year
Source: FDA adverse event reports by year received, 2016–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.
Recent recalls
- 2019-06-17TerminatedUnder Investigation by firm
Advance Enforcer 35 Focal Force PTA Balloon Catheter 6mm x 4cm, 5FR/50cm, Catalog Number ASB5-35-50-6-4, REF Number G35248 Product Usage: The Advance Enforcer 35 Focal-Force PTA Balloon Catheter is intended for percutaneous transluminal angioplasty (PTA) of lesions in peripheral arteries, including iliac, renal, popliteal, infrapopliteal, femoral and iliofemoral, as well as obstructive lesions of native or synthetic arteriovenous dialysis fistulae. Not for use in the cerebral or coronary vasculature.
Multiple complaints were received for balloons bursting below the rated burst pressure. Potential adverse events that may occur if an affected product is used include a delay in the procedure, additional intervention, vessel injury, and balloon fragmentation in the patient.
- 2019-06-17TerminatedUnder Investigation by firm
Advance Enforcer 35 Focal Force PTA Balloon Catheter 6mm x 4cm, 5FR/80cm, Catalog Number ASB5-35-80-6-4, REF Number G35252 Product Usage: The Advance Enforcer 35 Focal-Force PTA Balloon Catheter is intended for percutaneous transluminal angioplasty (PTA) of lesions in peripheral arteries, including iliac, renal, popliteal, infrapopliteal, femoral and iliofemoral, as well as obstructive lesions of native or synthetic arteriovenous dialysis fistulae. Not for use in the cerebral or coronary vasculature.
Multiple complaints were received for balloons bursting below the rated burst pressure. Potential adverse events that may occur if an affected product is used include a delay in the procedure, additional intervention, vessel injury, and balloon fragmentation in the patient.
- 2019-06-17TerminatedUnder Investigation by firm
Advance Enforcer 35 Focal Force PTA Balloon Catheter 6mm x 4cm, 5FR/135cm, Catalog Number ASB5-35-135-6-4, REF Number G35257 Product Usage: The Advance Enforcer 35 Focal-Force PTA Balloon Catheter is intended for percutaneous transluminal angioplasty (PTA) of lesions in peripheral arteries, including iliac, renal, popliteal, infrapopliteal, femoral and iliofemoral, as well as obstructive lesions of native or synthetic arteriovenous dialysis fistulae. Not for use in the cerebral or coronary vasculature.
Multiple complaints were received for balloons bursting below the rated burst pressure. Potential adverse events that may occur if an affected product is used include a delay in the procedure, additional intervention, vessel injury, and balloon fragmentation in the patient.
Monitor catheter, percutaneous, cutting/scoring automatically
DeviceWatch watches FDA adverse events and recalls for the products you care about, summarizes clinical narratives with AI and alerts you when a safety signal emerges.