Infant Heel Warmer (Chemical Heat Pack)
FDA adverse event reports and recall history for infant heel warmer (chemical heat pack) (product code MPO), from the public MAUDE corpus and the live FDA recall database.
- Total reports
- 501
- Deaths
- 0
- Injuries
- 57
- Malfunctions
- 428
Reports per year
Source: FDA adverse event reports by year received, 2012–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.
Recent recalls
- 2023-07-17Open, ClassifiedProcess control
NovaPlus Infant Heel Warmer, Instant Squeeze Activation, Catalog Number V11460-010
Cardinal Health has received an increase in burst/leaking complaints from mid-April to May regarding the above listed lot of the Novalplus Infant Heel Warmer with tape.
- 2022-01-19Open, ClassifiedNo Marketing Application
Heel Snuggler, Reference # 989805603411 99047
After the device was cleared by the FDA, Philips made changes to the labeling that do not fall within the existing FDA clearance. These devices with the modified labeling cannot be distributed or sold without new 510(k) clearance.
- 2015-08-21TerminatedPackage design/selection
Covidien Argyle Infant Heel Warmers (no tab) Item Code: MH00002N Product Usage: Intended for application of the right amount of heat to an infants heel for blood sampling
Infant Heel warmers may break during activation resulting in spillage of the contents on patients or healthcare worker causing potential skin/eye irritations
- 2015-08-21TerminatedPackage design/selection
Covidien Argyle Infant Heel Warmers (with tab) Item Code: MH00002T Product Usage: Intended for application of the right amount of heat to an infants heel for blood sampling.
Infant Heel warmers may break during activation resulting in spillage of the contents on patients or healthcare worker causing potential skin/eye irritations
- 2015-07-20TerminatedNo Marketing Application
Infant Heel Warmer and Infant Gel Warmer, sold under the brand names CooperSurgical and Fisher brand as follows : CooperSurgical PN 20418 WARMGEL INFANT HEEL WARMER - PRISM BOX OF 100; PN 24401 LIQUID INFANT HEEL WARMER - PRISM BOX OF 100; and Fisher Brand. PN 24647 FISHERBRAND GEL INFANT - HEEL WARMER BOX OF 100; PN 24646 FISHERBRAND LIQUID INFANT - HEEL WARMER- 100 PER BOX
Product marketed without a 510(k)
- TerminatedPackage design/selection
Infant Heel Warmer with Attachment Tab, Nonsterile, Active Product SKU MH00002T The infant heel warmer is activated at the point of care by squeezing the pouch. This mixes the substrates inside the pouch causing an exothermic chemical reaction, warming the pouch, which is then applied to the heel of an infant to aid in blood draws
A caution statement is being added to remind the user to activate away from face and infant as reports of pack rupture or leaking during activation may resulting in the contents splashing into the user s eyes or onto the infant. Potential harms associated with contact of the liquid inside the pouch are rash, burn and inflammation.
- TerminatedPackage design/selection
Infant Heel Warmer without Attachment Tab, Nonsterile, Active Product SKU MH00002N The infant heel warmer is activated at the point of care by squeezing the pouch. This mixes the substrates inside the pouch causing an exothermic chemical reaction, warming the pouch, which is then applied to the heel of an infant to aid in blood draws
A caution statement is being added to remind the user to activate away from face and infant as reports of pack rupture or leaking during activation may resulting in the contents splashing into the user s eyes or onto the infant. Potential harms associated with contact of the liquid inside the pouch are rash, burn and inflammation.
Monitor infant heel warmer (chemical heat pack) automatically