Calculator, Drug Dose
FDA adverse event reports and recall history for calculator, drug dose (product code NDC), from the public MAUDE corpus and the live FDA recall database.
- Total reports
- 2,093
- Deaths
- 2
- Injuries
- 178
- Malfunctions
- 1,903
Reports per year
Source: FDA adverse event reports by year received, 2012–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.
Recent recalls
- 2026-07-23Open, ClassifiedSoftware design
MiniMed Go App, Model: MMT-8600 (iOS), MMT-8601 (Android)
For MiniMed Go app version 1.0.0 and 1.1.0, when the Set day and night is turned OFF and missed dose or Correct high glucose alerts are enabled, the associated alerts will not be triggered during the nighttime. This could result in missed rapid-acting insulin doses without the user s knowledge, and may lead to delayed therapy and an increased risk of hyperglycemia.
- 2025-12-18Open, ClassifiedSoftware Design Change
InPen App, Model/CFN Number: MMT-8061 (Android Users)
When app is uninstalled and reinstalled, insulin pen software issue causes Choose Notification Style Screen to not show during setup so users can't choose to allow notifications to override phone settings when on mute/Do Not Disturb, and previously set up override permission is deleted, so audible and vibratory notifications not received, which may lead to delayed insulin therapy, hyperglycemia.
- 2025-09-02Open, ClassifiedUnder Investigation by firm
InPen App, CFN: MMT-8060 (iOS Users) and MMT-8061 (Android Users)
Medtronic MiniMed, Inc. is recalling InPen App for iOS and Android users due to software design errors that could lead to a missed short-acting insulin dose reminder and a recommendation to correct a high glucose value. It does not impact insulin delivery, long-acting insulin dose reminders, or CGM alerts, and users can still use the pen itself to calculate a dose, deliver insulin, record the dose date/time, and view CGM data. This issue was identified during internal testing before release in the US but after release to OUS customers, no complaints or MDRs related to this recall have been reported. Use of the affected device may result hyperglycemia by failing to alert the user and delay treatment of diabetes.
- 2022-12-20Open, ClassifiedSoftware design
InPen Diabetes Management App, part of the InPen System, REF: MMT-105ELBLNA, MMT-105ELGYNA, MMT-105ELPKNA, MMT-105ELPKNA, MMT-105NNBLNA, MMT-105NNGYNA, MMT-105NNPKNA.
Diabetes management app may crash immediately upon opening, while using the app and/or while app is running in the background, which may lead to inability to use the app. An error message may be displayed indicating the app has stopped or isn't responding. After the crash, a present dose reminder is not displayed, which may lead to delayed insulin therapy and potentially resulting in hyperglycemia.
- 2017-01-26TerminatedSoftware design
Accu-Chek Connect Diabetes Management App
iOS and Android: Under certain conditions the affected app versions may disregard historical bolus data potentially leading to an incorrect bolus insulin recommendation being provided to the user. iOS only: Pairing and using multiple meters with the Accu-Chek Connect app can under rare circumstances cause the bolus advisor to fail to offer a correction bolus recommendation within the eligible time window following a blood glucose measurement (10 15 minutes). Depending on the individual metabolic situation potentially incorrect bolus advice could lead to serious health consequences such as hypoglycemia. Both software issues may also cause the amount of active insulin displayed during the bolus calculation process to be incorrect and should not be used to manually calculate a bolus.
- 2016-11-16TerminatedSoftware design (manufacturing process)
Version 1.2.0 of Accu-Chek Connect Diabetes Management App (iOS) released on July 11, 2016
Roche Diabetes Care, Inc. initiated a voluntary recall for the version 1.2.0 of Accu-Chek Connect Diabetes Management App (iOS), released on July 11, 2016, due to the software containing a programming error (bug) in the Bolus Advisor feature, which could lead to an incorrect insulin bolus recommendation.
- TerminatedSoftware design
Accu-Chek Connect diabetes management software app, Catalog number 07562462001 / GTIN number 00365702700000, Catalog number 07250452001 / GTIN number 00365702700017 The Accu-Chek Connect Diabetes Management App is indicated as an aid in the treatment of diabetes. The software provides for electronic download of blood glucose meters, manual data entry, storage, display, transfer, and self-managing of blood glucose and other related health indicators which can be shown in report and graphical format. The Accu-Chek Bolus Advisor, as a component of the Accu-Chek Connect Diabetes Management App, is indicated for the management of diabetes by calculating an insulin dose or carbohydrate intake based on user-entered data
Certain iOS and Android App versions contain a program error (bug) in the Bolus Advisor feature. After pairing a meter with the app for the first time, a customer may encounter the rare condition in which the countdown timer is not displayed and correction bolus advice is not available for the most recent, valid glucose reading. This same blood glucose value may become available for bolus advice calculation at a later time (countdown timer is displayed).
- TerminatedSoftware change control
EndoTool SubQ (Versions 1.7.1, 1.7.4, 1.7.5)
Product was distributed prior to approval or clearance from FDA.
- TerminatedSoftware in the Use Environment
Insulia Diabetes Management Companion
The firm identified an issue with the basal calculator identified on the Android version of the Insulia application. The bug could lead from low-impact to high-impact hypoglycemia depending on the circumstances. A new update of Insulia is now available (version 1.6.42), which resolves this issue.
- TerminatedOther
Accu-Chek Connect Diabetes Management App, Catalog number 07562462001 / GTIN number 00365702700000 & Catalog number 07250452001 / GTIN number 00365702700017
Certain iOS and Android App versions contain a program error (bug) in the Bolus Advisor feature. Due to a software bug, when the OS region of the phone setting is changed, the unit of measure within the app may unexpectedly change. This creates a risk the app might not transfer the blood glucose result or the user might not correctly input numerical values for carbohydrate used for bolus advice.
Monitor calculator, drug dose automatically