System, Test, Automated, Antimicrobial Susceptibility, Short Incubation
FDA adverse event reports and recall history for system, test, automated, antimicrobial susceptibility, short incubation (product code LON), from the public MAUDE corpus and the live FDA recall database.
- Total reports
- 4,002
- Deaths
- 6
- Injuries
- 71
- Malfunctions
- 3,923
Reports per year
Source: FDA adverse event reports by year received, 2012–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.
Recent recalls
- 2026-10-09Open, ClassifiedUnder Investigation by firm
BD Phoenix Panel NMIC-312. Catalog Number: 449082.
Potential for quality control failures with Klebsiella pneumoniae ATCC BAA-1705 when using the CPO Ambler classification feature.
- 2026-10-09Open, ClassifiedUnder Investigation by firm
BD Phoenix Panel NMIC-306. Catalog Number: 449292.
Potential for quality control failures with Klebsiella pneumoniae ATCC BAA-1705 when using the CPO Ambler classification feature.
- 2026-09-18Open, ClassifiedPackaging process control
VITEK 2 AST-N461 TEST KIT, GRAM NEGATIVE SUSCEPTIBILITY TESTING - AUTOMATED, Reference Number 424811
The issue is related to card pouch damage associated with a portion of the lot which was produced on a specific card pouching device, in a specific timeframe, causing a potential risk of false resistant results. False resistant results may lead to the use of less effective antibiotic treatments or the association of multiple antibiotics having negative side effects, which could potentially impact patient s health.
- 2026-08-05Open, ClassifiedSoftware Design Change
REF # D23352, MicroScan LabPro V5.03 Panel Update 11 (US), Electronic installed on the following instruments: - B1018-280 MicroScan autoSCAN-4 Instrument - B1018-283 MicroScan WalkAway-40 plus Instrument - B1018-284 MicroScan WalkAway-96 plus Instrument - B1018-440 DxM-1040 MicroScan WalkAway Instrument - B1018-496 DxM-1096 MicroScan WalkAway Instrument
Analyzers contain incorrect algorithms which may affect antimicrobial susceptibility test results, leading to false susceptibility classifications."
- 2026-08-05Open, ClassifiedSoftware Design Change
REF # D23354, MicroScan LabPro V5.03 Panel Update 11 DVD Kit (US) on the following instruments: - B1018-280 MicroScan autoSCAN-4 Instrument - B1018-283 MicroScan WalkAway-40 plus Instrument - B1018-284 MicroScan WalkAway-96 plus Instrument - B1018-440 DxM-1040 MicroScan WalkAway Instrument - B1018-496 DxM-1096 MicroScan WalkAway Instrument
Analyzers contain incorrect algorithms which may affect antimicrobial susceptibility test results, leading to false susceptibility classifications."
- 2026-07-09Open, ClassifiedProcess change control
VITEK 2 AST-N447 TEST KIT Product reference 424620
Potential for false resistant carbapenem and/or cephalosporin results when analyzing samples with kits from affected lot that have pouch foil fractures.
- 2025-10-23Open, ClassifiedSoftware design
BD Phoenix M50 Instrument, labeled as the following: 1. BD Phoenix M50 Automated Microbiology System, Catalog Numbers: 443624, 44362409.
Product service credentials used by some BD technical support teams to access certain BD products were accessed by an unauthorized actor. Until these product service credentials are updated, there is a risk of unauthorized access that may impact the confidentiality, integrity and/or availability of the relevant products and associated data.
- 2025-09-29Open, ClassifiedUnder Investigation by firm
VITEK 2 Gram-negative AST Test Kits (Cards) containing Colistin formulation cs02n. VITEK 2 is an automated system consisting of instruments, software and reagent cards designed for the identification and antimicrobial susceptibility testing of bacteria and yeast. Export only
Potential for false resistant results for Colistin (cs02n) with multi-drug-resistant Pseudomonas aeruginosa and Acinetobacter baumannii complex used in specific cards
- 2024-12-03Open, ClassifiedSoftware design
BD Phoenix M50 Automated Microbiology System Version or Model: 443624 Catalog Number: 443624
BD confirmed that product service credentials used by some BD technical support teams to access certain BD products were accessed by an unauthorized actor. Until these product service credentials are updated, there is a risk of unauthorized access that may impact the confidentiality, integrity and/or availability of the relevant products and associated data.
- 2024-09-24Open, ClassifiedUnder Investigation by firm
BD Phoenix M50 instrument -Intended for the rapid identification (ID) and antimicrobial susceptibility testing (AST) of clinically significant bacteria. The BD Phoenix System provides rapid results for most aerobic and facultative anaerobic Grampositive bacteria. Catalog Number: 443624
Escherichia coli (E. coli) misidentification for clinical samples that were tested on BD Phoenix" M50 instrument and BD Phoenix" 100, may lead to misdiagnosis and inappropriate treatment of infections caused by E. coli,
Monitor system, test, automated, antimicrobial susceptibility, short incubation automatically