System, Photopheresis, Extracorporeal
FDA adverse event reports and recall history for system, photopheresis, extracorporeal (product code LNR), from the public MAUDE corpus and the live FDA recall database.
- Total reports
- 2,974
- Deaths
- 37
- Injuries
- 317
- Malfunctions
- 2,573
Reports per year
Source: FDA adverse event reports by year received, 2012–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.
Recent recalls
- 2026-04-16Open, ClassifiedProcess control
Brand Name: Cellex Photopheresis System Product Name: CELLEX Photopheresis Kit Model/Catalog Number: CLXUSA Software Version: Not Applicable Product Description: THERAKOS Photopheresis or extracorporeal photopheresis (ECP) is a photoimmune therapy where leukocytes are separated from whole blood via apheresis, combined with a photoactive drug (8-methoxypsoralen) and then exposed to ultraviolet A (UVA) light. All blood components, including the treated leukocytes, are returned to the patient. THERAKOS Photopheresis utilizes the THERAKOS CELLEX System to combine cell separation and photoactivation into a single, closed and sterile circuit. The THERAKOS CELLEX Photopheresis System collects the buffy coat (leukocyte-enriched blood) from the patient in a discontinuous flow process and intermittently returns the remaining plasma and erythrocytes to the patient. The buffy coat is passed through the photoactivation module where the drug is activated with a precise amount of UVA light determined by the characteristics of the individual patient s buffy coat. After photoactivation, the buffy coat is immediately returned to the patient s bloodstream. Component: Not Applicable
Increased difficulty in installing the centrifuge bowl onto the centrifuge bowl holder. Improper installation may cause the centrifuge bowl to dislodge, resulting in a broken bowl during the prime cycle or treatment phase, leading to a delay in patient treatment and blood loss.
- 2012-01-11TerminatedOther
Therakos Cellex Photopheresis System CellexUSA & Cellex. The photopheresis system is indicated for use in the ultraviolet-A (UVA) irradiation in the presence of the photoactive drug 8-methoxypsoralen (8-MOP) of extracorporeally circulating leukocyte-enriched blood in the palliative treatment of the skin manifestation of cutaneous T-cell lymphoma (CTCL) in person who have not been responsive to other forms of treatment. The Cellex procedural kits are designed to interface with the Cellex Photopheresis System to perform cell separation and photo activation in a single closed and sterile circuit. This is accomplished by use of an integrated centrifuge bowl and associated tubing for separation of whole blood into red blood cells, plasma, and buffy coat (leukocyte-enriched blood). All blood components are returned to the patient. The procedural kits are fully assembled as received by the customer and the customer is only required to place the components in the Cellex System to enable the system for use.
There have been three confirmed reports of post reinfusion hemolysis or hemoglobinuria in patients treated on the Therakos Cellex Photopheresis System.
- 2011-08-10TerminatedDevice Design
XT125 PHOTOPHERESIS PROCEDURAL KIT FOR USE WITH THE UVAR XTS INSTRUMENT; XT001 PHOTOPHERESIS PROCEDURAL KIT FOR USE WITH THE UVAR XTS INSTRUMENT Single-use, disposable, closed system with sterile fluid pathways. The device collects the buffy coat (leukocyte-enriched blood) from the patient in a discontinuous flow process and intermittenly returns the remaining plasma and erythocytes to the patient.
Increased complaints were received of tubing leaks at the junction between the tubing and Fluid Logic Module plastic ports of the kits.
- 2011-07-19TerminatedEmployee error
Therakos Cellex Procedural Kits. Designed to interface with the Cellex Photopheresis System to perform cell separation and photoactivation in a single, closed and sterile circuit.
Cellex Procedural kits may have minor tears in their Tyvek covers.
- 2011-04-28TerminatedProcess control
Therakos UVAR XTS Procedural Kits Is indicated for use in the ultraviolet-A (UVA) irradiation.
Therakos received an increased rate of complaints of centrifuge bowl leaks for certain lots of Therakos UVAR XTS Procedural Kits (Product Code XT 125).
- 2010-05-26TerminatedNonconforming Material/Component
Therakos CellEx Photopheresis System; Photopheresis Procedural Kits, for use with the THERAKOS CellEx Instrument;
Higher than normal blood leaks on THERAKOS CELLEX Procedural Kits that are intended for use with the THERAKOS CELLEX Photopheresis System
- 2006-09-14TerminatedOther
Therakos Photopheresis Procedural Kit for use with the UVAR XTS Instrument. (catalog number XT125)
Centrifuge Bowl may leak
- Open, ClassifiedUnder Investigation by firm
THERAKOS CELLEX Photopheresis Procedural Kit PRODUCT CODE: CLXUSA
Product was released prior to all testing being completed. If used and a drive tube leak were to occur, may cause illness or injury.
- Open, ClassifiedUnder Investigation by firm
THERAKOS CELLEX Photopheresis Procedural Kit, CLXUSA
This recall was initiated to recover a Dunnage Case, which is a case of three (3) CLXUSA kits. This Dunnage Case was shipped on 02-Dec-2021. Dunnage is used during the sterilization process and is not intended for customer use.
- TerminatedNonconforming Material/Component
The THERAKOS CELLEX Photopheresis System is a photoimmune therapy. Catalog CLXUSA. Manufacturing lot D342 UDI: 20705030200003 The system designed to separate the patient's blood into various fractions and collect the white blood cell (WBC) fraction. The WBC fraction is treated with a photoactive drug UVADEX (methoxsalen) Sterile Solution. The UVADEX is photo-activated by UVA light and results in the WBCs undergoing apoptosis
The Mallinckrodt (Therakos) post-market monitoring process detected an increased trend in tubing leaks within the pump tubing organizer, associated with a portion of Kits from Lot D 342.
Monitor system, photopheresis, extracorporeal automatically