NIUClass 3

Stent, Superficial Femoral Artery, Drug-Eluting

FDA adverse event reports and recall history for stent, superficial femoral artery, drug-eluting (product code NIU), from the public MAUDE/AEMS corpus and the live FDA recall database.

Total reports
2,836
Deaths
15
Injuries
1,988
Malfunctions
833

Reports per year

06
13
14
15
16
17
18
19
20
21
22
23
24
25
26

Source: FDA adverse event reports by year received, 20062026. Reporting paused during the 2026 MAUDE-to-AEMS transition and has since resumed.

Recent recalls

  • 2013-05-15TerminatedProcess control

    Zilver PTX Drug-Eluting Peripheral Stent - 6 mm X 80mm; 125 cm, Rx, Sterile Product Usage: The Zilver PTX Drug -Eluting Peripheral Stent is a self expanding, small metal, mesh tube (stent) with the outer surface coated with the drug Paclitaxel that can be Implanted in an artery in the thigh (femoropopliteal artery). The Paclitaxel coating helps Preven the artery from narrowing again (restenosis). The Zilver; PTX stent acts as a scaffold to: Hold open a narrowed artery in the thigh (femoropopliteal artery) caused by PAD . improve blood flow to the extremity.

    Cook Medical has received a small number of complaints related to the delivery system for the Zilver PTX Drug-Eluting Peripheral Stent involving fractures of the delivery system inner catheter after stent deployment, and separation of the inner catheter tip section (13 complaints with an occurrence rate of 0.043%) Two adverse events, including one death, occurred in cases where a tip separation

  • 2013-05-15TerminatedProcess control

    Zilver PTX Drug-Eluting Peripheral Stent, - 7 mm X 80mm; 125 cm, Rx, Sterile Product Usage: The Zilver PTX Drug -Eluting Peripheral Stent is a self expanding, small metal, mesh tube (stent) with the outer surface coated with the drug Paclitaxel that can be Implanted in an artery in the thigh (femoropopliteal artery). The Paclitaxel coating helps prevent the artery from narrowing again (restenosis). The Zilver; PTX stent acts as a scaffold to: Hold open a narrowed artery in the thigh (femoropopliteal artery) caused by PAD. Improve blood flow to the extremity.

    Cook Medical has received a small number of complaints related to the delivery system for the Zilver PTX Drug-Eluting Peripheral Stent involving fractures of the delivery system inner catheter after stent deployment, and separation of the inner catheter tip section (13 complaints with an occurrence rate of 0.043%) Two adverse events, including one death, occurred in cases where a tip separation

  • TerminatedReprocessing Controls

    ELUVIA Drug-Eluting Vascular Stent System - Product Usage: indicated for improving luminal diameter in the treatment of symptomatic de-novo or restenotic lesions in the native superficial femoral artery (SFA) and/or proximal popliteal artery with reference vessel diameters (RVD) ranging from 4.0-6.0 mm and total lesion lengths up to 190 mm.

    One batch of the ELUVIATM Drug Eluting Vascular Stent System incorrect placement of some 6x40mm ELUVIA stents into 6x100mm packaging.

Monitor stent, superficial femoral artery, drug-eluting automatically

DeviceWatch watches FDA adverse events and recalls for the products you care about, summarizes clinical narratives with AI and alerts you when a safety signal emerges.