Low Energy Direct Current Thermal Ablation System
FDA adverse event reports and recall history for low energy direct current thermal ablation system (product code OAB), from the public MAUDE corpus and the live FDA recall database.
- Total reports
- 233
- Deaths
- 9
- Injuries
- 143
- Malfunctions
- 78
Reports per year
Source: FDA adverse event reports by year received, 2012–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.
Recent recalls
- 2023-04-12Open, ClassifiedUnder Investigation by firm
NanoKnife 5-Probe Procedure Pack 15 cm - Activation. Indicated for the surgical ablation of soft tissue. Catalog Number: 20400111 UPN: H787204001110
Not programmed in accordance with specification. The programming affects the RFID function and does not allow the NanoKnife probes to be recognized by the NanoKnife generator resulting in a delay in procedure
- 2012-01-27TerminatedLabeling False and Misleading
NanoKnife Irreversible Electroporation System. System consists of multiple components: Generator, footswitch, disposable electrode. System is manufactured by AngioDynamics, Inc. --- Catalog #20300101 Intended use: the generator system with six outputs is indicated for surgical ablation of soft tissue.
Angiodynamics was marketing the NanoKnife System with the ablation zone estimator feature in the US without FDA clearance.
- 2011-09-23TerminatedComponent design/selection
ANGIODYNAMICS NANOKNIFE IRE GENERATOR, Catalog/REF No. 20300101, ANGIODYNAMICS, 603 Queensbury Ave., Queensbury, NY 12804 Intended to remove tissue and control bleeding by the use of high- frequency electrical current (surgical ablation of soft tissue).
Some units of Nanoknife have the potential to deliver a shortened pulse width without indicating an error. The affected units contain a "GOMe" circuit board which may cause the generator to drift from the calibration points set at the time of manufacture.
- TerminatedSoftware Manufacturing/Software Deployment
AngioDynamics NanoKnife Disposable Single Electrode Probes, Irreversible Electroporation (IRE) Device, 15 cm IRE Single Electrode RFID Activation, PG. Catalog No. 20400103
Programming issue affected RFID function of a single lot and did not allow the NanoKnife probes to be recognized by the NanoKnife generator.