Real Time Nucleic Acid Amplification System
FDA adverse event reports and recall history for real time nucleic acid amplification system (product code OOI), from the public MAUDE/AEMS corpus and the live FDA recall database.
- Total reports
- 2,702
- Deaths
- 0
- Injuries
- 23
- Malfunctions
- 2,673
Reports per year
Source: FDA adverse event reports by year received, 2011–2026. Reporting paused during the 2026 MAUDE-to-AEMS transition and has since resumed.
Recent recalls
- 2025-04-24Open, ClassifiedDevice Design
The Alinity m System is designed to automate the steps for nucleic acid testing. Model Numbers included in this recall are 08N53-001, 08N53-002, and 08N53-032 (Refurbished).
The design of the Alinity m System Solutions drawer enclosure may allow liquid (liquid waste or system solutions) originating in the System Solutions drawer to flow beyond the footprint of the instrument and into the walking-path of the user.
- 2025-01-10Open, ClassifiedEnvironmental control
Idylla INSTRUMENT, IVD, REF P0010. Intended for the detection of nucleic acid targets in various types of human samples, using disposable Test-specific ldylla Cartridges.
False-positive MSI-H results generated by the Idylla MSI Test.
- 2024-04-19Open, ClassifiedUnder Investigation by firm
Alinity m System, Part Number: 08N53-002
The Alinity m System camera detects whether tubes are "capped." If the camera detects a cap and the retention bar is not present on the sample rack, the system will generate the message 'Error - 6013: Has cap, no retention bar." Abbott has identified an issue concerning the use of certain third-party Universal Transport Medium (UTM) tubes on the Alinity m System. This issue has been observed with Copan UTM-RT 3 mL Without Beads Transport & Preservation Medium for Viral Molecular Diagnostics Testing, Part Number (PN) 3C047N. However, it is possible that other third party manufactured tubes could experience the same issue.
- 2024-02-26Open, ClassifiedProcess control
Wash station: M1000 Wash Station Assembly & DITI Slide, REF 04J72-28* (*This part number is a spare to the M2000sp.)
Products that were intended for export only which were not approved or registered in accordance with U.S. specifications were shipped to U.S. customers.
- 2024-01-23Open, ClassifiedUnder Investigation by firm
PANTHER FUSION System REF 9121010000 The Panther System is an integrated nucleic acid testing system which fully automates all steps necessary to perform Aptima assays from sample processing through amplification, detection, and data reduction. The Panther Fusion System is an integrated, fully automated, in vitro diagnostic (IVD) clinical multiplex test system. It is capable of performing nucleic acid based tests from sample processing through amplification, detection, and data reduction.
Due to potential false positive Flu B results when a sample was also SARS-CoV-2 positive.
- 2023-07-18Open, ClassifiedUnder Investigation by firm
Simplexa Direct Amplification Disc Kit, Rx Only, IVD The Simplexa Direct Amplification Disc Kit, REF: MOL1455 is 1 box containing 3 Direct Amplification Discs (DADs). Each box contains discs individually pouch-sealed and labeled with REF: MOL1452
There is a potential for the direct amplification disc to malfunction which may result in spillage of liquid from the disc.
- 2023-04-06Open, ClassifiedUnder Investigation by firm
Abbott Alinity m System, REF 08N53-002, containing Amplification Detetction Unit (ADU) #4.
The instrument was found to be installed without plunger assemblies in the clamp bar for an Amplification Detection Unit. After correction of the issue, the instrument was returned for use without calibration. This would result in a potential for delay of results and incorrect results.
- 2023-01-20Open, ClassifiedUnder Investigation by firm
Alinity m System. The Alinity m System is designed to automate the steps for nucleic acid testing, which includes sample processing, amplification, detection, and data reduction.
The firm identified the following three potential issues which may have impacted ADUs installed on distributed Alinity m Systems. 1. Systems which had their ADUs serviced may have an incorrect board value (calibration data file) referenced causing a misconfiguration of the calibration. 2. Inability of Maintenance and Diagnostics 2300 (Amp Detect Optical Calibrations) to recognize error handling and thermal hold errors during the calibration process, potentially leading to a misconfiguration of the calibration. 3. Under specific conditions during the ADU manufacturing process, an incorrect calibration may potentially have been configured.
- 2022-11-18Open, ClassifiedUnder Investigation by firm
Alinity m System, REF 08N53-002, For In Vitro Diagnostic Use
Abbott has identified potential performance issues for the Alinity m System software version 1.6.5: 1) In a unique scenario, the waste chute flapper was found to not open when the Systems Solution drawer was closed and locked. It was found that the software will cause the flapper door to stay in the closed position when the drawer is closed and locked if the flapper had previously been moved out of position. 2) Sample preparation drawer #1 barcode information is used instead of sample prep drawer #2 when the scanned data is not sent to the System Control Center before the next bottle barcode in sample prep drawer#2 is scanned. 3) Under a specific condition, while the system is processing tests and a new test request is made, when the level of the bulk solution is too low to process a test, the software would error stop the system and try to complete all in-process tests. An internal counter is reset which can cause reaction vessels (RVs) already present in an Amplification Detection Unit (or amp detect) to remain there and not be moved to waste so that more RVs can be placed into the amp detect for processing. 4) 4 to 6 replicates of the same auto calibrator orders can be run with 2 different sets of calibrator materials on the same rack. It was discovered during internal testing, when running 4 to 6 replicates of a calibrator, it is possible for the user to use 2 separate lots of material which is not detected by the System Control Center (SCC). If this were to occur, the calibration curve would be created using 2 separate material lots. In normal use scenario, this curve would be made using only 1 lot of material.
- 2022-05-13Open, ClassifiedUnder Investigation by firm
Verigene Processor SP, Models 10-0000-07/10-0000-07R a component of the Verigene system
Heater within the Verigene System may be outside of established temperature limits for assays which could result in false negative results.
Monitor real time nucleic acid amplification system automatically