Skip to main content
ODGClass 2

Endoscopic Ultrasound System, Gastroenterology-Urology

FDA adverse event reports and recall history for endoscopic ultrasound system, gastroenterology-urology (product code ODG), from the public MAUDE corpus and the live FDA recall database.

Total reports
17,739
Deaths
32
Injuries
318
Malfunctions
17,384

Reports per year

12
13
14
15
16
17
18
19
20
21
22
23
24
25
26

Source: FDA adverse event reports by year received, 2012–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.

Recent recalls

  • 2016-07-23TerminatedDevice Design

    Ultrasound Gastroscope Product Usage: Intended to provide optical visualization of, ultrasonic Visualization of, and therapeutic access to the Upper Gastrointestinal Tract including but not restricted to, the organs; tissues; and subsystems: Esophagus, Stomach, Duodenum, Small Bowel and underlying areas. The instrument is introduced via per orally when indications consistent with the requirement for procedure are observed in adult and pediatric patient populations.

    Pentax Medical Video Processor EPK-i5010 has been shipped with a reusable Pentax Medical water bottle assembly (OS-H4) which has not been cleared for use with this devices in the United States.

  • 2015-09-28TerminatedDevice Design

    EVIS EXERA II Gastrovideoscope OLYMPUS GF type UCT180; designed for endoscopic real-time ultrasound imaging, ultrasound guided needle aspiration and other endoscopic procedures within the upper gastrointestinal tract and surrounding organs.

    Insufficient information for specific cleaning accessories used for the cleaning and reprocessing of the GF-UCT180 ultrasound endoscope

  • TerminatedDevice Design

    Olympus Ultrasound Endoscope, Model number :GF-UM20 - Product Usage: used with other supporting equipment for endoscopic real-time ultrasound imaging and endoscopic surgery within the gastrointestinal tract.

    A new inspection step was added to the cleaning and disinfection section of the Instructions For Use (IFU) for this device in order to reduce the risk of causing patient infections.

  • TerminatedDevice Design

    Olympus Ultrasound Endoscope, Model number: GF-UM160 - Product Usage: used for diagnostic ultrasound imaging or fluid flow analysis of the human body.

    A new inspection step was added to the cleaning and disinfection section of the Instructions For Use (IFU) for this device in order to reduce the risk of causing patient infections.

  • TerminatedDevice Design

    Olympus Ultrasound Endoscope, Model number: GF-UC160P-OL5 - Product Usage: used for diagnostic ultrasound imaging or fluid flow analysis of the human body.

    A new inspection step was added to the cleaning and disinfection section of the Instructions For Use (IFU) for this device in order to reduce the risk of causing patient infections.

  • TerminatedDevice Design

    Olympus Ultrasound Endoscope, Model number: GF-UCT160-OL5 - Product Usage: used for diagnostic ultrasound imaging or fluid flow analysis of the human body.

    A new inspection step was added to the cleaning and disinfection section of the Instructions For Use (IFU) for this device in order to reduce the risk of causing patient infections.

  • TerminatedDevice Design

    Trident Endoscopic Ultrasonic Biopsy Needle (FNB)

    The combination of the firmness of the mass, the potential endoscope and elevator position combined with the force required to penetrate the mass may contribute to the breakage of the needle in the human body, which could cause injury.

  • TerminatedDevice Design

    Trident Endoscopic Ultrasonic Aspiration Needle (FNA)

    The combination of the firmness of the mass, the potential endoscope and elevator position combined with the force required to penetrate the mass may contribute to the breakage of the needle in the human body, which could cause injury.

  • TerminatedError in labeling

    Ultrasound Video Gastroscope, Model EG-3670URK, The product is intended to be used in combination with the ultrasound scanning unit made by Hitachi Medical Corporation, and the PENTAX Medical video processor to provide images for observation, diagnosis, imaging, or therapy of the upper gastrointestinal tract from esophagus through duodenum.

    Error in reprocessing instructions in Revisions 13 and 14 of IFU Z845, used with EG-3670URK and EG-3870UTK ultrasound gastroscopes. In the disinfection step, the IFU has a warning indicating the cleaning detergent solution should remain in contact with the all internal channels and external endoscope surfaces. The disinfecting solution, not the cleaning detergent, should be used for this step.

  • TerminatedError in labeling

    Ultrasound Video Gastroscope, Model EG-3870UTK. The product is intended to be used in combination with the ultrasound scanning unit made by Hitachi Medical Corporation, and the PENTAX Medical video processor to provide images for observation, diagnosis, imaging, or therapy of the upper gastrointestinal tract from esophagus through duodenum.

    Error in reprocessing instructions in Revisions 13 and 14 of IFU Z845, used with EG-3670URK and EG-3870UTK ultrasound gastroscopes. In the disinfection step, the IFU has a warning indicating the cleaning detergent solution should remain in contact with the all internal channels and external endoscope surfaces. The disinfecting solution, not the cleaning detergent, should be used for this step.

Monitor endoscopic ultrasound system, gastroenterology-urology automatically