Suture, Surgical, Absorbable, Polydioxanone

FDA adverse event reports and recall history for suture, surgical, absorbable, polydioxanone (product code NEW), from the public MAUDE corpus and the live FDA recall database.

Total reports
16,282
Deaths
41
Injuries
3,433
Malfunctions
12,527

Reports per year

12
13
14
15
16
17
18
19
20
21
22
23
24
25
26

Source: FDA adverse event reports by year received, 2012–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.

Recent recalls

  • 2025-09-19Open, ClassifiedUnder Investigation by firm

    STRATAFIX Spiral PDS Plus Bidirectional Knotless Tissue Control Device, SFX SPI PDS+ BI VIO 14X14IN(36X36CM) USP1(M4) D/A MO-4. Product Code: SXPP2B400-11.

    Potential for barb non-engagement.

  • 2025-04-03Open, ClassifiedProcess control

    PDSPLUSVIO30IN(75CM)USP6-0(M0.7)D/AC-1MP

    Identified curing issues with the silicone during the needles manufacturing process.

  • 2025-04-03Open, ClassifiedProcess control

    PDSIIVIO30IN(75CM)USP6-0(M0.7)D/AC-1

    Identified curing issues with the silicone during the needles manufacturing process.

  • 2025-04-03Open, ClassifiedProcess control

    PROLENE BLU 18IN(45CM) USP6-0(M0.7) S/A P-1 PRM

    Identified curing issues with the silicone during the needles manufacturing process.

  • 2025-04-03Open, ClassifiedProcess control

    ETHIBOND EXCEL GRN 36IN(90CM) USP2-0(M3) D/A MH

    Identified curing issues with the silicone during the needles manufacturing process.

  • 2024-09-30Open, ClassifiedUnder Investigation by firm

    PDS Plus Antibacterial (polydioxanone) Suture, Product Numbers PDP072H40, PDP259H50, PDP416H50, PDP441H50, PDP442H50, PDP489G50, PDP489H40, PDP497G51, PDP497H16, PDP497H40, PDP498H40, PDP502V05, PDP507V05, PDP510V05, PDP515V05, PDP662H50, PDP663H50, PDP684H40, PDP9615H40, PDP9625H40, PDP9626H40, PDP9631H40, PDP9715H15, PDP9715H40, PDP9733H16, PDP9861H40, PDP9865H40, XWPDP421H40, XWPDP423H40, XWPDP442H40, XWPDP443H16

    Internal testing on returned product from this lot confirmed that some PDS" II (polydioxanone) Sutures from this lot did not meet tensile strength requirement.

  • 2024-09-30Open, ClassifiedUnder Investigation by firm

    PDS II (polydioxanone) Suture, Product Numbers Z844G-50, D10065-05, D10085-05, D6776-05, D6776-05, D9297-05, D9297-05, Z259H-50, Z259H-50, Z259H-50, Z259H-50, Z259H-50, Z259H-50, Z259H-50, Z259H-50, Z259H-50, Z259H-50, Z259H-50, Z259H-50, Z259H-50, Z259H-50, Z259H-50, Z416H-50, Z416H-50, Z416H-50, Z416H-50, Z416H-50, Z416H-50, Z416H-50, Z416H-50, Z421H-50, Z421H-50, Z421H-50, Z422H-50, Z422H-50, Z422H-50, Z422H-50, Z422H-50, Z422H-50, Z422H-50, Z423H-50, Z423H-50, Z423H-50, Z442H-50, Z442H-50, Z442H-50, Z443H-50, Z443H-50, Z443H-50, Z443H-50, Z443H-50, Z443H-50, Z443H-50, Z443H-50, Z489G-50, Z489G-50, Z489G-50, Z489G-50, Z493G-50, Z493G-50, Z493G-50, Z493G-50, Z493G-50, Z494G-50, Z495G-05, Z495G-05, Z495G-05, Z496G-50, Z496G-50, Z496G-50, Z497G-50, Z497G-50, Z497G-50, Z497G-50, Z497G-50, Z497G-50, Z497G-50, Z497G-50, Z497G-50, Z497G-50, Z503G-50, Z504G-50, Z507G-50, Z507G-50, Z683G-05, Z833G-05, Z833G-05, Z844G-50

    Internal testing on returned product from this lot confirmed that some PDS" II (polydioxanone) Sutures from this lot did not meet tensile strength requirement.

  • 2024-06-14Open, ClassifiedUnder Investigation by firm

    PDS Plus Antibacterial (polydioxanone) Suture - Intended for use in general soft tissue approximation, including use in paediatric cardiovascular tissue, in microsurgery and in ophthalmic surgery. These sutures are particularly useful where the combination of an absorbable suture and extended wound support (up to six weeks) is desirable. PRODUCT CODE: PDP304H PDP316H PDP341H

    Issue on a specific packaging machine resulted in a hole in the primary packaging, product sterility compromised and could result in infection .

  • 2023-05-10Open, ClassifiedUnder Investigation by firm

    PDS Plus Antibacterial (polydioxanone) Suture - PDS PLUS CLR 27IN(70CM) USP3-0(M2) S/A PC-25 PRIME. Intended for use in general soft tissue approximation, including use in pediatric cardiovascular tissue. Product Code: PDP9625H

    Suture breakage during intra-operative use could result in poor performance of the impacted product because the intended benefit of tissue approximation and/or ligation may not be achieved

  • 2022-12-14Open, ClassifiedUnder Investigation by firm

    PDS II (polydioxanone) Suture- VIOLET MONOFILAMENT 27IN 0 S/A CT-1 Product Code: Z340H For use in soft tissue approximation, including use in pediatric cardiovascular tissue where growth is expected to occur and ophthalmic surgery

    knot tensile strengths were observed to be significantly below the minimum knot tensile, will result in breakage prior to use or intra-operatively rather than premature post-operative breakage