Sterilant, Medical Devices
FDA adverse event reports and recall history for sterilant, medical devices (product code MED), from the public MAUDE corpus and the live FDA recall database.
- Total reports
- 1,871
- Deaths
- 3
- Injuries
- 425
- Malfunctions
- 562
Reports per year
Source: FDA adverse event reports by year received, 2012–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.
Recent recalls
- 2022-08-26Open, ClassifiedEnvironmental control
STERIS CORP Revital-Ox Resert Hydrogen Peroxide High-Level Disinfectant Revital-Ox RESERT RTU Liquid 4 Liter Container Max 21 Day Reuse Model Number: 4455AW
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
- 2022-08-26Open, ClassifiedEnvironmental control
a. McKesson DISINFECTANT, GLUTARALDEHYDE 28DAY (4GL/CS) MGM68, Model Number: 68-102800. b. METREX RESEARCH McKesson 14-Day Glutaraldehyde Solution DISINFECTANT, GLUTARALDEHYDE 14DAY (4GL/CS) MGM68 Model Number: 68-101400. c. METREX RESEARCH McKesson SOLUTION, GLUT 2.5% 28DAY QT (16/CS) PSS744 Catalog # 341. d. METREX RESEARCH LLC MetriCide DISINFECTANT, METRICIDE 14DAYGLUTARALDEHYDE 1GL (4/CS) Catalog # 10-1400. e. METREX RESEARCH LLC MetriCide28 DISINFECTANT, METRICIDE GLUT LNG LIFE GL (4GL/CS) Catalog # 10-2800
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
- 2019-07-05TerminatedError in labeling
Regimen 28 Day - Glutaraldehyde Sterilizing and Disinfecting Solution
Immersion time stated on the bottle is 45 minutes while the product instructions say 90 minutes.
- 2016-09-02TerminatedNonconforming Material/Component
Acecide-C High Level Disinfectant & Sterilant 875 mL. Packaging is 6 inner cartons in one master carton;1 inner carton consists of a set of two liquids (1 acid & 1 buffer) banded together.
Certain lots of the acid solution in Acecide-C are degrading faster than normal.
- 2015-09-16TerminatedUnder Investigation by firm
CIDEX OPA Solution, P/N 20390 for use as a high level disinfectant for reprocessing heat sensitive medical devices.
Advanced Sterilization Products (ASP) is reminding all CIDEX OPA Solution customers to follow the warnings and rinsing procedure in their current labelling around the criticality of proper rinsing of medical device following disinfection with CIDEX OPA according to the CIDEX OPA Solution IFU (Instruction For Use).
- 2015-08-25TerminatedDevice Design
CIDEX Activated Dialdehyde Solution, 4 bottles/case, P/N: 2266
Advanced Sterilization Products (ASP) is recalling the CIDEX Activated Dialdehyde Solution because it may lack the adequate data to support the entire 24-month shelf life.
- 2013-06-26TerminatedLabeling Change Control
Quick Connect QLC1677E, QRC1699E, QPC1721E and QFC1683E. STERIS Corporation, Mentor, OH 44060 QLC1677E, QRC1699E, QPC1721E and QFC1683E contain processing instructions, a Quick Reference Wall Chart and a flow unit. The flow unit consists of a Quick Reference Card, tubing, tethers and adapters for connecting specified devices to Flexible Processing Trays/Containers.
During a routine label review, the firm identified four SYSTEM 1E Quick Connects that will be updated to reflect the removal of obsolete device models. This labeling review process also highlighted the opportunity to combine two existing quick connects into a single product while eliminating the obsolete models.
- 2012-09-07TerminatedNonconforming Material/Component
The SYSTEM 1E Liquid Sterilant Processing System. The SYSTEM 1E Liquid Sterilant Processing System is intended for the liquid chemical sterilization of manually cleaned immersible and reusable critical and semi-critical heat sensitive medical devices, including endoscopes and their accessories.
The firm initiated a recall after complaint investigations revealed that when the units are left on and not in use at night and/or on weekends the hose disconnects/separates resulting in water leakage that can cause damage to property.
- 2012-08-28TerminatedDevice Design
QKC1692E The QKC1692E Quick Connect device connects endoscopes to the sterilization tray for C1220E liquid chemical sterilization processor.
STERIS has been actively seeking devices for quick connect revalidation projects. STERIS was unable to obtain two (2) of the needed devices; these two devices represent the design of seven (7) models in one quick connect. Due to the unavailability of these two validation devices, STERIS will remove 7 models from the labeling of QKC1692E and will re-label all QKC1692Es in the field..
- 2012-08-27TerminatedProcess control
CIDEX OPA, High Level Disinfectant Solution, Glutaraldehyde-free (0.55% ortho-phthaladehyde) solution. The CIDEX OPA solution is intended for automated and manual cleaning and disinfection.
The recall was initiated because Advanced Sterilization Products has discovered instances of missing labels on bottles of CIDEX OPA Solution, Part # 20390 manufactured between May 2011 and April 2012.
Monitor sterilant, medical devices automatically
DeviceWatch watches FDA adverse events and recalls for the products you care about, summarizes clinical narratives with AI and alerts you when a safety signal emerges.