Skip to main content
OVDClass 2

Intervertebral Fusion Device With Integrated Fixation, Lumbar

FDA adverse event reports and recall history for intervertebral fusion device with integrated fixation, lumbar (product code OVD), from the public MAUDE corpus and the live FDA recall database.

Total reports
1,271
Deaths
11
Injuries
792
Malfunctions
462

Reports per year

12
13
14
15
16
17
18
19
20
21
22
23
24
25
26

Source: FDA adverse event reports by year received, 2012–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.

Recent recalls

  • 2026-05-20Open, ClassifiedProcess design

    OneLIF Interbody Cage (REF/Description): 010-107-1007/Small 7 Degree 10mm X 7mm, 010-107-1007-2/Small 7 Degree 10mm X 7mm - Oblique, 010-107-1209/Small 7 Degree 12mm X 9mm, 010-107-1411/Small 7 Degree 14mm X 11mm, 010-107-1613/Small 7 Degree 16mm X 13mm, 010-113-1006/Small, 13 Degree 10mm X 6mm, 010-113-1006-2/Small, 13 Degree, 10mm X 6mm - Oblique, 010-113-1208/Small 13 Degree 12mm X 8mm, 010-113-1410/Small 13 Degree 14mm X 10mm, 010-113-1612/Small 13 Degree 16mm X 12mm, 010-119-1407/19 Degree 14mm X 7mm, 010-119-1609/Small 19 Degree 16mm X 9mm, 010-119-1811/Small 19 Degree 18mm X 11mm, 010-125-1606/Small 25 Degree 16mm X 6mm, 010-125-1808/Small 25 Degree 18mm X 8mm, 010-207-1007/Medium 7 Degree 10mm X 7mm, 010-207-1007-2/Medium 7 Degree 10mm X 7mm - Oblique, 010-207-1209/Medium 7 Degree 12mm X 9mm, 010-207-1411/Medium 7 Degree 14mm X 11mm, 010-207-1613/Medium 7 Degree 16mm X 13mm, 010-213-1206/Medium 13 Degree 12mm X 6mm, 010-213-1408/Medium 13 Degree 14mm X 8mm, 010-213-1610/Medium 13 Degree 16mm X 10mm, 010-213-1812/Medium 13 Degree 18mm X 12mm, 010-219-1406/Medium 19 Degree 14mm X 6mm, 010-219-1608/Medium 19 Degree 16mm X 8mm, 010-219-1810/Medium 19 Degree 18mm X 10mm, 010-225-1807/Medium 25 Degree 18mm X 7mm, 010-225-2009/Medium 25 Degree 20mm X 9mm, 010-307-1006/Large 7 Degree,10mm X 6mm, 010-307-1006-2/Large 7 Degree 10mm X 6mm - Oblique, 010-307-1208/ Large 7 Degree 12mm X 8mm, 010-307-1410/Large 7 Degree 14mm X 10mm, 010-307-1612/Large 7 Degree 16mm X 12mm, 010-313-1205/Large 13 Degree 12mm X 5mm, 010-313-1407/Large 13 Degree 14mm X 7mm, 010-313-1609/Large 13 Degree 16mm X 9mm, 010-313-1811/Large 13 Degree 18mm X 11mm, 010-313-2013/Large 13 Degree 20mm X 13mm, 010-319-1606/Large 19 Degree 16mm X 6mm, 010-319-1808/ Large 19 Degree 18mm X 8mm, 010-319-2010/ Large 19 Degree 20mm X 10mm, 010-325-2007/Large 25 Degree 20mm X 7mm, 010-325-2209/ Large 25 Degree 22mm X 9mm, Implant Kits containing OneLIF Interbody Cages, Name/Kit Family: Kit-OneLIF-10 mm (716)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (717)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (718)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (719)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (722)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (723)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (724)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (733)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (734)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (737)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (738)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (746)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (751)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (761)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (762)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (787)/Kit-OneLIF-10 mm, Kit-OneLIF- IMPOneLIF Implant Kit (701)/Kit-OneLIF-IMP - OneLIF Implant Kit, Kit-OneLIF-IMP - OneLIF Implant Kit (702)/Kit-OneLIF-IMP - OneLIF Implant Kit, Kit-OneLIF-IMP - OneLIF Implant Kit (703) NL/Kit-OneLIF-IMP - OneLIF Implant Kit, Kit-OneLIF-IMP - OneLIF Implant Kit (704)/Kit-OneLIF-IMP - OneLIF Implant Kit, Kit-OneLIF-IMP - OneLIF Implant Kit (705)/Kit-OneLIF-IMP - OneLIF Implant Kit, Kit-OneLIF-IMP - OneLIF Implant Kit (706)/Kit-OneLIF-IMP - OneLIF Implant Kit, Kit-OneLIF-IMP - OneLIF Implant Kit (708)/Kit-OneLIF-IMP - OneLIF Implant Kit, Kit-OneLIF-IMP - OneLIF Implant Kit (709)/Kit-OneLIF-IMP - OneLIF Implant Kit,

    Intervertebral body fusion system straight inserter failed to properly attach to affected cages, which may result in an inability to securely engage the implant with the inserter, which could lead to procedural delay and/or the need to select an alternative implant.

