Skip to main content
MOIClass 2

System, Test, Immunological, Antigen, Tumor

FDA adverse event reports and recall history for system, test, immunological, antigen, tumor (product code MOI), from the public MAUDE/AEMS corpus and the live FDA recall database.

Total reports
208
Deaths
0
Injuries
0
Malfunctions
193

Reports per year

12
13
14
15
16
17
18
19
20
21
22
23
24
25
26

Source: FDA adverse event reports by year received, 20122026. Reporting paused during the 2026 MAUDE-to-AEMS transition and has since resumed.

Recent recalls

  • 2022-08-29Open, ClassifiedUnder Investigation by firm

    Dimension Vista LOCI CA 15-3 Flex reagent cartridge. In Vitro Diagnostic for the quantitative measurement of CA 15-3. Siemens Material Number (SMN)/Ref Catalog Number : 10464344/K6456

    Negative bias with patient samples. The mean patient sample bias range from -26% to -33%,potential risk of delay in appropriate treatment and management decisions.

  • 2018-05-04TerminatedError in labeling

    IMMULITE /IMMULITE 1000 BR-MA (CA15-3)

    Concentrations for the level of biotin that does not interfere is not currently listed in the instruction for use for the Dimension Vista CTNI, Dimension Vista MMB, Dimension Sirolimus, ADVIA Centaur FOL, ADVIA Centaur aHBcM and the following IMMULITE assays, 3gAllergy Specific IgE, anti-HBc, BR-MA (CA15-3), CEA, Folic Acid, Gastrin, OM-MA (CA125), Thyroglobulin, and Vitamin B12, and these assays are susceptible to interference from biotin.

  • 2018-05-04TerminatedError in labeling

    IMMULITE 2000/IMMULITE 2000 XPi BR-MA (CA15-3)

    Concentrations for the level of biotin that does not interfere is not currently listed in the instruction for use for the Dimension Vista CTNI, Dimension Vista MMB, Dimension Sirolimus, ADVIA Centaur FOL, ADVIA Centaur aHBcM and the following IMMULITE assays, 3gAllergy Specific IgE, anti-HBc, BR-MA (CA15-3), CEA, Folic Acid, Gastrin, OM-MA (CA125), Thyroglobulin, and Vitamin B12, and these assays are susceptible to interference from biotin.

  • 2012-12-11TerminatedDevice Design

    Siemens Healthcare BR Assay for CA 27.29 250 Test Kits 03896216 50 Test Kit 02419937 Ref Kit (US only) 10340081 Intended Use The ADVIA Centaur BR assay is an in vitro diagnostic test for the quantitative serial determination of cancer antigen CA 27.29 in human serum using the ADVIA Centaur and ADVIA Centaur XP systems. The test is intended for use as an aid in monitoring patients previously treated for Stage II or Stage Ill breast cancer.

    Operating range of the ADVIA Centaur systems is 18C to 30C (64.4F to 86.0F), a change in room temperature may increase or decrease assay results, depending on the specific assay and ADVIA Centaur system used.

  • 2012-10-28TerminatedUnder Investigation by firm

    Siemens ADVIA Centaur BR Assay: Catalog Number: US: REF 02419937 (10214722) (50 Tests); REF 03896216(10334837)(250 Tests) Not CE Marked; OUS: REF03204829(10310271)(50 Tests); REF 01157807(1 0314722) (250 Tests) CE Marked; REF 09010686 (10340081)(250 Tests). The ADVIA Centaur BR assay is an in vitro diagnostic test for the quantitative serial determination of cancer antigen CA 27.29 in human serum using the ADVIA Centaur and ADVIA Centaur XP systems. The test is intended for use as an aid in monitoring patients previously treated for Stage II or Stage Ill breast cancer. Serial testing for CA 27.29 in the serum of patients who are clinically free of disease should be used in conjunction with other clinical methods used for the early detection of cancer recurrence. The test is also intended for use as an aid in the management of breast cancer patients with metastatic disease by monitoring the progression or regression of disease in response to treatment.

    BR Assay for CA 27.29 Calibration Interval Change due quality control (QC) results outside of the expected ranges.

  • 2007-07-11TerminatedOther

    AxSYM CA 15-3 Reagent Pack; list 03B42; 100 test pack; an in vitro diagnostic; Abbott Laboratories, Abbott Park, IL 60064

    Failure of the AxSYM instrument to open (actuate) some reagent packs from certain lots. The defects can result in probe crashes if not detected prior to placement of the reagent kit on the AxYM instrument.

  • 2006-07-18TerminatedOther

    CA 15-3 Controls, list 9C08-10; an In Vitro diagnostic for use with the AxSYM or IMx CA 15-3 reagents, calibrators and instruments; each kit contains 2 bottles with 115D8:DF3 reactive determinants (human) prepared in TRIS buffer with protein (bovine) stabilizers to yield the following assay values (U/ml): Control L - 35 and Control H - 150; Abbott Laboratories, Abbott Park, IL 60064

    The CA 15-3 Controls replicates out of range high resulting in invalid runs.

  • 2003-03-26TerminatedOther

    Abbott AxSYM CA 15-3 Master Calibrator Pack, list 3B42-30; the pack contains two 4-mL bottles, with Master Calibrator 1 having 0 U/mL and Master Calibrator 2 having 60 U/mL; Abbott Laboratories, Abbott Park, IL 60064

    The CA 15-3 Calibrators may cause depressed values for the CA 15-3 High Control and for patient specimens.

  • 2003-03-26TerminatedOther

    CA 15-3 Calibrators, list 9C08-01; the pack contains six 4-mL bottles of 115D8:DF3 reactive determinants prediluted in TRIS buffer with protein stabilizers to yield the following assay values: A - 0 u/mL, B - 15 U/mL, C - 60 U/mL, D - 120 U/mL, E - 180 U/mL and F - 250 U/mL; Abbott Laboratories, Abbott Park, IL 60064

    The CA 15-3 Calibrators may cause depressed values for the CA 15-3 High Control and for patient specimens.

  • 2003-03-26TerminatedOther

    IMx CA 15-3 Reagent Pack, list 6A75-22; this reagent pack contains one 4-mL bottle of Mode 1 Calibrator along with other reagents; Abbott Laboratories, Abbott Park, IL

    The CA 15-3 Calibrators may cause depressed values for the CA 15-3 High Control and for patient specimens.

Monitor system, test, immunological, antigen, tumor automatically

DeviceWatch watches FDA adverse events and recalls for the products you care about, summarizes clinical narratives with AI and alerts you when a safety signal emerges.