Processor, Cervical Cytology Slide, Automated
FDA adverse event reports and recall history for processor, cervical cytology slide, automated (product code MKQ), from the public MAUDE corpus and the live FDA recall database.
- Total reports
- 411
- Deaths
- 0
- Injuries
- 40
- Malfunctions
- 243
Reports per year
Source: FDA adverse event reports by year received, 2011–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.
Recent recalls
- 2013-08-28TerminatedSoftware design
PrepStain AG. For use in the screening and detection of cervical cancer.
PrepStain system rack setting was incorrectly entered on one PrepStain system during a routine preventive maintenance visit by a BD Field Service Representative.
- 2013-08-28TerminatedSoftware design
PrepStain Tecan US-I. For use in the screening and detection of cervical cancer.
PrepStain system rack setting was incorrectly entered on one PrepStain system during a routine preventive maintenance visit by a BD Field Service Representative.
- 2013-08-28TerminatedSoftware design
PrepStain Tecan US-II. For use in the screening and detection of cervical cancer.
PrepStain system rack setting was incorrectly entered on one PrepStain system during a routine preventive maintenance visit by a BD Field Service Representative.
- 2012-07-26TerminatedNonconforming Material/Component
PrepStain Kit: 1.2 DiTi Cone. The PrepStain System is a liquid-based thin layer cell preparation process. The PrepStain System produces SurePath Slides that are intended for use in the screening and detection of cervical cancer, pre-cancerous lesions, atypical cells and all other cytologic categories as defined by The Bethesda System for Reporting Cervical/Vaginal Cytology Diagnoses. The 1.2 DiTi is the tip on the PrepStain to which a pipette tip attaches and dispenses patient samples onto a glass microscope slide.
Some Prep Stain Kit, 1.2 DiTi Cones have been found to be leaking and the PrepStain Preparation kits contain incorrect preventive maintenance (PM) instructions.
- 2012-07-26TerminatedNonconforming Material/Component
PrepStain Upgrade Kit: 1.3.0.3/1.2 DiTi Cone. The PrepStain System is a liquid-based thin layer cell preparation process. The PrepStain System produces SurePath Slides that are intended for use in the screening and detection of cervical cancer, pre-cancerous lesions, atypical cells and all other cytologic categories as defined by The Bethesda System for Reporting Cervical/Vaginal Cytology Diagnoses. The 1.2 DiTi is the tip on the PrepStain to which a pipette tip attaches and dispenses patient samples onto a glass microscope slide.
Some Prep Stain Kit, 1.2 DiTi Cones have been found to be leaking and the PrepStain Preparation kits contain incorrect preventive maintenance (PM) instructions.
- 2012-07-26TerminatedNonconforming Material/Component
PrepStain Slide Processor. The PrepStain System is a liquid-based thin layer cell preparation process. The PrepStain System produces SurePath Slides that are intended for use in the screening and detection of cervical cancer, pre-cancerous lesions, atypical cells and all other cytologic categories as defined by The Bethesda System for Reporting Cervical/Vaginal Cytology Diagnoses. The 1.2 DiTi is the tip on the PrepStain to which a pipette tip attaches and dispenses patient samples onto a glass microscope slide.
Some Prep Stain Kit, 1.2 DiTi Cones have been found to be leaking and the PrepStain Preparation kits contain incorrect preventive maintenance (PM) instructions.
- 2012-07-26TerminatedNonconforming Material/Component
PrepStain PM Kit, T-AG (Catalog 490588); PrepStain PM Kit Europe (Catalog 490671); PrepStain Prep Only PM Kit, T-AG (Catalog 490954); PrepStain PM Kit XLP, T-US-II (Catalog 490424); PrepStain Prep PM Kit XLP T-US-II (Catalog 491118); PrepStain Prep Only PM Kit XLP, T-US-II (490423); PrepStain Prep Only PM Kit, T-US (Catalog 490429); PrepStain PM Kit, T-US (Catalog 490626); PrepStain PM Kit Europe, T-US (Catalog 490428). The PrepStain System is a liquid-based thin layer cell preparation process. The PrepStain System produces SurePath Slides that are intended for use in the screening and detection of cervical cancer, pre-cancerous lesions, atypical cells and all other cytologic categories as defined by The Bethesda System for Reporting Cervical/Vaginal Cytology Diagnoses. The 1.2 DiTi is the tip on the PrepStain to which a pipette tip attaches and dispenses patient samples onto a glass microscope slide.
Some Prep Stain Kit, 1.2 DiTi Cones have been found to be leaking and the PrepStain Preparation kits contain incorrect preventive maintenance (PM) instructions.
- 2012-06-15TerminatedComponent design/selection
PrepStain Syringing Pipettes 240 (Catalog #490517)
The PrepStain Syringing Pipettes may not dispense the minimum volume of 7.5 mL of the sample due to leaking. This event resulted in an increase in complaints of leaking syringing pipettes
- 2012-06-15TerminatedComponent design/selection
PrepStain GYN 480 Disposable Test Kit with PreCoat Slides (Catalog #490523
The PrepStain Syringing Pipettes may not dispense the minimum volume of 7.5 mL of the sample due to leaking. This event resulted in an increase in complaints of leaking syringing pipettes
- 2012-06-15TerminatedComponent design/selection
PrepStain Installation Kit (PreCoat), Catalog/Model # 490528
The PrepStain Syringing Pipettes may not dispense the minimum volume of 7.5 mL of the sample due to leaking. This event resulted in an increase in complaints of leaking syringing pipettes
Monitor processor, cervical cytology slide, automated automatically