  • 2025-12-11Open, ClassifiedDevice Design

    Brand Name: Pillar" SA PTC Spacer System Product Name: Pillar" SA PTC Spacer System Model/Catalog Number: 39-9012SP 33W X 28L X 12.5H, 7DEG PTC IMPLANT; 39-9014SP 33W X 28L X 14H, 7DEG PTC IMPLANT; 39-9016SP 33W X 28L X 16H, 7DEG PTC IMPLANT; 39-9018SP 33W X 28L X 18H, 7DEG PTC IMPLANT; 39-9212SP 33W X 28L X 12.5H, 12DEG PTC IMPLANT; 39-9214SP 33W X 28L X 14H, 12DEG PTC IMPLANT; 39-9216SP 33W X 28L X 16H, 12DEG PTC IMPLANT; 39-9218SP 33W X 28L X 18H, 12DEG PTC IMPLANT; 39-2012SP 37W X 28L X 12.5H, 7DEG PTC IMPLANT; 39-2014SP 37W X 28L X 14H, 7DEG PTC IMPLANT; 39-2016SP 37W X 28L X 16H, 7DEG PTC IMPLANT; 39-2018SP 37W X 28L X 18H, 7DEG PTC IMPLANT; 39-2212SP 37W X 28L X 12.5H, 12DEG PTC IMPLANT; 39-2214SP 37W X 28L X 14H, 12DEG PTC IMPLANT; 39-2216SP 37W X 28L X 16H, 12DEG PTC IMPLANT; 39-2218SP 37W X 28L X 18H, 12DEG PTC IMPLANT;

    Labeling contains claims that are not consistently present.

  • 2025-09-11Open, ClassifiedDevice Design

    TiBrid Stand Alone Intervertebral Body Fusion Device, SA Angled Punch Awl, Sizes: 30 degree, 45 degree and 60 degree.

    Failure of fusion system instruments in the field.

  • 2025-09-11Open, ClassifiedDevice Design

    TiBrid Stand Alone Intervertebral Body Fusion Device, TiBrid SA Angled Screwdrivers, Sizes 30 degree, 45 degree, and 60 degree angles.

    Failure of fusion system instruments in the field.

  • 2025-03-31Open, ClassifiedDevice Design

    Monterey AL Implant Inserter; 10/12mm; Catalog 48019100.

    Potential for the gold unlock button to separate from the inserter.

  • 2025-03-31Open, ClassifiedDevice Design

    Monterey AL Implant Inserter; 14/16mm; Catalog 48019120.

    Potential for the gold unlock button to separate from the inserter.

  • 2025-03-31Open, ClassifiedDevice Design

    Monterey AL Implant Inserter; 18/20mm; Catalog 48019130.

    Potential for the gold unlock button to separate from the inserter.

  • 2025-03-31Open, ClassifiedDevice Design

    Monterey AL Implant Inserter; 22mm; Catalog 48019140.

    Potential for the gold unlock button to separate from the inserter.

  • 2025-02-19Open, ClassifiedLabeling Change Control

    Irix-A Integrated Lumbar Fusion System, Model Number X080-423011-08PC (Spacer Assy, 42 X 30 X 8, 11mm, PC (STERILE)) and Model Number X080-382819-08PC (Spacer ASSY, 38 X 28 X 8, 19mm, PC, (STERILE))

    Certain implants in the Irix-A Lumbar Fusion System were distributed while labeled with an incorrect expiration date.

  • 2024-02-23Open, ClassifiedDevice Design

    IdentiTi ALIF Standalone Interbody System: Part Number/Description 132-10-85-250 / IdentiTi ALIF SA Graft Bolt, 08.5 x 25 mm 132-10-85-300 / IdentiTi ALIF SA Graft Bolt, 08.5 x 30 mm 132-10-85-350 / IdentiTi ALIF SA Graft Bolt, 08.5 x 35 mm 132-10-90-250 / IdentiTi ALIF SA Graft Bolt, 09 x 25 mm 132-10-90-300 / IdentiTi ALIF SA Graft Bolt, 09 x 30 mm 132-10-90-350 / IdentiTi ALIF SA Graft Bolt, 09 x 35 mm 232-10-85-250 / IdentiTi ALIF SA Graft Bolt, 08.5 x 25 mm 232-10-85-300 / IdentiTi ALIF SA Graft Bolt, 08.5 x 30 mm 232-10-85-350 / IdentiTi ALIF SA Graft Bolt, 08.5 x 35 mm 232-11-90-250 / IdentiTi ALIF SA Cannulated Graft Bolt, 09 x 25 mm 232-11-90-300 / IdentiTi ALIF SA Cannulated Graft Bolt, 09 x 30 mm 232-11-90-350 / IdentiTi ALIF SA Cannulated Graft Bolt, 09 x 35 mm The IdentiTi ALIF Standalone Interbody System is an integrated intervertebral body fusion device for use in anterior lumbar interbody fusion (ALIF) procedures.

    Due to reports of intraoperative graft bolt implantation breakages.

Monitor intervertebral fusion device with integrated fixation, lumbar automatically