Pump, Infusion, Implanted, Programmable
FDA adverse event reports and recall history for pump, infusion, implanted, programmable (product code LKK), from the public MAUDE corpus and the live FDA recall database.
- Total reports
- 117,815
- Deaths
- 1,639
- Injuries
- 75,258
- Malfunctions
- 38,593
Reports per year
Source: FDA adverse event reports by year received, 2012–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.
Recent recalls
- 2026-06-25CompletedProcess control
Intera Oncology INTERA 3000 Hepatic Artery Infusion Pump, AP-0300H, implantable drug delivery device
There is the potential for leakage along the drug pathway from the pump through the end of the catheter.
- 2026-05-14Open, ClassifiedUnder Investigation by firm
Medtronic SynchroMed Flex Infusion Mode - A810 Clinician Programmer Software Application, loaded on the following devices: 1. PROG CT900A CLINICIAN TABLET US, Model Number CT900A; 2. PROG CT900B CLINICIAN TABLET EU -UK, Model Number CT900B; 3. PROG CT900C CLINICIAN TABLET UK, Model Number CT900C; 4. PROG CT900D CLINICIAN TABLET GLOBAL, Model Number CT900D; 5. PROG CT900E CLINICIAN TABLET GLBL SMSNG, Model Number CT900E; 6. PROG CT900F CLINICIAN TABLET GLBL SMSNG, Model Number CT900F; implanted programmable infusion pump
Software issue: When programming the SynchroMed Pump with the A810 CP App in Flex Infusion Mode, the infusion steps should automatically be arranged in consecutive order, from earliest to latest. However, in rare cases, the steps may be displayed out of order. If this occurs, the programmed Flex Infusion schedule will not be delivered at the intended time, if at all.
- 2025-04-01Open, ClassifiedUnder Investigation by firm
A820 myPTM Software Application associated with Medtronic SynchroMed Pump and Infusion System.
Product complaints were received describing the A820 myPTM app taking longer than expected for patients to interact with their implantable pump.
- 2024-12-27Open, ClassifiedUnder Investigation by firm
SynchroMed II Model 8637 and SynchroMed III Model 8667 implantable pumps used with PROG CT900B CLINICIAN TABLET EU -UK with the A810 CP app version 2.x, CFN CT900B
Software issues were identified in application version 2.x.
- 2024-12-27Open, ClassifiedUnder Investigation by firm
SynchroMed II Model 8637 and SynchroMed III Model 8667 implantable pumps used with PROG CT900C CLINICIAN TABLET UK with the A810 CP app version 2.x, CFN CT900C
Software issues were identified in application version 2.x.
- 2024-12-27Open, ClassifiedUnder Investigation by firm
SynchroMed II Model 8637 and SynchroMed III Model 8667 implantable pumps used with PROG CT900D CLINICIAN TABLET GLOBAL with the A810 CP app version 2.x, CFN CT900D
Software issues were identified in application version 2.x.
- 2024-12-27Open, ClassifiedUnder Investigation by firm
SynchroMed II Model 8637 and SynchroMed III Model 8667 implantable pumps used with PROG CT900E CLINICIAN TABLET GLBL SMSNG and PROG CT900E CLINICIAN TABLET PAPER EN with the A810 CP app version 2.x, CFN CT900E
Software issues were identified in application version 2.x.
- 2024-10-21Open, ClassifiedDevice Design
SynchroMed II Infusion Pump. Product Number: 8637-20
Potential for error and inability to interrogate pump due to data corruption in the pump memory.
- 2024-07-12Open, ClassifiedDevice Design
Medtronic Ascenda Intrathecal Catheter, Models: a) 8780 (114 cm); b) 8781 (140 cm); c) 8784 (catheter pump revision kit)
Design update to the Ascenda Intrathecal Catheter Models 8780, 8781, and 8784. The intent of the design update is to reduce the potential for tissue growth into the Ascenda catheter connector which may potentially lead to catheter occlusion.
- 2023-12-22Open, ClassifiedUnder Investigation by firm
Medtronic SynchroMed II, Model 8637-20, Programmable pump
Medtronic recently identified that if the SynchroMed II pump switches into telemetry mode due to electromagnetic interference (EMI) from an MRI scan, while the pump is sounding an alarm, the pump will not resume drug delivery after leaving the MRI magnetic field, which is inconsistent with the current labeling.
Monitor pump, infusion, implanted, programmable automatically
DeviceWatch watches FDA adverse events and recalls for the products you care about, summarizes clinical narratives with AI and alerts you when a safety signal emerges.