LXTClass 2

Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component, Metal Composite

FDA adverse event reports and recall history for appliance, fixation, nail/blade/plate combination, multiple component, metal composite (product code LXT), from the public MAUDE corpus and the live FDA recall database.

Total reports
781
Deaths
0
Injuries
474
Malfunctions
304

Reports per year

12
13
14
15
16
17
18
19
20
21
22
23
24
25
26

Source: FDA adverse event reports by year received, 2012–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.

Recent recalls

  • 2022-06-28TerminatedProcess control

    DVR Anatomic, Model No. DVRAR. Distal Volar Radius Anatomical Plate System used for the fixation of fractures and osteotomies involving the distal radius.

    One lot of DVRASL plates were incorrectly etched and may have been placed into kits as DVRAR or DVRASR plates

  • 2016-03-08TerminatedPackage design/selection

    Crosslock DVR ePAK -Sterile, Ref 81180050, are complete surgical sets including plates, screws, and instruments for orthopedic fixation The Cross lock DVR ePAK Volar Rim Plates system is intended for fixation of fractures, mal unions, and osteotomies involving the distal radius. The DVR Wrist ePAK system is intended stabilization and fixation of small bone fragments in fresh fractures, revision procedures, joint fusion and reconstructions of small bones of the hand, foot, radius, ulna, ankle, humerus, scapula and pelvis, particularly in osteopenic bone.

    Components of the DVR ePAKs can oxidize during shipment and storage prior to use. A delay in surgery, less than 30 minutes, may occur if the surgical staff proceeds to re-sterilize the product. If oxidative material would flake off a component contained within the ePAK and enter the surgical field/wound an inflammatory reaction may occur. The inflammatory reaction may result in a foreign body reaction and/or metal hypersensitivity. Oxidized components have failed cytotoxicity testing.

  • 2015-04-14TerminatedNonconforming Material/Component

    LOW PROFILE NON-LOCK SCREW; 2.7mm x 22 mm; REF 131227222; LOT RM105E. Provides the orthopaedic surgeon a means of bone fixation and helps generally in the management of fractures and reconstructive surgeries.

    A Biomet investigation found that the Low Profile Non-Locking Screw 2.7mmx22mm screw (Part Number: 131227222 Lot: RM105E) was anodized green and should have been anodized gold per the print.

  • 2009-06-30TerminatedProcess control

    TransFx External Fixation System Drill Bit, Quick Connect, 2.0 mm diameter, 125 mm length, nonsterile, Zimmer, Warsaw, IN; REF 4450-54. TransFx drill bits can be found in the following kits: TransFx Intermediate External Fixation System, REF 4450-05-10; TransFx Large External Fixation System, REF 4450-010-10 and TransFx Small External Fixation System, REF 4450-00050-15. For use in pre-drilling bone to accept a fixation pin, which are part of the external fixation structure that the surgeon assembles intra-operatively to stabilize bone fractures from the extremities.

    Drill tip is out of specifications, reducing the efficiency of drilling action, requiring more force and potentially heating the bone surface during operation.

  • 2009-06-30TerminatedProcess control

    TransFx External Fixation System Drill Bit, Quick Connect, 2.5 mm diameter, 180 mm length, nonsterile, Zimmer, Warsaw, IN; REF 4450-56. TransFx drill bits can be found in the following kits: TransFx Intermediate External Fixation System, REF 4450-05-10; TransFx Large External Fixation System, REF 4450-010-10 and TransFx Small External Fixation System, REF 4450-00050-15. For use in pre-drilling bone to accept a fixation pin, which are part of the external fixation structure that the surgeon assembles intra-operatively to stabilize bone fractures from the extremities.

    Drill tip is out of specifications, reducing the efficiency of drilling action, requiring more force and potentially heating the bone surface during operation.

  • 2009-06-30TerminatedProcess control

    TransFx External Fixation System Drill Bit, Quick Connect, 3.0 mm diameter, 180 mm length, nonsterile, Zimmer, Warsaw, IN; REF 4450-58. TransFx drill bits can be found in the following kits: TransFx Intermediate External Fixation System, REF 4450-05-10; TransFx Large External Fixation System, REF 4450-010-10 and TransFx Small External Fixation System, REF 4450-00050-15. For use in pre-drilling bone to accept a fixation pin, which are part of the external fixation structure that the surgeon assembles intra-operatively to stabilize bone fractures from the extremities.

    Drill tip is out of specifications, reducing the efficiency of drilling action, requiring more force and potentially heating the bone surface during operation.

  • 2009-06-30TerminatedProcess control

    TransFx External Fixation System Drill Bit, Quick Connect, 3.5 mm diameter, 195 mm length, nonsterile, Zimmer, Warsaw, IN; REF 4450-61. TransFx drill bits can be found in the following kits: TransFx Intermediate External Fixation System, REF 4450-05-10; TransFx Large External Fixation System, REF 4450-010-10 and TransFx Small External Fixation System, REF 4450-00050-15. For use in pre-drilling bone to accept a fixation pin, which are part of the external fixation structure that the surgeon assembles intra-operatively to stabilize bone fractures from the extremities.

    Drill tip is out of specifications, reducing the efficiency of drilling action, requiring more force and potentially heating the bone surface during operation.

  • 2009-06-30TerminatedProcess control

    TransFx External Fixation System Drill Bit, Quick Connect, 4.0 mm diameter, 195 mm length, nonsterile, Zimmer, Warsaw, IN; REF 4450-63. TransFx drill bits can be found in the following kits: TransFx Intermediate External Fixation System, REF 4450-05-10; TransFx Large External Fixation System, REF 4450-010-10 and TransFx Small External Fixation System, REF 4450-00050-15. For use in pre-drilling bone to accept a fixation pin, which are part of the external fixation structure that the surgeon assembles intra-operatively to stabilize bone fractures from the extremities.

    Drill tip is out of specifications, reducing the efficiency of drilling action, requiring more force and potentially heating the bone surface during operation.

  • 2009-06-30TerminatedProcess control

    TransFx External Fixation System Drill Bit, Quick Connect, 4.5 mm diameter, 230 mm length, nonsterile, Zimmer, Warsaw, IN; REF 4450-65. TransFx drill bits can be found in the following kits: TransFx Intermediate External Fixation System, REF 4450-05-10; TransFx Large External Fixation System, REF 4450-010-10 and TransFx Small External Fixation System, REF 4450-00050-15. For use in pre-drilling bone to accept a fixation pin, which are part of the external fixation structure that the surgeon assembles intra-operatively to stabilize bone fractures from the extremities.

    Drill tip is out of specifications, reducing the efficiency of drilling action, requiring more force and potentially heating the bone surface during operation.

  • 2009-06-30TerminatedProcess control

    TransFx External Fixation System Drill Bit, Quick Connect, 4.0 mm diameter, 230 mm length, nonsterile, Zimmer, Warsaw, IN; REF 4450-67. TransFx drill bits can be found in the following kits: TransFx Intermediate External Fixation System, REF 4450-05-10; TransFx Large External Fixation System, REF 4450-010-10 and TransFx Small External Fixation System, REF 4450-00050-15. For use in pre-drilling bone to accept a fixation pin, which are part of the external fixation structure that the surgeon assembles intra-operatively to stabilize bone fractures from the extremities.

    Drill tip is out of specifications, reducing the efficiency of drilling action, requiring more force and potentially heating the bone surface during operation.

Monitor appliance, fixation, nail/blade/plate combination, multiple component, metal composite automatically