PKLClass 2

Hemostatic Metal Clip For The Gi Tract

FDA adverse event reports and recall history for hemostatic metal clip for the gi tract (product code PKL), from the public MAUDE/AEMS corpus and the live FDA recall database.

Total reports
7,065
Deaths
18
Injuries
618
Malfunctions
6,426

Reports per year

12
14
15
16
17
18
19
20
21
22
23
24
25
26

Source: FDA adverse event reports by year received, 20122026. Reporting paused during the 2026 MAUDE-to-AEMS transition and has since resumed.

Recent recalls

  • 2026-04-13Open, ClassifiedProcess control

    Instinct Plus Endoscopic Clipping Device REF: G58010 Rx Only, Sterile EO

    Due to increased in complaints their is the potential for endoscopic clipping device to malfunction.

  • 2023-06-21Open, ClassifiedUnder Investigation by firm

    Single Use Repositionable Clip QUICKCLIP PRO 165 CM 10/BX- Intended for hemostatic metal clip for the GI Tract Catalog Number: HX-202LR.A

    Complaints on clip deployment occurring during clinical procedures, leading to hazardous situations where clips fail to achieve their expected function or deploy incorrectly, potentially causing harms to patients. The complaints include reports that: 1) the clip arm does not open when the user pushes the slider, 2) the clip arm does not close when the user pulls the slider and 3) the clip detaches from the target tissue earlier than expected, after being deployed in a procedure.

  • 2023-06-21Open, ClassifiedUnder Investigation by firm

    Single Use Repositionable Clip QUICKCLIP PRO 165CM 5/BX- Intended for hemostatic metal clip for the GI Tract Catalog Number: HX-202LR.B

    Complaints on clip deployment occurring during clinical procedures, leading to hazardous situations where clips fail to achieve their expected function or deploy incorrectly, potentially causing harms to patients. The complaints include reports that: 1) the clip arm does not open when the user pushes the slider, 2) the clip arm does not close when the user pulls the slider and 3) the clip detaches from the target tissue earlier than expected, after being deployed in a procedure.

  • 2023-06-21Open, ClassifiedUnder Investigation by firm

    Single Use Repositionable Clip QUICKCLIP PRO 230 CM 10/BX- Intended for hemostatic metal clip for the GI Tract Catalog Number: HX-202UR.A

    Complaints on clip deployment occurring during clinical procedures, leading to hazardous situations where clips fail to achieve their expected function or deploy incorrectly, potentially causing harms to patients. The complaints include reports that: 1) the clip arm does not open when the user pushes the slider, 2) the clip arm does not close when the user pulls the slider and 3) the clip detaches from the target tissue earlier than expected, after being deployed in a procedure.

  • 2023-06-21Open, ClassifiedUnder Investigation by firm

    Single Use Repositionable Clip QUICKCLIP PRO 230CM 5/BX- Intended for hemostatic metal clip for the GI Tract Catalog Number: HX-202UR.B

    Complaints on clip deployment occurring during clinical procedures, leading to hazardous situations where clips fail to achieve their expected function or deploy incorrectly, potentially causing harms to patients. The complaints include reports that: 1) the clip arm does not open when the user pushes the slider, 2) the clip arm does not close when the user pulls the slider and 3) the clip detaches from the target tissue earlier than expected, after being deployed in a procedure.

  • TerminatedDevice Design

    US Endoscopy Padlock Clip defect closure system Ref: C910001 The Padlock Clip is indicated for use in flexible endoscopy and for the compression of tissue in the gastrointestinal tract, for hemostasis or for treating lesions of the wall of the gastrointestinal organs.

    Potential esophageal laceration during a patient procedure

Monitor hemostatic metal clip for the gi tract automatically