Pacemaker/Icd/Crt Non-Implanted Components
FDA adverse event reports and recall history for pacemaker/icd/crt non-implanted components (product code OSR), from the public MAUDE corpus and the live FDA recall database.
- Total reports
- 63,843
- Deaths
- 2
- Injuries
- 98
- Malfunctions
- 63,737
Reports per year
Source: FDA adverse event reports by year received, 2013–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.
Recent recalls
- TerminatedUnknown/Undetermined by firm
Merlin@Home Software Model EX2000 v8.2.2 for Merlin@Home Transmitter Models EX1150, EX1150W, EX1100, and EX1100W. The Merlin@home Transmitters are a component of St. Jude Medicals (SJM) Remote Care System. These Merlin@home transmitters are used to retrieve information from the SJM implantable medical device (IMD) using RF telemetry. The data is then sent back to the Merlin.net Patient Care Network through a built-in modem or, alternatively, through a USB cellular adapter or a USB based Broadband Kit. The transmitted data enables clinicians to remotely monitor a patients IMD using a secure, encrypted, password-protected web site, thus reducing the number of in-clinic visits a patient needs to make to their physician. The transmitter monitors the patients device daily between scheduled follow-ups and can alert a clinician if it detects anything out of the norm.
New pacemaker firmware was developed to further mitigate the risk of unauthorized access to our pacemakers that utilize radio frequency (RF) communications.
- Open, ClassifiedNonconforming Material/Component
Programmers & Remote Monitoring Software Apps: 2090 CareLink Programmer
There is a potential for Medtronic programmer and remote monitoring software applications to display an inaccurate remaining longevity estimate for a subset of implanted cardiac device models. This issue does not impact device functionality.
- Open, ClassifiedNonconforming Material/Component
Programmers & Remote Monitoring Software Apps: 29901 CareLink Encore Programmer
There is a potential for Medtronic programmer and remote monitoring software applications to display an inaccurate remaining longevity estimate for a subset of implanted cardiac device models. This issue does not impact device functionality.
- Open, ClassifiedNonconforming Material/Component
Programmers & Remote Monitoring Software Apps: CareLink SmartSync Device Manager
There is a potential for Medtronic programmer and remote monitoring software applications to display an inaccurate remaining longevity estimate for a subset of implanted cardiac device models. This issue does not impact device functionality.
- Open, ClassifiedSoftware design
Medtronic CareLink SmartSync Device Manager application software (D00U005) used by the following Medtronic Cobalt and Crome implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds): Cobalt XT VR: DVPA2D1, DVPA2D4; Cobalt VR: DVPB3D1, DVPB3D4; Cobalt XT DR: DDPA2D1, DPA2D4; Cobalt DR: DDPB3D1, DDPB3D4; Cobalt XT HF: DTPA2D4, DTPA2D1; Cobalt XT HF; Quad: DTPA2QQ, DTPA2Q1; Cobalt HF: DTPB2D4, DTPB2D1; Cobalt HF Quad: DTPB2QQ, DTPB2Q1. Crome VR: DVPC3D1, DVPC3D4; Crome DR: DDPC3D1, DDPC3D4; Crome HF: DTPC2D4, DTPC2D1; Crome HF Quad: DTPC2QQ, DTPC2Q1.
The processing of collected episode data may fail due to a software error.
- Open, ClassifiedSoftware design
Medtronic CareLink 2090 Programmer
Vulnerabilities have been identified in the SDN download process that may allow an individual with malicious intent to update the programmers with non-Medtronic software during an SDN download.
- Open, ClassifiedSoftware design
Medtronic CareLink Encore 29901 Programmer
Vulnerabilities have been identified in the SDN download process that may allow an individual with malicious intent to update the programmers with non-Medtronic software during an SDN download.
- TerminatedSoftware design
MEDTRONIC CARELINK 2090 programmer
There is an error in how the programmer calculates and displays the remaining longevity value during a period of time (up to 2 years) prior to the device reaching its Recommended Replacement Time (RRT).
- TerminatedSoftware design
MEDTRONIC CARELINK 29901 Programmer for Medtronic and Vitaron Devices
The Medtronic CareLink Encore 29901 programmer system is comprised of prescription devices indicated for use in the interrogation and programming of implantable medical devices. Prior to use, refer to the Programmer Reference Guide as well as the appropriate programmer software and implantable device technical manuals for more information related to specific implantable device models. Programming should be attempted only by appropriately trained personnel after careful study of the technical manual for the implantable device and after careful determination of appropriate parameter values based on the patient's condition and pacing system used. The Medtronic CareLink Encore 29901 programmer must be used only for programming implantable devices manufactured by Medtronic or Vitatron.
- TerminatedDevice Design
Medtronic Power Supply 26907, Medtronic CareLink Encore(TM) 29901 Programmer
There is a potential for an electrical short circuit to develop in a subset of 479 Model 26907 power supply cable plugs that connect to the Medtronic Model 29901 CareLink Encore programmer. A failure mode of the power supply has been found where an electrical short occurs on the low voltage direct current connector and results in the connector heating up, generating smoke, melting and in some cases disconnecting from the programmer.
Monitor pacemaker/icd/crt non-implanted components automatically
DeviceWatch watches FDA adverse events and recalls for the products you care about, summarizes clinical narratives with AI and alerts you when a safety signal emerges.