MOFClass 2

Guide, Wire, Catheter, Neurovasculature

FDA adverse event reports and recall history for guide, wire, catheter, neurovasculature (product code MOF), from the public MAUDE corpus and the live FDA recall database.

Total reports
670
Deaths
15
Injuries
180
Malfunctions
472

Reports per year

08
12
13
15
16
17
18
19
20
21
22
23
24
25
26

Source: FDA adverse event reports by year received, 2008–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.

Recent recalls

  • 2024-05-31Open, ClassifiedComponent design/selection

    The Synchro SELECT Guidewires are a steerable guidewire family with a shapeable tip and are available in straight and pre-shaped versions. The outside diameters of the guidewires are 0.014in. The guidewires are compatible with existing microcatheters used in common procedures such as those used in endovascular diagnosis and therapy of neurovascular disease. The distal portion of the guidewire tip is radiopaque. For lubricity, the distal portion of the device is coated with a hydrophilic polymer and the proximal portion of the guidewire is coated with polytetrafluoroethylene (PTFE). The torque device included with the guidewire attaches to the proximal end of the wire and functions as a steering guide. Rotation of this device facilitates guidewire placement into the appropriate vessel by precise directional manipulation of the guidewire tip. The introducer included with the guidewire is intended to aid insertion of the guidewire into the catheter hub and/or hemostasis valve.

    Stryker Neurovascular has observed an increased frequency of PTFE coating damage occurring on the Synchro Guidewires that may be caused by the practice of backloading the guidewire through the optional introducer accessory. This issue is limited to certain lots of the Synchro Guidewire that contain an older version of the introducer accessory. Due to variation in the manufacturing process of the supplier of the introducer accessory, certain lots of introducers have sharper than intended edges that can peel off the PTFE coating when physicians use a technique known as backloading. Users with impacted product in their inventory are cautioned not to use this backloading technique.

  • TerminatedUnder Investigation by firm

    MicroVention Terumo Traxcess 7 Mini Guidewire with Hydrophilic Coating, Ref GW0721006M, 1 guidewire per package, sterile, RX.

    There is a potential for the hydrophilic coating on the guidewire to delaminate from the guidewire.