MUEClass 2

Full Field Digital, System, X-Ray, Mammographic

FDA adverse event reports and recall history for full field digital, system, x-ray, mammographic (product code MUE), from the public MAUDE corpus and the live FDA recall database.

Total reports
670
Deaths
5
Injuries
54
Malfunctions
498

Reports per year

12
13
14
15
16
17
18
19
20
21
22
23
24
25
26

Source: FDA adverse event reports by year received, 2012–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.

Recent recalls

  • 2026-05-12Open, ClassifiedDevice Design

    SELENIA DIMENSIONS MAMMOGRAPHY SYSTEMS. Model Numbers: SDA-SYS-3000-2D-HTC, SDA-SYS-3000-3D, SDM-SYS-6000-2D, SDM-SYS-6000-3D, SDM-00001-2D, SDM-00001-3D, SDM-00001-M2D, SDM-00001-M3D, SDM-05000-2A2, SDM-05000-2A3, SDM-05000-2AC, SDM-05000-2D2, SDM-05000-2D3, SDM-05000-2DC, SDM-05000-3D2, SDM-05000-3D3, SDM-05000-3DC, SDM-SYS-9000-2D, SDM-SYS-9000-3D.

    Complaints have been received of systems developing loose, missing, or broken internal bolts over time.

  • 2025-12-09Open, ClassifiedRadiation Control for Health and Safety Act

    Senographe Pristina, Pristina Serena, and Pristina Serena 3D. for screening and diagnostic mammography

    Some Senographe Pristina systems either have a missing or an incorrect X-Ray Warning label, which is a nonconformance to FDA Regulation 21 CFR 1020.30(j).

  • 2025-12-03Open, ClassifiedComponent design/selection

    MAMMOMAT Revelation;

    There were instances where the operator table was sold together with a bus-installation kit. The operator table is not intended and not released for bus installations as it was designed for stationary operation only.

  • 2025-12-03Open, ClassifiedComponent design/selection

    MAMMOMAT Inspiration;

    There were instances where the operator table was sold together with a bus-installation kit. The operator table is not intended and not released for bus installations as it was designed for stationary operation only.

  • 2025-12-03Open, ClassifiedComponent design/selection

    MAMMOMAT Fusion;

    There were instances where the operator table was sold together with a bus-installation kit. The operator table is not intended and not released for bus installations as it was designed for stationary operation only.

  • 2024-02-08Open, ClassifiedRadiation Control for Health and Safety Act

    Senographe Pristina

    X-ray exposure termination audible signal on Senographe Pristina with Efficiency Suite 2 intermittently not emitted per 21CFR 1020.31(j) requirements.

  • 2020-06-16TerminatedDevice Design

    MAMMOMAT Revelation, model no. 11343300 - Product Usage: is intended to be used for mammography exams, screening, diagnostics, biopsies and dual energy procedures under the supervision of medical professionals.

    If an examination is interrupted when the InSpect Projection View (PV) is active and the vacuum biopsy system is being used, the tube arm's manual movement range is decreased from +/-15 deg. to +/- 6 deg. This limitation may lead to difficulties in removing the vacuum biopsy system.

  • 2020-02-18TerminatedProcess control

    Selenia Dimensions Mammography System, Model number ASY-04160, Selenia Dimensions Gantry, UDI 15420045505636,

    The recalling firm has received reports from the field of the C-arm on the Selenia Dimensions system seizing or dropping during field repairs and cleaning of the systems.

  • 2018-01-17TerminatedEmployee error

    Aspire HD Plus (FDR MS-2500), also known as Aspire HD Plus F and Aspire HD-s (FDR MS-2000), also known as Aspire HD Plus S). Mammography machine The Fujifilm Digital Mammography Systems, Aspire HD Plus (FDR MS-2500) and Aspire HD-s (FDR MS-2000), generate full-field digital mammography images that can, as other fall-field digital mammography systems, be used for screening and diagnosis of breast cancer and are intended for use in the same clinical applications as traditional screen-film mammography systems.

    There was a case that the swivel arm went down unintentionally when it was moved up/down.

  • 2018-01-17TerminatedEmployee error

    ASPIRE Cristalle (FDR MS-3500) Mammography machine The Fujifilm Digital Mammography Systems, Aspire HD Plus (FDR MS-2500) and Aspire HD-s (FDR MS-2000), generate full-field digital mammography images that can, as other fall-field digital mammography systems, be used for screening and diagnosis of breast cancer and are intended for use in the same clinical applications as traditional screen-film mammography systems.

    There was a case that the swivel arm went down unintentionally when it was moved up/down.

Monitor full field digital, system, x-ray, mammographic automatically