Skip to main content
MRNClass 2

Apparatus, Nitric Oxide Delivery

FDA adverse event reports and recall history for apparatus, nitric oxide delivery (product code MRN), from the public MAUDE/AEMS corpus and the live FDA recall database.

Total reports
2,044
Deaths
30
Injuries
139
Malfunctions
1,828

Reports per year

12
13
14
15
16
17
18
19
20
21
22
23
24
25
26

Source: FDA adverse event reports by year received, 20122026. Reporting paused during the 2026 MAUDE-to-AEMS transition and has since resumed.

Recent recalls

  • 2025-10-06Open, ClassifiedSoftware design

    NOxBOXi Nitric Oxide Delivery System; Model Numbers: NOXBOX-I

    Fluctuations may be more likely to occur if the total flow through the device sensor (used by the device to determine flow) is less than the minimum device requirement of 0.5 LPM. In addition, dose fluctuations have been observed when Bunnel LifePulse HFJV system or conventional ventilators have the following device settings: low service pressures generally less than 3.0 PSI, a pressure difference greater than 5.0cm H2O; or at breath per minute rates of 300 (5 Hz).

  • 2025-10-06Open, ClassifiedSoftware design

    NOxBOXi Nitric Oxide Delivery System; Model Numbers: NOXBOX-I

    The unexpected internal fault / system diagnostic error to result when a rapid succession of button selections is made to the user interface without waiting for the device to respond to user prompts.

  • 2025-09-09Open, ClassifiedComponent design/selection

    NOxBOXi Nitric Oxide Delivery System

    Replacement of the internal Sample Pump due to high failure rates.

  • 2023-12-15Open, ClassifiedProcess control

    VERO BIOTECH GENOSYL DS (Delivery System) Generation 3 Cassette, Model Number 602722-01; generates and delivers nitric oxide (NO) for inhalation at the point of use

    Cassette did not meet an 1n-process flow criterion.

  • 2023-06-09TerminatedComponent change control

    NOxBOXi Nitric Oxide Delivery System, NOXBOX-I, REQNOXBOX-I, Nitric Oxide Delivery Device

    Device may experience malfunctions due to misalignment of the check valve in the manifold. If the device fails, there is a risk of nitric oxide leak, oxygen gas leak, or earlier cylinder gas changeover than expected or not at all. In a worst case scenario, there may be an interruption in therapy which could lead to oxygen desaturation of the patient and associated complications.

  • 2022-09-13Open, ClassifiedUnder Investigation by firm

    INOmax DSIR -(delivery system) delivers INOMAX (nitric oxide for inhalation) therapy gas into the inspiratory limb of the patient breathing circuit in a way that provides a constant concentration of nitric oxide (NO), as set by the user, Model Number: 10007

    The INOmax DSIR Delivery System, manufactured by Mallinckrodt Pharmaceuticals, is not compatible with the upgraded version of the Leoni Plus ventilator, manufactured by Lowenstein Medical. The Leoni Plus ventilator was recently upgraded in 2021 by Lowenstein Medical, with new hardware and software, and the upgraded version cannot be appropriately connected with the INOmax DSIR as instructed in the INOmax DSIR Operators Manual (OM). The INOmax DSIR OM requires the use of a one-way valve with the Leoni Plus ventilator during High Frequency Oscillatory Ventilation mode and warns that use of the device without a one-way valve may result in high Nitric Oxide (NO) delivery.

  • 2021-06-23TerminatedSoftware design

    GENOSYL DS; Nitric Oxide Delivery System Combination Product NDA202860

    Reports received of NO measured below desired dose during transition between primary console and backup console since Software Version 2.2.3 was uploaded to consoles in the field.

  • 2015-02-27TerminatedProcess change control

    Ikaria, INOmax DSIR (Delivery System), Model 10007. Nitric oxide delivery system for use with ventilators.

    Potential delivery failure alarm condition. INOmax DSIR with software version 2.0.4 and a certain variant of the monitor display may trigger a delivery failure alarm when the display brightness is set to its lowest level. If this condition occurs, therapy will be interrupted.

  • 2014-12-12TerminatedProcess change control

    INOMAX DSIR Nitric Oxide delivery system, Model 10007, software version 3.1.0., delivers INOMAX (nitric oxide for inhalation) therapy gas and provides continuous monitoring of inspired O2, NO2, and NO, and comprehensive alarm systems. The primary clinical settings are Neonatal Intensive Care Units (NICU) and to transport of neonates.

    An issue has been identified in the INOmax DSIR system that could result in monitored Nitric Oxide (NO) concentration reporting lower than expected. This issue only pertains to those devices manufactured using a specific version of the Monitoring Circuit Board.

  • 2014-05-02TerminatedSoftware Design Change

    INOmax DSIR, model 10007, Nitric Oxide Delivery System, with Software version 3.0.0. The INOmax DS delivery system delivers INOMAX (nitric oxide for inhalation) therapy gas and provides continuous monitoring of inspired O2, NO2, and NO, and a comprehensive alarm system. The primary targeted clinical setting is the Neonatal Intensive Care Unit (NICU) and secondary targeted clinical setting is the transport of neonates.

    An anomaly impacts the monitoring system of INOmax DSIR Plus (software 3.0.0). When the integrated pneumatic backup switch is activated with low or no gas flow through the injector module, an intended NO cell failure alarm will occur. In addition, an unintended NO2 cell failure alarm may also occur. The NO/NO2 cell failure alarms are caused by an average NO/NO2 circuit concentration greater tha

Monitor apparatus, nitric oxide delivery automatically

DeviceWatch watches FDA adverse events and recalls for the products you care about, summarizes clinical narratives with AI and alerts you when a safety signal emerges.