Posterior Cervical Screw System
FDA adverse event reports and recall history for posterior cervical screw system (product code NKG), from the public MAUDE corpus and the live FDA recall database.
- Total reports
- 2,233
- Deaths
- 15
- Injuries
- 1,452
- Malfunctions
- 764
Reports per year
Source: FDA adverse event reports by year received, 2015–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.
Recent recalls
- 2021-01-27TerminatedEmployee error
Yukon Polyaxial Screw Size 4.0x26mm (catalog number 7601-04026), component of YUKON Spinal System
The Yukon polyaxial screw used in the Yukon Spinal System was incorrectly laser marked with size 5.0 mm (diameter) & 24 mm (length).
- TerminatedProcess control
YUKON Polyaxial Screws (3.5 X 12 mm, Owning Site: Leesburg USA 3.5 X 14 mm, 3.5 X 32 mm, 3.5 X 34 mm, 3.5 X 38 mm, 3.5 X 40 mm, 4.0 X 16 mm, 4.0 X 18 mm, 4.0 X 20 mm, 4.0 X 22 mm, 4.0 X 24 mm, 4.0 X 26 mm, 4.0 X 28 mm, 4.0 X 30 mm, 5.0 X 22 mm, 5.0 X 28 mm, 5.0 X 30 mm) Catalog Numbers 7601-03512, 7601-03514, 7601-03532, 7601-03534, 7601-03538L, 7601-03540L, 7601-04016L, 7601-04018L, 7601-04020L, 7601-04022L, 7601-04024L, 7601-04026L, 7601-04028L, 7601-04030, 7601-05022, 7601-05028, 7601-05030
Screws had a manufacturing error that could prevent the screw from mating with the screw inserter or set screw as intended. This event was reported to FDA as part of Stryker's retrospective review of K2M product field actions; the action was completed in 2018.
- TerminatedProcess control
Synapse System 3.5mm TI Cancellous Polyaxial Screws, Part No. 04.615.026 for use in spinal surgery
The subject product measures 4.0mm in diameter instead of 3.5mm.
- TerminatedNonconforming Material/Component
Yukon OCT Spinal System Yukon Anti-Torque Alignment Tube - Product Usage: The Anti-Torque Alignment Tube (the Alignment Tube ) provides counter rotation to reduce the torque applied to the anatomy when the size 15 Torque limiting shaft (the shaft is used to tighten the Yukon Set Screw to the specified torque.
Shafts do not pass through the inner cannula of the Anti-Torque Alignment Tube from Lot KTGG due to a manufacturing nonconformance. This could result in a delay of surgery or result in surgical intervention later on.
- TerminatedNonconforming Material/Component
ORTHOFIX, Connector System, SMALL SET SCREW, REF 79-2003
Orthofix is conducting a voluntary recall of the Connector System Large Set Screw (PN: 79-2002) with affected Lot Numbers 001 and 002, and Small Set Screw (PN: 79-2003) with affected Lot Numbers 002, 003 and 004. Through the company's manufacturing controls, size discrepancies were noted with the drive feature for both the large and small sets screws. As a result, it is possible that (1) the set screw drive feature will not accept the Set Screw Driver (PN: 79-1006) or (2) the driver may get wedged in the corners of the drive feature (false bottom) which may lead to a stripped set screw.
- TerminatedNonconforming Material/Component
ORTHOFIX, Connector System, LARGE SET SCREW, REF 79-2002
Orthofix is conducting a voluntary recall of the Connector System Large Set Screw (PN: 79-2002) with affected Lot Numbers 001 and 002, and Small Set Screw (PN: 79-2003) with affected Lot Numbers 002, 003 and 004. Through the company's manufacturing controls, size discrepancies were noted with the drive feature for both the large and small sets screws. As a result, it is possible that (1) the set screw drive feature will not accept the Set Screw Driver (PN: 79-1006) or (2) the driver may get wedged in the corners of the drive feature (false bottom) which may lead to a stripped set screw.
Monitor posterior cervical screw system automatically
DeviceWatch watches FDA adverse events and recalls for the products you care about, summarizes clinical narratives with AI and alerts you when a safety signal emerges.