Skip to main content
OHRClass 2

Catheter Care Tray

FDA adverse event reports and recall history for catheter care tray (product code OHR), from the public MAUDE corpus and the live FDA recall database.

Total reports
1,270
Deaths
1
Injuries
387
Malfunctions
881

Reports per year

14
15
16
17
18
19
20
21
22
23
24
25
26

Source: FDA adverse event reports by year received, 2014–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.

Recent recalls

  • 2026-06-17Open, ClassifiedNonconforming Material/Component

    MEDLINE Urethral Catheterization Tray, Red Rubber Latex, Medline Product Number/SKU (REF), French sizes: 1. DYNC1816, 15 Fr; 2. DYND10350, 15 Fr.

    During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.

  • 2026-06-17Open, ClassifiedNonconforming Material/Component

    MEDLINE Pre-connected Urethral Catheterization Tray and Bag, Red Rubber Latex, 15 Fr, Medline Product Number/SKU (REF) DYND10407

    During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.

  • 2026-06-17Open, ClassifiedNonconforming Material/Component

    MEDLINE Urethral Catheterization Tray and Bag, Red Rubber Latex, 15 Fr, Medline Product Number/SKU (REF) DYND18350

    During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.

  • 2026-06-17Open, ClassifiedNonconforming Material/Component

    MEDLINE Intermittent Latex Catheter Coude Pre-Connected to Collection Bag, 14 Fr, 1000mL, Medline Product Number/SKU (REF) URO4RR14C

    During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.

  • 2026-06-17Open, ClassifiedNonconforming Material/Component

    MEDLINE Intermittent Latex Catheter Coude Pre-Connected to Collection Bag, 16 Fr, 1000mL, Medline Product Number/SKU (REF) URO4RR16C

    During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.

  • 2026-06-17Open, ClassifiedNonconforming Material/Component

    MEDLINE Medical Procedure Kits labeled as: TONSIL PACK CATH, Medline Kit Number/SKU DYNJ65199A

    During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.

  • 2025-11-14Open, ClassifiedUnder Investigation by firm

    MEDLINE convenience kits labeled as: PORT INSERTION PACK, REF DYNJ81597

    Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing SafeAir Smoke Evacuation Pencils. Stryker Instruments has expanded the scope of their recall of SafeAir Smoke Evacuation Pencils previously initiated in October 2024, adding an additional 47 lots to the scope of the recall. The initial recall was due to the potential for product to activate without manual input when plugged into a power source or remain active after buttons are released.

  • 2024-12-05Open, ClassifiedUnder Investigation by firm

    MEDLINE convenience kits labeled as: PORT INSERTION PACK, REF DYNJ81597

    Medline is recalling medical convenience manufactured using certain lots of SafeAir Smoke Evacuation Pencil which were subsequently recalled by Stryker.

  • 2024-10-23Open, ClassifiedUnder Investigation by firm

    Medline CENTRAL CATHETER INSERTION TRA, REF CVI680C; medical procedure convenience kit

    It has been determined that certain Convenience kits that were packaged in breather pouch lot 323080002 may result in the potential of using non-sterile product if the vendor seal of the pouch is opened/unsealed.

  • 2024-09-06Open, ClassifiedUnder Investigation by firm

    Medline Convenience kits labeled as: 1) CVC INSERTION PACK, Pack Number CVI4720A ; 2) CVC INSERTION PACK, Pack Number CVI4720AH; 3) CATHETER INSERTION KIT CLABSI , Pack Number CVI4880; 4) CHEST INSERTION PACK-LF , Pack Number DYNJ17634D ; 5) URETEROSCOPY, Pack Number DYNJ45882A ; 6) TONSIL PACK CATH, Pack Number DYNJ65199A ; 7) CATH PLACEMENT RIB FRACTURE , Pack Number DYNJ66475B ; 8) UROLOGY ROBOTIC PACK, Pack Number DYNJ80211C ; 9) UROLOGY ROBOTIC PACK, Pack Number DYNJ80211CH; 10) UROLOGY PK, Pack Number DYNJ80252C ; 11) PORT INSERTION PACK , Pack Number DYNJ81597

    Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that may pose a risk to patient health.

Monitor catheter care tray automatically