FDA adverse event reports and recall history for handpiece, rotary bone cutting (product code KMW), from the public MAUDE corpus and the live FDA recall database.

Total reports
309
Deaths
1
Injuries
267
Malfunctions
41

Reports per year

09
11
14
15
16
17
18
19
20
21
22
23
24
25
26

Source: FDA adverse event reports by year received, 2009–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.

Recent recalls

  • 2024-02-20Open, ClassifiedLabeling False and Misleading

    W&H Surgical handpiece S-12. For Surgical treatment of dental hard tissue Model Number: 30061000

    1:2 speed increasing surgical handpieces have been incorrectly laser marked during production. The specific laser marking should read 1:2, but instead it reads 1:1 on the body of the handpiece

  • 2018-10-24TerminatedDevice Design

    Surgical Handpiece, Model: SGS-E, Order Code: H084, Nakanishi Inc. The SGS/SGA Surgical Handpieces are intended for surgical use. Typical usage includes dental implant use: drilling and tapping the maxilla and mandible.

    Reprocessing instructions for the SGS/SGA Surgical Handpieces have been updated with instructions of manual cleaning and specific parameters for autoclave sterilization . The previous operation manuals associated with the SGS/SGA products must be replaced by new operation manuals.

  • 2018-10-24TerminatedDevice Design

    Surgical Handpiece, Model: SGS-E2G Order Code: H185, Nakanishi Inc. The SGS/SGA Surgical Handpieces are intended for surgical use. Typical usage includes dental implant use: drilling and tapping the maxilla and mandible.

    Reprocessing instructions for the SGS/SGA Surgical Handpieces have been updated with instructions of manual cleaning and specific parameters for autoclave sterilization . The previous operation manuals associated with the SGS/SGA products must be replaced by new operation manuals.

  • 2018-10-24TerminatedDevice Design

    Surgical Handpiece, Model: SGA-E Order Code: H083, Nakanishi Inc. The SGS/SGA Surgical Handpieces are intended for surgical use. Typical usage includes dental implant use: drilling and tapping the maxilla and mandible.

    Reprocessing instructions for the SGS/SGA Surgical Handpieces have been updated with instructions of manual cleaning and specific parameters for autoclave sterilization . The previous operation manuals associated with the SGS/SGA products must be replaced by new operation manuals.

  • 2018-10-24TerminatedDevice Design

    Surgical Handpiece, Model: SGA-E2G Order Code: H184, Nakanishi Inc. The SGS/SGA Surgical Handpieces are intended for surgical use. Typical usage includes dental implant use: drilling and tapping the maxilla and mandible.

    Reprocessing instructions for the SGS/SGA Surgical Handpieces have been updated with instructions of manual cleaning and specific parameters for autoclave sterilization. The previous operation manuals associated with the SGS/SGA products must be replaced by new operation manuals.

  • 2018-10-24TerminatedDevice Design

    Micro Surgery Handpiece SGS, Model: SGS-ES, Order Code: H264, Manufactured by: Nakanishi Inc. The SGS/SGA Surgical Handpieces are intended for surgical use. Typical usage includes dental implant use: drilling and tapping the maxilla and mandible.

    Reprocessing instructions for the SGS/SGA Surgical Handpieces have been updated with instructions of manual cleaning and specific parameters for autoclave sterilization. The previous operation manuals associated with the SGS/SGA products must be replaced by new operation manuals.

  • 2018-10-24TerminatedDevice Design

    Micro Surgery Handpiece SGS, Model: SGS-E2S, Order Code: H266, Manufactured by: Nakanishi Inc The SGS/SGA Surgical Handpieces are intended for surgical use. Typical usage includes dental implant use: drilling and tapping the maxilla and mandible..

    Reprocessing instructions for the SGS/SGA Surgical Handpieces have been updated with instructions of manual cleaning and specific parameters for autoclave sterilization. The previous operation manuals associated with the SGS/SGA products must be replaced by new operation manuals.

  • 2018-10-24TerminatedDevice Design

    Micro Surgery Handpiece SGA, Model SGA-ES, Order Code:H263, Manufactured by: Nakanishi Inc. The SGS/SGA Surgical Handpieces are intended for surgical use. Typical usage includes dental implant use: drilling and tapping the maxilla and mandible.

    Reprocessing instructions for the SGS/SGA Surgical Handpieces have been updated with instructions of manual cleaning and specific parameters for autoclave sterilization. The previous operation manuals associated with the SGS/SGA products must be replaced by new operation manuals.

  • 2018-10-24TerminatedDevice Design

    Micro Surgery Handpiece SGA, Model SGA-E2S Order Code: H265, Manufactured by: Nakanishi Inc. The SGS/SGA Surgical Handpieces are intended for surgical use. Typical usage includes dental implant use: drilling and tapping the maxilla and mandible.

    Reprocessing instructions for the SGS/SGA Surgical Handpieces have been updated with instructions of manual cleaning and specific parameters for autoclave sterilization. The previous operation manuals associated with the SGS/SGA products must be replaced by new operation manuals.

  • Open, ClassifiedDevice Design

    OsteoMed OsteoPower Power Console: -OsteoMed OsteoPower Power Console 2 REF 450-0021-00 -OsteoMed D-Power Control Console 2 REF 450-0021-01 -Power Control Console 2 REF ECOS-450-0021-00 -OsteoMed OsteoPower Power Console 2i REF 450-0005-00 -OsteoMed D-Power Control Console 2i REF 450-0005-01 -Power Control Console 2i REF ECOS-450-0005-00 OsteoMed Footswitches: -OsteoMed BMF Footswitch REF 450-0390 -D-Bi-Directional Master Footswitch REF 450-0380 -D-Uni-Directional Footswitch II REF 450-0350 -D-Bi-Directional Footswitch II REF 450-0360 -D-Uni-Directional Master Footswitch REF 450-0370

    Due to consoles not in compliance with the latest electrical safety standards IEC 60601 and as a result shocked a patient.

Monitor handpiece, rotary bone cutting automatically