NEYClass 2

System, Ablation, Microwave And Accessories

FDA adverse event reports and recall history for system, ablation, microwave and accessories (product code NEY), from the public MAUDE corpus and the live FDA recall database.

Total reports
1,807
Deaths
45
Injuries
718
Malfunctions
1,036

Reports per year

12
13
14
15
16
17
18
19
20
21
22
23
24
25
26

Source: FDA adverse event reports by year received, 2012–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.

Recent recalls

  • 2023-01-12Open, ClassifiedProcess control

    COVIDIEN Emprint Ablation Catheter with Thermosphere Technology, Bronchoscopic Ablation, REF CA108L1

    The product is labeled with the incorrect expiration date.

  • 2016-03-17TerminatedNonconforming Material/Component

    Short Percutaneous Antenna with Thermosphere" Technology Standard Percutaneous Antenna with Thermosphere" Technology; The product is packaged in a blister with a tyvek lid, contained in a display box. The numbers of units per carton is 1. Long Percutaneous Antenna with Thermosphere" Technology Product Usage: The Covidien Emprint Ablation System is intended for use in percutaneous,laparoscopic, and intraoperative coagulation (ablation) of soft tissue, including partial or complete ablation of non-resectable liver tumors. The CovidienEmprint Ablation System is not intended for use in cardiac procedures

    Medtronic is voluntarily recalling specific item codes and production lots of Covidien Emprint" Percutaneous Antenna with Thermosphere" Technology due to the potential for the generator to shut down in the precence of high temperature measurements in the Emprint" Percutaneous Antenna.

  • TerminatedProcess control

    Covidien Emprint Ablation Catheter with Thermosphere Technology Bronchoscopic Ablation, CFN CA108L1

    The firm identified an error in the device temperature monitoring circuitry which is designed to interrupt energy delivery in the event of a primary cooling system failure. Use of an affected device in conjunction with a primary cooling system failure could lead to the applicator tip overheating and associated heat damage which could result in an allergic/toxic reaction, unintended tissue ablation, tissue burn, and/or a delay of treatment while an alternate device is located.

  • TerminatedUnder Investigation by firm

    Accu2i pMTA Applicator, 14 cm STANDARD, REF/Catalog # 900-600, Model # (UPN) H7879006000, STERILE --- This is only for use with a Sulis V Generator, REF 806-10. Product Usage: The Acculis Accu2i pMTA Microwave Tissue Ablation Applicator is indicated for the intraoperative coagulation of soft tissue.

    The firm is recalling product that has the potential to not function during use. The product affected by this recall will not deliver the desired microwave energy due to coolant ingress in an electrical connection; this condition results in an error code High Reflective Power displayed on the microwave generator.

  • TerminatedUnder Investigation by firm

    Accu2i pMTA Applicator, 29 cm LONG, REF/Catalog # 900-601, Model # (UPN) H7879006010, STERILE --- This is only for use with a Sulis V Generator, REF 806-10. Product Usage: The Acculis Accu2i pMTA Microwave Tissue Ablation Applicator is indicated for the intraoperative coagulation of soft tissue.

    The firm is recalling product that has the potential to not function during use. The product affected by this recall will not deliver the desired microwave energy due to coolant ingress in an electrical connection; this condition results in an error code High Reflective Power displayed on the microwave generator.

  • TerminatedUnder Investigation by firm

    Accu2i pMTA Applicator, 19 cm INTERMEDIATE, REF/Catalog # 900-602, Model # (UPN) H7879006020, STERILE --- This is only for use with a Sulis V Generator, REF 806-10. Product Usage: The Acculis Accu2i pMTA Microwave Tissue Ablation Applicator is indicated for the intraoperative coagulation of soft tissue.

    The firm is recalling product that has the potential to not function during use. The product affected by this recall will not deliver the desired microwave energy due to coolant ingress in an electrical connection; this condition results in an error code High Reflective Power displayed on the microwave generator.

  • TerminatedUnder Investigation by firm

    Accu2i pMTA Applicator, 14 cm STANDARD, REF/Catalog # 900-600-US, Model # (UPN) H787900600US0, STERILE --- This is only for use with a Sulis V Generator, REF 806-10. Product Usage: The Acculis Accu2i pMTA Microwave Tissue Ablation Applicator is indicated for the intraoperative coagulation of soft tissue.

    The firm is recalling product that has the potential to not function during use. The product affected by this recall will not deliver the desired microwave energy due to coolant ingress in an electrical connection; this condition results in an error code High Reflective Power displayed on the microwave generator.

  • TerminatedUnder Investigation by firm

    Accu2i pMTA Applicator, 29 cm LONG, REF/Catalog # 900-601-US, Model # (UPN) H787900601US0, STERILE --- This is only for use with a Sulis V Generator, REF 806-10. Product Usage: The Acculis Accu2i pMTA Microwave Tissue Ablation Applicator is indicated for the intraoperative coagulation of soft tissue.

    The firm is recalling product that has the potential to not function during use. The product affected by this recall will not deliver the desired microwave energy due to coolant ingress in an electrical connection; this condition results in an error code High Reflective Power displayed on the microwave generator.

  • TerminatedUnder Investigation by firm

    Accu2i pMTA Applicator, 19 cm INTERMEDIATE, REF/Catalog # 900-602-US, Model # (UPN) H787900602US0, STERILE --- This is only for use with a Sulis V Generator, REF 806-10. Product Usage: The Acculis Accu2i pMTA Microwave Tissue Ablation Applicator is indicated for the intraoperative coagulation of soft tissue.

    The firm is recalling product that has the potential to not function during use. The product affected by this recall will not deliver the desired microwave energy due to coolant ingress in an electrical connection; this condition results in an error code High Reflective Power displayed on the microwave generator.

  • Open, ClassifiedDevice Design

    Covidien Emprint Percutaneous Antenna with Thermosphere Technology, 15cm, Material CA15L1

    The firm received consumer reports of the ceramic trocar tip of the Emprint ablation antenna disengaging from the needle shaft post-ablation. This can result in the tip remaining in the patient.