Catheter, Nephrostomy
FDA adverse event reports and recall history for catheter, nephrostomy (product code LJE), from the public MAUDE corpus and the live FDA recall database.
- Total reports
- 821
- Deaths
- 8
- Injuries
- 261
- Malfunctions
- 543
Reports per year
Source: FDA adverse event reports by year received, 2012–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.
Recent recalls
- 2025-11-20Open, ClassifiedProcess control
Brand Name: EZDilate Nephrostomy Balloon Catheter and Inflation Device Kit Product Name: Olympus Nephro-EZDilate Nephrostomy Balloon Catheter Model/Catalog Number: BPCN0815K Product Description: The Nephro-EZDilate nephrostomy balloon catheter is a reinforced catheter attached to a distal dilatation balloon. It has a radiopaque tip and two radiopaque markers positioned on the catheter inside the balloon that define the working length
Device kits contained an incorrectly sized 4mm x 4cm urinary tract balloon catheter (BURS0404) instead of the correct 8mm x 15cm nephrostomy tract balloon catheter (BPCN0815). All other components in the kits were correctly included and packaged in the kit.
- 2025-09-05Open, ClassifiedUnder Investigation by firm
Boston Scientific NephroMax Kit, Material Numbers (UPN): 1. M0062101180; 2. M0062101600. (Urology)
The Encore 26 Inflation Device is used with balloon dilation catheters to create and monitor pressure in the balloon and to deflate the balloon. An internal investigation determined that during use, foreign material particles could migrate from affected Encore 26 Inflation Devices into a balloon dilatation catheter.
- 2016-03-24TerminatedLabeling mix-ups
UltraxxTM Nephrostomy Balloon Set The Ultraxx" Nephrostomy Balloon Set consists of a radiopaque balloon catheter, radiopaque sheath, and inflation device. The Ultraxx" Nephrostomy Balloon Set is used to dilate the musculofascia, renal capsule, and parenchyma to establish and maintain a percutaneous tract.
Cook Medical is voluntarily recalling specific lots of the Ultraxx" Nephrostomy Balloon Set and the Ascend AQ Ureteral Dilation Balloon Catheter Set because of a labeling mix-up. Cook Medical has received reports of packages labeled as Ultraxx" Nephrostomy Balloon Set actually containing the Ascend AQ Ureteral Dilation Balloon Catheter Set and vice versa.
- 2016-03-24TerminatedLabeling mix-ups
Ascend AQ Ureteral Dilation Balloon Catheter Set The Ascend AQ Ureteral Dilation Balloon Catheter Set consists of a radiopaque balloon catheter and inflation device. The Ascend AQ Ureteral Dilation Balloon Catheter Set is used for transluminal dilation of ureteral strictures or ureteral dilation prior to ureteroscopy and stone manipulation.
Cook Medical is voluntarily recalling specific lots of the Ultraxx" Nephrostomy Balloon Set and the Ascend AQ Ureteral Dilation Balloon Catheter Set because of a labeling mix-up. Cook Medical has received reports of packages labeled as Ultraxx" Nephrostomy Balloon Set actually containing the Ascend AQ Ureteral Dilation Balloon Catheter Set and vice versa.
- 2013-07-09TerminatedProcess control
NephroMax high Pressure Nephrostomy Balloon Catheter. Material Number/Catalog Number: M0062101400/210-140, M062101440/210144 Dilation of the nephrostomy tract.
Incorrect sized renal sheath was packaged with the NephroMax high pressure nephrostomy Balloon Catheter.
- 2013-07-09TerminatedProcess control
NephroMax high Pressure nephrostomy Balloon Catheter Kit. Material Number/Catalog Number: M0062101600/210-160, M0062101640/210-164 Dilation of the nephrostomy tract.
Incorrect sized renal sheath was packaged with the NephroMax high pressure nephrostomy Balloon Catheter.
- 2007-06-05TerminatedOther
Boston Scientific PCNL (Percutaneous Tract) Kit with Nephromax Balloon; Part Number: M0064501110
Mislabeled: Outer box unit has correct part number but product is incorrectly identiified as a Percutaneous Tract Kit with Nephromax Balloon. The box actually contains a Percutaneous Tract Kit with Amplatz Dilator and Sheath Kit
- TerminatedNo Marketing Application
Percutaneous Neonatal Pigtail Nephrostomy Set, Reference Part Number 080106, Order Number G14508. The set contains a 22 gage, 5 cm Chiba Needle; an 18 gage, 5 cm trocar needle; an .038-inch (.97 mm) diameter, 30 cm PTFE-coated stainless steel; a 5, 6, and 7 French, 10 cm radiopaque dilator; a 6 French, 7 cm radiopaque pigtail catheter; a retention disc with pull tie; and a 10 French, 10 cm clear, polyvinylchloride connecting tube with one-way stopcock.
The indwell time for the Percutaneous Neonatal Pigtail Nephrostomy Set and Pediatric Nephrostomy Stent Set is decreasing from 4 months to 4 weeks. The current IFU is in process of being updated and will be provided with orders placed following implementation.
- TerminatedNo Marketing Application
Pediatric Nephrostomy Stent Set, Reference Part Numbers 080208-S6 and 080208-S7, Order Numbers G15034 and G15035. The set contains a radiopaque pigtail catheter and stent, retention disc with pull tie, and a 10 French, 30 cm polyvinylchloride connecting tube with one-way stopcock.
The indwell time for the Percutaneous Neonatal Pigtail Nephrostomy Set and Pediatric Nephrostomy Stent Set is decreasing from 4 months to 4 weeks. The current IFU is in process of being updated and will be provided with orders placed following implementation.
- TerminatedUnknown/Undetermined by firm
X-Force Nephrostomy Balloon Dilation Catheter Kit with Inflation Device, PTFE Sheath, 8 mm x 15 cm, Product Code 996081 The product is a dual lumen catheter with a 24 (8mm) or 30Fr (10mm) balloon mounted on the distal tip. It has a radiopaque tip and radiopaque marker beneath the balloon. It is recommended for use in the dilation of the nephrostomy tact and for placement of the working sheath.
Contract manufacturer packaged an incorrect sheath with the product. Packaged 30Fr instead of 24Fr.
Monitor catheter, nephrostomy automatically