Cannula, Manipulator/Injector, Uterine
FDA adverse event reports and recall history for cannula, manipulator/injector, uterine (product code LKF), from the public MAUDE corpus and the live FDA recall database.
- Total reports
- 1,804
- Deaths
- 2
- Injuries
- 248
- Malfunctions
- 1,484
Reports per year
Source: FDA adverse event reports by year received, 2012–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.
Recent recalls
- 2023-12-06Open, ClassifiedComponent design/selection
Cooper Surgical H/S ELLIPTOSPHERE CATH 5 Fr -For administering contrast media during Hysterosalpingography or Hysterosonography procedures to detect uterine pathology such as polyps, fibroids, adhesions or endometrial thickening, and/or patency of fallopian tubes. Model: 61-3005
Incorrect bonding material used for assembly of the catheter, may fail, resulting in detachment of component parts, broken components, or leakage. This breakage may result in delay of procedure
- 2023-12-06Open, ClassifiedComponent design/selection
Cooper Surgical H/S ELLIPTOSPHERE CATH 5 FR Set-For administering contrast media during Hysterosalpingography or Hysterosonography procedures to detect uterine pathology such as polyps, fibroids, adhesions or endometrial thickening, and/or patency of fallopian tubes Model: 61-4005
Incorrect bonding material used for assembly of the catheter, may fail, resulting in detachment of component parts, broken components, or leakage. This breakage may result in delay of procedure
- 2023-12-06Open, ClassifiedComponent design/selection
CooperSurgical H/S CATHETER SET FR 5 with Integated Stylet Model: 61-4005S For administering contrast media during Hysterosalpingography or Hysterosonography procedures to detect uterine pathology such as polyps, fibroids, adhesions or endometrial thickening, and/or patency of fallopian tubes
Incorrect bonding material used for assembly of the catheter, may fail, resulting in detachment of component parts, broken components, or leakage. This breakage may result in delay of procedure
- 2022-08-26Open, ClassifiedEnvironmental control
Mckesson GYNECATH H/S CATHETER 5FR (10/CS) Model Number: 11-19610
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
- 2015-06-08TerminatedDevice Design
Ally Uterine Positioning System (Ally UPS) used in the mounting and positioning and holding of uterine manipulators during gynecological laproscopic procedures Model: Ally UPS
Design of device may expose user to injury to fingers or body parts
- 2014-09-05TerminatedPackaging process control
CooperSurgical H/S CATHETER (PART # 61-5007) Intended for Hysterosonography/ Hysterosalpingography.
Unsealed pouches.
- 2014-06-11TerminatedProcess control
Cooper Surgical TAMPA CATHETER 5 French 33 cm. Intended for Hysterosonography. Model Number: 61-2005
Sterility of the device may be compromised due to unsealed pouch
- 2012-11-26TerminatedDevice Design
Cooper Surgical RUMI Arch, Reusable Uterine Manipulator Handle Ref: UMH700 The intended use of the RUMI Arch is in an operative endoscopy (laparoscopy) where a uterus is present and where positioning of the uterus, fallopian tubes and ovaries is desirable. These types of surgeries include laparoscopic tubal ligation, and/or operative laparoscopy
Excessive pressure exerted on the uterine tip when attached to the Arch may, under certain circumstances, cause the internal steel rod of the tip to protrude through the cleaning hole on the shaft of the handle
- 2012-11-26TerminatedDevice Design
Cooper Surgical Advincula Arch , Reusable Uterine Manipulator Handle Ref: UMH750 The intended use of the RUMI Arch is in an operative endoscopy (laparoscopy) where a uterus is present and where positioning of the uterus, fallopian tubes and ovaries is desirable. These types of surgeries include laparoscopic tubal ligation, and/or operative laparoscopy
Excessive pressure exerted on the uterine tip when attached to the Arch may, under certain circumstances, cause the internal steel rod of the tip to protrude through the cleaning hole on the shaft of the handle
- 2008-09-02TerminatedProcess control
ClearView Uterine Manipulator, Long tip version with disposable Sound-Dilator (DSD-1000), REF CV2000, 9 cm, Sterile, Clinical Innovations, 747 West 4170 South, Murray, Utah 84123.
Inadequate package seal could compromise sterility.
Monitor cannula, manipulator/injector, uterine automatically
DeviceWatch watches FDA adverse events and recalls for the products you care about, summarizes clinical narratives with AI and alerts you when a safety signal emerges.