Skip to main content
LSLClass 2

Dna-Reagents, Neisseria

FDA adverse event reports and recall history for dna-reagents, neisseria (product code LSL), from the public MAUDE/AEMS corpus and the live FDA recall database.

Total reports
190
Deaths
0
Injuries
89
Malfunctions
95

Reports per year

11
13
14
15
16
17
18
19
20
21
22
23
24
25
26

Source: FDA adverse event reports by year received, 20112026. Reporting paused during the 2026 MAUDE-to-AEMS transition and has since resumed.

Recent recalls

  • 2015-08-20TerminatedIncorrect or no expiration date

    Xpert CT /NG Vaginal/Endocervical/ Specimen Collection Kit: Part number CT/NGSWAB-50; Microbiology: The Cepheid X pert CT /NG Vagina Endocervical/ Specimen Collection Kit is designed to collect, preserve and transport patient Chlamydia trachomatis and Neisseria gonorrhoeae DNA in endocervical and vaginal specimens from symptomatic and asymptomatic individuals prior to analysis with the Cepheid Xpert CT/NG Assay.

    Firm determined that some lots of CT/NG Swab 50 and CT/NG Urine collection kits contain a sub component with an expiry date that is earlier than the kit expiration date.

  • 2015-08-20TerminatedIncorrect or no expiration date

    Xpert CT/NG Urine Specimen Collection Kit Part number GXCT/NGURINE-50; Microbiology: Xpert CT/NG Urine Specimen Collection Kit is designed to preserve and transport Chlamydia trachomatis and Neisseria gonorrhoeae DNA in first-catch male and female urine specimens from symptomatic and asymptomatic individuals prior to analysis with the Cepheid Xpert CT/NG Assay.

    Firm determined that some lots of CT/NG Swab 50 and CT/NG Urine collection kits contain a sub component with an expiry date that is earlier than the kit expiration date.

  • 2005-02-03TerminatedOther

    BD ProbeTec ET Urine Processing Kit, 4x25, catalog 440454, packaged in a carton containing 4 storage bags identified as BD ProbeTec ET Urine Processing Pouches, 25 pouches/bag. Carton and bags labeled in part ***Becton, Dickinson and Company, 7 Loveton Circle, Sparks, MD 21152***

    An in-vitro diagnostic kit reagent may cause false negative clinical results in patients tested for sexually transmitted infections.

  • 2004-08-19TerminatedOther

    Amplicor [3] AV-HRP. (Avidin-HRP (Horseradish Peroxidase) BGG Conjugate.) Kit: Amplicor CT/NG NG Detection Kit

    An increased frequency of 'blue foci' that potentially can cause elevated A450 background in microwell plate wells after the addition of conjugate reagent during PCR detection.

  • 2004-08-19TerminatedOther

    Amplicor [3] AV-HRP, Avidin-HRP Horseradish Peroxidase) BGG Conjugate. Kit: AMPLICOR NG Detection Kit, Export IVD/CE.

    An increased frequency of 'blue foci' that potentially can cause elevated A450 background in microwell plate wells after the addition of conjugate reagent during PCR detection.

  • 2003-08-20TerminatedOther

    Becton Dickinson's ProbeTec (tm) ET Instrument

    Component of in vitro diagnostic device was incorrectly installed causing incorrect reporting of patient results.

  • TerminatedEquipment maintenance

    Xpert CT/NG Catalog GXCT/NG-120 Batch 1000049568, Catalog GXCT/NG-10 Batch 1000053075, Catalog GXCT/NG-10 Batch 1000056156 and Catalog GXCT/NG-10 Batch 1000056157. Is a qualitative in vitro real-time PCR test.

    Cepheid has received reports that some of our customers have experienced a higher number of temperature drift errors than would typically be expected when using this test and its intended use specimen types. Investigation of the issue also revealed an occasional probability of false negative results which remains consistent with package insert claims.

Monitor dna-reagents, neisseria automatically

DeviceWatch watches FDA adverse events and recalls for the products you care about, summarizes clinical narratives with AI and alerts you when a safety signal emerges.