Implant, Endosseous, Orthodontic
FDA adverse event reports and recall history for implant, endosseous, orthodontic (product code OAT), from the public MAUDE/AEMS corpus and the live FDA recall database.
- Total reports
- 423
- Deaths
- 0
- Injuries
- 407
- Malfunctions
- 15
Reports per year
Source: FDA adverse event reports by year received, 2007–2026. Reporting paused during the 2026 MAUDE-to-AEMS transition and has since resumed.
Recent recalls
- 2013-09-12TerminatedDevice Design
Brand Name: Vector TAS 8mm Screw. The Device Listing Number is E670535. Vector T AS 8mm Screw is intended to provide a fixed anchorage point for attachment of orthodontic appliances to facilitate the orthodontic movement of teeth. It is used temporarily and removed after orthodontic treatment has been completed. It is surgically placed in the bone of the upper or lower jaw arches to provide support for orthodontic devices.
Ormco Corporation is voluntarily recalling one lot of 8 MM Vector TAS Screws because some screws of this lot were machined undersized at the top of the threads may potentially cause a weakening of the screw. Use of the undersized product may result in breakage of the screw during insertion or removal when torqued, possibly requiring surgical removal of any remaining pieces.
- 2007-12-04TerminatedDevice Design
Straumann Palatal Implant , 04.1 L 4.2 mm, Article Number: 042.335S, Straumann USA., LLC, Andover, Massachusetts, 01810-1008
Sticking Transfer Piece: The Transfer Piece which accompanies the Palatal Implant may get stuck during implant requiring removal by using mechanical force.
- 2007-12-04TerminatedDevice Design
Straumann Palatal Implant , 4.8 L 4.2 mm, Article Number: 042.336S, Straumann USA., LLC, Andover, Massachusetts, 01810-1008
Sticking Transfer Piece: The Transfer Piece which accompanies the Palatal Implant may get stuck during implant requiring removal by using mechanical force.
Monitor implant, endosseous, orthodontic automatically