FDA adverse event reports and recall history for product code lnx (product code LNX), from the public MAUDE corpus and the live FDA recall database.

Total reports
590
Deaths
16
Injuries
109
Malfunctions
380

Reports per year

09
10
11
12
13
14
15
16
17
18
19
20
21
22
23

Source: FDA adverse event reports by year received, 2009–2023. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.

Recent recalls

  • 2014-01-22TerminatedSoftware design

    ORCHESTRA/ORCHESTRA PLUS Programmer

    Sorin has voluntarily issued a notification to physicians related to the overestimation of the residual longevity displayed by its programmer (ORCHESTRA or ORCHESTRA PLUS) during a follow-up exam of patients implanted with a REPLY or ESPRIT pacemaker.

  • 2013-12-18TerminatedUnknown/Undetermined by firm

    ABACUS TPN Calculation Software, Baxa: 8300-0045 ABACUS CE (Calculator Edition), 8300-0046 ABACUS SE (Single-Workstation Edition), 8300-0047 ABACUS ME (Multi-Workstation Edition) ABACUS is a Windows - based order entry TPN software application for comprehensive calculations and label printing. The primary use of Abacus is for TPN order calculations. Other uses of Abacus include: non-TPN volumes administered to patients on TPN therapy in order calculations; CVVH calculations; Cardioplegia (Buckburg) calculations; Epidural calculations; Dilutions from concentrated ingredient; Combinations of multiple ingredients; Fractional doses from larger volumes

    Baxter Healthcare Corporation is issuing a voluntary correction for ABACUS Total parenteral Nutrition (TPN) Calculation Software versions 2.0 through 3.0 due to the possibility that a patient could receive another patient's TPN order as a result of data corruption.

  • 2012-05-25TerminatedLabeling design

    Abacus TPN Calculating Software Product Usage: LIMS Windows-based program for order entry and TPN calculation.

    Ordering salt-based ingredients on an ion-based template resulted in medication errors when calculating amounts of ingredients required.

  • 2011-12-02TerminatedDevice Design

    SoftPath GUI Versions: 4.3.0.10, 4.3.0.11, and 4.3.0.12. SoftPath is a laboratory information system to be used in a clinical laboratory by knowledgeable, trained, and experienced personnel. SoftPath enables the user to document steps in the receiving and processing of surgical, gynecological, medical cytological, and autopsy case specimens within the Pathology labs, for the purpose of generating Pathology reports for requesting physicians and label creation for identification of specimen/case material.

    On 06/13/2011, SCC Soft Computer initiated a correction on the following SoftPath GUI Versions: 4.3.0.10, 4.3.0.11, and 4.3.0.12. A client reported when they verified the contents of the fax reports, they found one of the specimen diagnoses to be missing. Parts A through D were present as well as Part F, however, Part E diagnosis was missing for one of the patient reports. All affected clients wer

  • 2011-12-02TerminatedSoftware design

    SoftReports version 1.1.8.2.4 with Hot Fixes 1.7199.7 or 1.7200.7. SoftReports is a web-based report launching tool to be used in a hospital or laboratory environment by knowledgeable, trained, and experienced personnel. SoftReports enables the user to schedule and generate reports for the purpose of viewing and printing data stored in SCC product databases.

    On 10/25/2011 SCC Soft Computer, Clearwater, FL initiated a correction on the SoftReports version 1.1.8.2.4 with Hot Fixes 1.7199.7 or 1.7200.7. When SoftReports Designer opened a layout, the system did not load the layout with the previously saved value in the NewPage property. This caused multiple patient records to be printed on the same page. All affected clients were notified on the issue on

  • 2011-11-23TerminatedSoftware design

    SoftLab GUI versions 4.0.2.8-4.0.2.10; 4.0.3.12-4.0.3.13; and 4.0.4.1-4.0.4.4. SoftLab is a laboratory information system to be used in a medical research or clinical laboratory by knowledgeable, trained, and experienced personnel. SoftLab enables the user to handle data-processing and laboratory activities, for the purpose of storing, managing, querying, and report laboratory data.

