Catheter, Hemodialysis, Triple Lumen, Non-Implanted
FDA adverse event reports and recall history for catheter, hemodialysis, triple lumen, non-implanted (product code NIE), from the public MAUDE corpus and the live FDA recall database.
- Total reports
- 829
- Deaths
- 26
- Injuries
- 136
- Malfunctions
- 663
Reports per year
Source: FDA adverse event reports by year received, 2012–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.
Recent recalls
- 2026-07-02Open, ClassifiedNonconforming Material/Component
5605150 Power-Trialysis Short-Term Straight Dialysis Catheter 13 Fr Triple-Lumen UDI-DI Code: 00801741086953 5605240 Power-Trialysis Short-Term Straight Dialysis Catheter 13 Fr Triple-Lumen UDI-DI Code: 00801741086977 5605300 Power-Trialysis Short-Term Straight Dialysis Catheter 13 Fr Triple-Lumen UDI-DI Code: 00801741086984 5615120 Power-Trialysis Short-Term Curved Extension Dialysis Catheter 13 Fr Triple-Lumen UDI-DI Code: 00801741075087 5615200 Power-Trialysis Short-Term Curved Extension Dialysis Catheter 13 Fr Triple-Lumen UDI-DI Code: 00801741075100 5615200 Power-Trialysis Short-Term Curved Extension Dialysis Catheter 13 Fr Triple-Lumen UDI-DI Code: 00801741075100 5615200 Power-Trialysis Short-Term Curved Extension Dialysis Catheter 13 Fr Triple-Lumen UDI-DI Code: 00801741075100 5655120 Power-Trialysis Short-Term Alphacurve Dialysis Catheter 13 Fr Triple-Lumen UDI-DI Code: 00801741066054 The Power-Trialysis" Short-Term Dialysis Catheter, with a third internal lumen for intravenous therapy, power injection of contrast media, and central venous pressure monitoring, is indicated for use in attaining short-term (less than 30 days) vascular access for hemodialysis, hemoperfusion, and apheresis treatments. The catheter is intended to be inserted in the jugular, femoral, or subclavian vein as required. The maximum recommended infusion rate is 5 mL/sec for power injection of contrast media.
Due to lidocaine ampoules being manufactured under deficient manufacturing practices by manufacturer, lidocaine ampoules are currently being recalled by supplier.
- 2026-07-02Open, ClassifiedNonconforming Material/Component
5855240 Power-Trialysis Slim-Cath Straight Short-Term Dialysis Catheter 12 Fr Triple-Lumen UDI-DI Code: 00801741107177 5855300 Power-Trialysis Slim-Cath Straight Short-Term Dialysis Catheter 12 Fr Triple-Lumen UDI-DI Code: 00801741107184 5865120 Power-Trialysis Slim-Cath Precurved Extension Short-Term Dialysis Catheter 12 Fr Triple-Lumen UDI-DI Code: 00801741107191 5865200 Power-Trialysis Slim-Cath Precurved Extension Short-Term Dialysis Catheter 12 Fr Triple-Lumen UDI_DI Code: 00801741107214 5865240 Power-Trialysis Slim-Cath Precurved Extension Short-Term Dialysis Catheter 12 Fr Triple-Lumen UDI-DI Code: 00801741107221 The Power-Trialysis Slim-Cath Short-Term Dialysis Catheter, with a third internal lumen for intravenous therapy, power injection of contrast media, and central venous pressure monitoring, is indicated for use in attaining short-term (less than 30 days) vascular access for hemodialysis, hemoperfusion, and apheresis treatments. The catheter is intended to be inserted in the jugular, femoral, or subclavian vein as required. The maximum recommended infusion rate is 5 mL/sec for power injection of contrast media.
Due to lidocaine ampoules being manufactured under deficient manufacturing practices by manufacturer, lidocaine ampoules are currently being recalled by supplier.
- 2026-06-13Open, ClassifiedNonconforming Material/Component
Description/REF: AHDC KIT: 3-L 12 FR X 16 CM/AK-12123-PCDC; AHDC KIT: 3-L 12 FR X 20 CM/AK-15123-PCDC; AHDC KIT: 3-L 12 FR X 20 CM/CDC-15123-XP1A; AHDC KIT: 3-L 12 FR X 25 CM/CDC-16123-XP1A; AHDC KIT: 3-L 12 FR X 25 CM/CDC-16123-XPN1A; AHDC AGB U-BEND KIT: 2-L 12 FR X 20 CM/CDC-25122-XN1A-U; LBCVC AGB KIT: 3-L 12 FR X 16 CM/CDC-22123-XN1A
Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries.
- 2025-11-04Open, ClassifiedUnder Investigation by firm
MAHURKAR" Acute High Pressure Triple Lumen Catheter Tray Curved Extensions with IC Components: 12 Fr/Ch (4.0 mm) x 13 cm;
A potential issue with the seal integrity of header bag packaging.
- 2025-11-04Open, ClassifiedUnder Investigation by firm
MAHURKAR" Acute High Pressure Triple Lumen Catheter Tray Curved Extensions with IC Components: 12 Fr/Ch (4.0 mm) x 16 cm;
A potential issue with the seal integrity of header bag packaging.
- 2025-11-04Open, ClassifiedUnder Investigation by firm
MAHURKAR" Acute High Pressure Triple Lumen Catheter Tray Curved Extensions with IC Components: 12 Fr/Ch (4.0 mm) x 20 cm;
A potential issue with the seal integrity of header bag packaging.
- 2025-11-04Open, ClassifiedUnder Investigation by firm
MAHURKAR" Acute High Pressure Triple Lumen Catheter Tray Straight Extensions with IC Components: 12 Fr/Ch (4.0 mm) x 20 cm;
A potential issue with the seal integrity of header bag packaging.
- 2025-11-04Open, ClassifiedUnder Investigation by firm
MAHURKAR" Acute High Pressure Triple Lumen Catheter Tray Straight Extensions with IC Components: 12 Fr/Ch (4.0 mm) x 24 cm;
A potential issue with the seal integrity of header bag packaging.
- 2025-11-04Open, ClassifiedUnder Investigation by firm
MAHURKAR" Acute High Pressure Triple Lumen Catheter Tray Straight Extensions with IC Safety Components: 12 Fr/Ch (4.0 mm) x 24 cm;
A potential issue with the seal integrity of header bag packaging.
- 2023-08-11Open, ClassifiedProcess control
MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 13 cm, Curved Extensions, IC Tray REF:8888101001HP
Catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or excessive MDX may dislodge from the catheter lead to full catheter obstruction resulting in delay to treatment and partial obstruction resulting in reduced flow, or particulate dislodgement that may result in delay to treatment, hemolysis, embolism/embolus or thrombosis/thrombus
Monitor catheter, hemodialysis, triple lumen, non-implanted automatically
DeviceWatch watches FDA adverse events and recalls for the products you care about, summarizes clinical narratives with AI and alerts you when a safety signal emerges.