FDA adverse event reports and recall history for implant, dermal, for aesthetic use (product code LMH), from the public MAUDE corpus and the live FDA recall database.

Total reports
23,036
Deaths
14
Injuries
19,307
Malfunctions
1,645

Reports per year

12
13
14
15
16
17
18
19
20
21
22
23
24
25
26

Source: FDA adverse event reports by year received, 2012–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.

Recent recalls

  • 2016-08-19TerminatedNonconforming Material/Component

    Radiesse (+) Lidocaine, Injectable Implant 1.5cc, Rx only, P/N: 8063M0K1, Merz Aesthetics. Product Usage: RADIESSE is indicated for sub-dermal implantation for the correction of moderate to severe facial wrinkles and folds.

    Complaints reported of Radiesse (+) unable to be expelled from the syringe and/or needle.

  • 2015-12-03TerminatedOther

    Radiesse (+) Lidocaine Injectable Implant 1.5cc, P/N: 8063M0K1. Indicated for subdermal implantation for the correction of moderate to severe facial wrinkles and folds.

    The expiration date encoded in the UDI barcode on the label on the product carton and the peelable patient identification label is incorrect.

  • 2015-03-30TerminatedStorage

    Vitagel; indicated in surgical procedures (other than in neurosurgical and ophthalmic) as an adjunct to hemostasis when control of bleeding by ligature or conventional procedures is ineffective or impractical.

    Several shipments of Vitagel product (a surgical hemostat) required to be stored at 2 -8 degrees Celsius (C) were not delivered to customers within the validated delivery time.

  • 2014-08-08TerminatedDevice Design

    Enhancement Medical expression. model # EXP-150. Injectable Gel 1.5cc Syringe with hyaluronic acid concentration of 20 mg/ml. Product Usage: Expression is labelled as an Intranasal Splint, intended to minimize bleeding and edema and to prevent adhesions between the septum and the nasal cavity. However, Expression is currently being used off-label as a subcutaneously injected dermal filler.

    Manufacturer made a change in the production process that resulted in a change in final gel weight. RECALL EXPANDED 7/8/2014 Firm expanded their recall to include all lots of product.

  • 2011-11-01TerminatedNonconforming Material/Component

    MERZ AESTHETICS, Radiesse Volume Advantage 1.5 CC, Injectable Implant, Part Number 8071M0H1 and 8071M0K1, Sterile, Rx ONLY. Part Number 8071M4K1 distributed in Canada Part Number 8071M5 distributed in Europe/Middle East/Africa Part Number 8071M9 distributed in Hong Kong, Korea, Malaysia.

    There is a potential leakage of product that may occur with Merz Aesthetics 1.5cc syringes of Radiesse Dermal Filler. When using the syringe, Health Care Professionals should be alerted to the potential for product that my leak pass the plunger at the proximal end of the syring. The defect may result in infection requiring medical intervention to prevent impairment or damage. To date Merz has n

  • 2011-05-19TerminatedEmployee error

    Radiesse, Injectable Implant - 1.3cc Syringe, Sterile, Sterilized Using Steam, Use Before 2013-01, Ref 8046M7, Lot 1024417, Manufactured by: BioForm Medical Inc, 4133 Courtney Rd #10 Franksville, WI 53126. Subdermal implantation for the correction of moderate to severe facial wrinkles and folds, such as masolabial folds and it is also intended for restoration and/or correction of the signs of facial fat loss (lipoatrophy) in people with human immunodeficiency virus.

    Merz Aesthetics, Inc. is recalling 1 lot (1034417) of Radiesse 1.3 cc Implant. The lot was supposed to be scrapped because it had failed to meet its Loss On Drying (LOD) specification. Unfortunately it had been marked as approved, released and shipped.

  • TerminatedOther

    Revanesse Versa, PN40081

    Labeling error. The product is labeled with an 18 month expiration date, however the product is approved for 12 months.