Marker, Radiographic, Implantable
FDA adverse event reports and recall history for marker, radiographic, implantable (product code NEU), from the public MAUDE/AEMS corpus and the live FDA recall database.
- Total reports
- 1,839
- Deaths
- 1
- Injuries
- 1,162
- Malfunctions
- 646
Reports per year
Source: FDA adverse event reports by year received, 2012–2026. Reporting paused during the 2026 MAUDE-to-AEMS transition and has since resumed.
Recent recalls
- 2026-05-27Open, ClassifiedLabeling mix-ups
HydroMARK Breast Biopsy Site Marker, Model Number: 4010-02-15-T3
Incorrect inner package labeling. Boxes labeled 4010-02-15-T3 markers may contain individual packages labeled as T4 Markers.
- 2026-02-17Open, ClassifiedUnder Investigation by firm
MOLLI 2 System System includes: MM1000 (Pack of 10) - MOLLI Introducer (8cm, 14G) and MOLLI Marker UDI (GTIN): 00850024195028 MS-00009 (Pack of 10) - MOLLI Introducer (12cm, 14G) and MOLLI Marker UDI (GTIN): 00850024195134 MP2001 - MOLLI 2 Wand UDI (GTIN): 00850024195141 MP2015A - MOLLI OncoPen UDI (GTIN): 00850024195219 MT2001 - MOLLI 2 Tablet UDI (GTIN): 00850024195158 The MOLLI Marker is intended to be placed percutaneously in soft tissue to temporarily mark a surgical site intended for surgical removal. the MOLLI Marker is located and surgically removed with the target tissue.
Due to a need to update the product labeling (Instructions for Use IFU) as a result of the magnetized surgical tools dislodging markers using during target tissue removal procedures.
- 2025-10-24Open, ClassifiedProcess control
Magseed Pro 7 cm soft tissue marker. Model Number: MS2-17-1-07. Product Description: The Endomag Magseed Pro Magnetic Marker System is a sterile, single use device composed of a marker preloaded in a 17ga Needle Delivery System. The Endomag Magseed Pro Marker is indicated for use to radiographically mark soft tissue during a surgical procedure or for future surgical procedures.
Potential for contamination with cotton fibers.
- 2025-10-24Open, ClassifiedProcess control
Magseed Pro 12 cm soft tissue marker. Model Number: MS2-17-1-12. Product Description: The Endomag Magseed Pro Magnetic Marker System is a sterile, single use device composed of a marker preloaded in a 17ga Needle Delivery System. The Endomag Magseed Pro Marker is indicated for use to radiographically mark soft tissue during a surgical procedure or for future surgical procedures.
Potential for contamination with cotton fibers.
- 2025-01-06Open, ClassifiedDevice Design
EnVisio Patient Pad Transition Cushion, Model/Catalog Number: ENVISIO-SYS Product Description: The EnVisio Patient Pad Transition Cushion is a component of the EnVisio Navigation System which is a medical device that detects the presence of the SmartClip Soft Tissue Marker(s) and provides real-time three-dimensional navigation during surgery.
There is a potential safety risk caused by physical disassociation of the back section from the bed frame of Skytron model series 3600, 6500, 6600 and 6700 operating room beds when used with the EnVisio Navigation System.
- 2024-11-26Open, ClassifiedDevice Design
BioZorb Marker. Radiographic soft tissue marker.
Firm received reports of adverse events/complications that include serious injuries, such as pain, infection, rash, device migration, device erosion, seroma, and discomfort.
- 2024-05-03Open, ClassifiedUnknown/Undetermined by firm
BioZorb Marker-BioZorb 3D Bioabsorbable Marker-Intended Use radiographic marking of sites in soft tissue. 1) F0405 BioZorb Marker 4cm x 5cm 2) F0404 BioZorb Marker 4cm x 4cm 3) F0331 BioZorb Marker 1cm x 3cm x 3cm 4) F0231 BioZorb Marker 1cm x 3cm x 2cm 5) F0221 BioZorb Marker 1cm x 3cm x 2cm 6) F0304 BioZorb Marker 3cm x 4cm 7) F0303 BioZorb Marker 3cm x 3cm 8) F0203 BioZorb Marker 2cm x 3cm 9) F0202 BioZorb Marker 2cm x 2cm
Safety Information: Complaints describe complications/adverse events that include pain, infection, rash, device migration, device erosion, seroma, discomfort, or other complications from feeling the device in the breast, and the need for additional medical treatment to remove the device
- 2023-01-30TerminatedUnder Investigation by firm
Mammotome MammoStar Biopsy Site Identified, REF STAR1401
The label incorrectly identifies the product as STAR1401, a 1x3mm Barbell shaped marker pre-loaded in a 14 gauge needle, while the packaging contains STAR1402, a 1x5mm Tribell shaped marker pre-loaded in a 14 gauge needle.
- 2022-09-30Open, ClassifiedIncorrect or no expiration date
The BioZorb LP Marker is an implantable radiopaque marker used to facilitate visualization of a soft tissue site. The BioZorb LP Marker is comprised of a bioabsorbable PLA (polylactic acid) component which resorbs completely in 1 year or more and a permanent component (titanium). The BioZorb LP Marker is provided sterile for single use and is implantable.
Hologic BioZorb LP Marker, (Product number F0221), lot number 22A0RL, contained removable labels with an expiration date that indicated the product had already expired. The expiration date on the removable label is actually manufacturing date, and it's different from the one on the shelf box label, which is the correct expiration date.
- 2013-06-10TerminatedProcess control
BARD Biopsy Systems (A Business Unit of Bard Peripheral Vascular, Inc.), UltraClip Dual Trigger Breast Tissue Marker 17g X 10cm needle, Ultrasound Enhanced Coil Marker 5 Pack; Sterilize, Non-pyrogenic, Rx only, Single Use; Bard Peripheral Vascular, Inc., 1625 West 3rd Street, Tempe, AZ 85281 USA. PK4320100 Rev. 0 03/10 The UltraClip Dual Trigger breast tissue marker is a sterile, single use device comprised of a disposable introducer, and a metal implantable tissue marker with polyvinyl alcohol (PVA).
Bard Peripheral Vascular (BPV) is initiating this recall because a combination of Bard UltraClip Dual Trigger Tissue Marker's and support catheters (Microsheath and Usher) were inadvertently distributed to customers without completing the sterilization process (non sterile).
Monitor marker, radiographic, implantable automatically
DeviceWatch watches FDA adverse events and recalls for the products you care about, summarizes clinical narratives with AI and alerts you when a safety signal emerges.