    On 05/03/2011 SCC Soft Computer, Clearwater, FL initiated a correction on the following SoftLab GUI versions 4.0.2.8-4.0.2.10; 4.0.3.12-4.0.3.13; and 4.0.4.1-4.0.4.4. A client reported the PT and INR results did not get flagged. The PROT should have been flagged as HIGH and the INR should have been flagged as PANIC. Neither test was flagged. All affected clients were notified of the issue on 05/03

  • 2011-11-23TerminatedSoftware change control

    SCC Soft Computer, Softlab Laboratory information System with DORA versions: 2.0.0.4 with Hot Fix 1.2808, 1.2809, 1.2956, or 1.3476; 2.0.0.5; 2.0.0.6; 2.0.0.7 with Hot Fix 1.6595; and 2.1.0.3.

    Potential for patient data to be sent to the wrong patient record. A software change affected how the Softlab handles recovery from certain types of table access conflicts. When such conflicts occur, this function has the potential to cause faulty writing to the database.

  • 2011-06-30TerminatedSoftware design

    SoftReports versions 1.1.6.x and 1.1.7.x SoftReports version 1.1.6 released 2/2008; Version 1.1.7 released 6/2008. SCC Soft Computer 5400 Tech Data Drive Clearwater, FL 33760. SoftReports is a report designer and web-based report launching tool to be used by knowledgeable, trained, and experienced personnel. It provides the ability to create ad hoc, user defined queries, and reports using data from other SCC products. SoftReports enables users to customize standard reports generated by SCC systems to meet their specific needs. This system performs the following functions: 1. The ability to customize report layouts using beginner, intermediate, or advanced options. 2. The ability to import and export reports. 3. The ability to provide custom desktops to launch SQL's within other SCC applications. 4. The ability to preview and modify layouts specific to users' needs. 5. The ability to modify elements of reports using the comprehensive control formatting toolbar. 6. The ability to incorporate charts and graphics into reports. 7. The ability to modify existing query templates. 8. The ability to create independent and embedded queries. 9. The ability to pars SQL statements and notify users of errors within the syntax. 10. The ability to easily update criteria parameters to indicate input needed for executing queries. 11. The ability to view and print any report that is created using the Designer. 12. The ability to schedule, execute, and track ad-hoc reports. This system does not perform the following functions: 1. SoftReports is not configured to write back to the database. 2. SoftReports does not support user-created Ad Hoc reports being used for diagnostic purposes. 3. SoftReports is not intended to alter the meaning of patient result reports (test reports).

    SCC Soft Computer is issuing a correction for SoftReports Versions 1.1.6.x released 02/2008 and 1.1.7.x released 06/2008. Potential hazard - when using SoftReports Layout Designer, results displayed in the report may not match the patient information in the report header.

  • 2010-11-15TerminatedSoftware design

    SoftID.NET versions 2.0.0.6 - 2.0.0.19. SCC Soft Computer 5400 Tech Data Drive Clearwater, FL 33760. SoftID.NET is to be used in a medical or clinical laboratory setting by knowledgeable, trained, and experienced professionals. SoftID.NET supports bedside collection of patient specimens used in diagnostic testing. This product is integrated with Soft Computer's SoftLab, SoftScape and can also interface with all other SCC laboratory systems and Hospital Information Systems (HIS).

    SCC Soft Computer issued a Correction for SoftID.NET versions 2.0.0.6 - 2.0.0.19. When using the Cancel functionality, if the collection list refreshes during the transaction and changes the position of the patient on the list, the wrong patient order can be cancelled. All affected clients have been notified.

  • 2010-10-20TerminatedSoftware design

    SoftWeb version 4.0.4.10.1. SCC Soft Computer, 5400 Tech Data Drive, Clearwater, FL 33760. Manufacture Date: 05/16/2006.

    If a case number is re-used in SoftWeb with SoftPath, the SoftPath case is linked to the original MRN in SoftWeb. This causes SoftPath cases being linked to the wrong patient in SoftWeb. Incorrect results on the patient with original MRN and no results on the patient with the second MRN (and the reused case number) are displayed in SoftWeb Results Query. SCC Soft Computer issued a Correction for