Skip to main content
NWXClass 3

Catheter, Percutaneous Transluminal Coronary Angioplasty (Ptca), Cutting/Scoring

FDA adverse event reports and recall history for catheter, percutaneous transluminal coronary angioplasty (ptca), cutting/scoring (product code NWX), from the public MAUDE corpus and the live FDA recall database.

Total reports
7,883
Deaths
53
Injuries
475
Malfunctions
7,354

Reports per year

12
13
14
15
16
17
18
19
20
21
22
23
24
25
26

Source: FDA adverse event reports by year received, 2012–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.

Recent recalls

  • 2010-08-18TerminatedProcess design

    AngioSculpt Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheter, cutting/scoring, Easy Exchange (EX) delivery system, Manufactured by AngioScore, Fremont, CA. Catheter is for use in the treatment of hemodynamically significant coronary artery stenosis, including in-stent restenosis and complex type C lesions, for improving myocardial perfusion. The devices are available in balloon diameters of 2.0-3.5 mm in 0.5 mm increments, and in scoring balloon lengths of 10 to 15 nm. Catheter length is 139 cm and is compatible with 0.014 inch guide wires and 6F guide catheters. The EX catheter is supplied sterile and intended for single use.

    Separations of the PTCA catheter shaft at the guide wire exit port have occurred during treatment of coronary artery stenosis. Fragments of the catheter may become lodged in coronary arteries, which resulting in serious patient outcome and require emergency coronary bypass surgery and medical intervention to remove the retained device fragments.

  • 2008-01-23TerminatedProcess control

    AngioSculpt PTCA Scoring Balloon Catheter, Catalog (REF) Number: 2027-2510, size 2.5mm x 10 mm, AngioScore, Inc., Fremont, CA.

    Mis-labeling: Product labeling on product pouch indicates wrong size, the correct size is on the product carton

  • TerminatedProcess control

    Spectranetics AngioSculpt PTCA Scoring Balloon Catheter, balloon outer diameter 2.5mm, balloon length 15mm, REF 2027-2515, Sterile, Rx. The firm name on the label is Spectranetics Corporation, Fremont, CA.

    The incorrect outer carton box was used for the product.

Monitor catheter, percutaneous transluminal coronary angioplasty (ptca), cutting/scoring automatically

DeviceWatch watches FDA adverse events and recalls for the products you care about, summarizes clinical narratives with AI and alerts you when a safety signal emerges.