LYJClass 3

Stimulator, Autonomic Nerve, Implanted For Epilepsy

FDA adverse event reports and recall history for stimulator, autonomic nerve, implanted for epilepsy (product code LYJ), from the public MAUDE corpus and the live FDA recall database.

Total reports
52,136
Deaths
2,928
Injuries
18,507
Malfunctions
30,578

Reports per year

12
13
14
15
16
17
18
19
20
21
22
23
24
25
26

Source: FDA adverse event reports by year received, 2012–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.

Recent recalls

  • 2025-01-31Open, ClassifiedDevice Design

    SenTiva, REF: 1000; SenTiva Duo, REF: 1000-D; Generators Part of the VNS Therapy System. Used for Vagus Nerve Stimulation (VNS).

    Vagus nerve stimulator generators may stop delivering stimulation due to an internal, mechanically activated component that may become stuck in a closed position, which may result in patients returning to baseline seizure frequency or depressive symptoms and in the longer term, patients may be required to undergo generator replacement surgery to replace their device.

  • 2025-01-31Open, ClassifiedDevice Design

    SenTiva, REF: 1000C, Generators Part of the VNS Therapy System with Microburst Stimulation. Used for Vagus Nerve Stimulation (VNS).

    Vagus nerve stimulator generators may stop delivering stimulation due to an internal, mechanically activated component that may become stuck in a closed position, which may result in patients returning to baseline seizure frequency or depressive symptoms and in the longer term, patients may be required to undergo generator replacement surgery to replace their device.

  • 2025-01-30Open, ClassifiedDevice Design

    Brand Name: SenTiva" Product Name: VNS Therapy" SenTiva", Model 1000 Model/Catalog Number: 1000 Software Version: NA Product Description: The LivaNova VNS Therapy" System, used for Vagus Nerve Stimulation (VNS), consists of an implantable VNS Therapy generator, lead, and external programming system used to change stimulation settings. The generator is an implantable, multi-programmable pulse generator that delivers electrical signals to the vagus nerve through the lead. The generator is housed in a hermetically sealed titanium case and is powered by a single battery. Electrical signals are transmitted from the generator to the vagus nerve by the lead. The lead and the generator make up the implantable portion of the VNS Therapy System. The VNS Therapy Programming System includes a computer pre-installed with VNS Therapy programming software and a programming wand. The physician uses the programming system to read and change generator settings. The Model 1000 Generator was commercially approved for distribution in the U.S. on October 3, 2017 via PMA P970003/S210. Component: The LivaNova VNS Therapy" System, used for Vagus Nerve Stimulation (VNS).

    Their is a potential that implanted pulse generators may stop delivering therapy due to an internal component issue.

  • 2025-01-30Open, ClassifiedDevice Design

    Brand Name: SenTiva DUO" Product Name: VNS Therapy SenTiva DUO", Model 1000-D Model/Catalog Number: 1000-D Software Version: NA Product Description: The LivaNova VNS Therapy" System, used for Vagus Nerve Stimulation (VNS), consists of an implantable VNS Therapy generator, lead, and external programming system used to change stimulation settings. The generator is an implantable, multi-programmable pulse generator that delivers electrical signals to the vagus nerve through the lead. The generator is housed in a hermetically sealed titanium case and is powered by a single battery. Electrical signals are transmitted from the generator to the vagus nerve by the lead. The lead and the generator make up the implantable portion of the VNS Therapy System. The VNS Therapy Programming System includes a computer pre-installed with VNS Therapy programming software and a programming wand. The physician uses the programming system to read and change generator settings. The Model 1000-D Generator was commercially approved for distribution in the U.S. on May 13, 2020 via PMA P970003/S231. Component: The LivaNova VNS Therapy" System, used for Vagus Nerve Stimulation (VNS).

    Their is a potential that implanted pulse generators may stop delivering therapy due to an internal component issue.

  • 2019-12-20TerminatedDevice Design

    VNS Therapy, SenTiva, Model # 1000, (01)05425025750405, Rx Only for vagal nerve stimulation (VNS) therapy

    Sentiva generators may reset and become disabled within 60 days of enabling. Physicians can re-enable but devices will continue to be susceptible to resets. If a device experiences this issue, this may lead to additional surgery (premature replacement of generator), or patients may return to baseline seizure frequency or depressive symptoms due to cessation of intended stimulation.

  • 2016-01-15TerminatedDevice Design

    VNS Therapy AspireSR Generator Model 106. Indicated for use as an adjunctive therapy in reducing the frequency of seizures.

    Recall being initiated in response to three reports of "Burst Watchdog Timeout" events occurring with the Model 106 AspireSR Generator, resulting in a device reset condition where stimulation output is disabled.

  • 2016-01-13TerminatedManufacturing material removal

    Cyberonics Vagus Nerve Stimulator VNS Therapy AspireSR Generator, Model 106

    Certain Model 106 Pulse Generators demonstrate delays in sensing during use of the 'Verify Heartbeat Detection' feature and exhibit the potential for decreased battery longevity.

  • 2015-11-17TerminatedManufacturing material removal

    Cyberonics Vagus Nerve Stimulator (VNS) Therapy AspireSR Generator, Model 106

    Certain Model 106 Pulse Generators demonstrate delays in sensing during use of the 'Verify Heartbeat Detection' feature and exhibit potential for decreased battery longevity.

  • 2015-04-27TerminatedNonconforming Material/Component

    VNS Therapy DemiPulse Generator Model 103 and VNS Therapy DemiPulse Duo Generator Model 104; CYBERONICS, INC.

    The pulse generators have a lower battery longevity than specified in their design requirement as a result of the devices being inadvertently left in a programmed ON state during manufacture.

  • 2014-12-23TerminatedLabeling mix-ups

    Cyberonics VNS Therapy AspireHC, Model 105 Generator. The Cyberonics VNS Therapy System consists of the implantable VNS Therapy Generator, the VNS Therapy Lead, and the external programming system used to change the stimulation settings. Indicated for use as an adjunctive therapy in reducing the frequency of seizures.

    The recalled product was distributed with an incorrect serial number printed on the device's label.

Monitor stimulator, autonomic nerve, implanted for epilepsy automatically