Fluorometer, For Clinical Use
FDA adverse event reports and recall history for fluorometer, for clinical use (product code KHO), from the public MAUDE corpus and the live FDA recall database.
- Total reports
- 4,551
- Deaths
- 0
- Injuries
- 23
- Malfunctions
- 4,513
Reports per year
Source: FDA adverse event reports by year received, 2012–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.
Recent recalls
- 2020-02-11TerminatedSoftware design
AIA-360 Automated Immunoassay Analyzer, Product Code 019945
A display screen software issue on the analyzer causes the display to freeze when display screen is touched at the same time as a command from the instrument firmware, causing the instrument to stop. As a result, the run is aborted and the results are not retrievable.
- 2017-01-27TerminatedComponent change control
Genetic Screening Processor (GSP); PerkinElmer Part Number= 2021-0010 Intended for in vitro quantitative and qualitative determination of analytes in body fluids.
The firm became aware that the ball bushing axels are defective in the GSP instrument. During exposure to humid conditions, there is a risk that the axle material will corrode. Deterioration of the axle can be observed as deposits of corrosion on the defective axles. Ball bushing axles may corrode and corrosion particle may cause contamination in the wells or the corrosion will cause movement issues and possibly spillage and causing contamination of the plate wells.
- 2017-01-27TerminatedComponent change control
Spare Disk Remover Module; Product Number: 61001820 Intended for in vitro quantitative and qualitative determination of analytes in body fluids.
The firm became aware that the ball bushing axels are defective in the GSP instrument. During exposure to humid conditions, there is a risk that the axle material will corrode. Deterioration of the axle can be observed as deposits of corrosion on the defective axles. Ball bushing axles may corrode and corrosion particle may cause contamination in the wells or the corrosion will cause movement issues and possibly spillage and causing contamination of the plate wells.
- 2015-07-02TerminatedDevice Design
Infinite F500, in vitro diagnostic. Product Usage: The Tecan Infinite F500 is a multifunctional microplate reader with injector option. The Infinite F500 provides high performance for the vast majority of todays microplate applications and research and is robotic compatible. The Infinite F500 is intended as a general purpose laboratory instrument for professional use, supporting common microplate conforming to the ANS/SBS standards.
A firmware bug causes incorrect values to be displayed for specific measured luminescence values.
- 2015-07-02TerminatedDevice Design
Infinite M1000 PRO in vitro diagnostic. Product Usage: The Tecan Infinite M1000 PRO is a multifunctional monochromator based microplate reader that provides high performance for the vast majority of todays microplate applications and research. In addition to offering absorbance and fluorescence intensity measurements, the Infinite M1000 PRO can also perform fluorescence polarization and luminescence measurements (including luminescence scans) as well as Amplified Luminescent Proximity Homogeneous Assays (AlphaScreen and AlphaLISA). The Infinite M1000 PRO is also robotic compatible and offers a built-in stacker option as well as a external injector module. The Infinite M1000 PRO is intended as a general purpose laboratory instrument for professional use, supporting common microplates conforming to the ANS/SBS standards.
A firmware bug causes incorrect values to be displayed for specific measured luminescence values.
- 2015-07-02TerminatedDevice Design
Infinite M1000 in vitro diagnostic. Product Usage: The Tecan Infinite M1000 is a multifunctional monochromator based microplate reader that provides high performance for the vast majority of todays microplate applications and research. Besides offering absorbance and fluorescence intensity measurements, the Infinite M1000 allows the measurement of fluorescence polarization and luminescence. The Infinite M1000 is also robotic compatible and offers a built-in stacker option as well as a external injector module. The Infinite M1000 is intended as a general purpose laboratory instrument for professional use, supporting common microplates conforming to the ANS/SBS standards.
A firmware bug causes incorrect values to be displayed for specific measured luminescence values.
- 2015-07-02TerminatedDevice Design
Infinite 200 PRO in vitro diagnostic. Product Usage: The Tecan Infinite 200 PRO is a multifunctional microplate reader with injector option. The Infinite 200 PRO provides high performance for the vast majority of todays microplate applications and research and is robotic compatible. The Infinite 200 PRO has been designed as a general purpose laboratory instrument for professional use, supporting common 6 to 384 well microplates conforming to the ANS/SBS standards.
A firmware bug causes incorrect values to be displayed for specific measured luminescence values.
- 2015-07-02TerminatedDevice Design
Infinite 200 in vitro diagnostic. Product Usage: The Tecan Infinite 200 is a multifunctional microplate reader with injector option. The Infinite 200 provides high performance for the vast majority of todays microplate applications and research and is robotic compatible. The Infinite 200 PRO has been designed as a general purpose laboratory instrument for professional use, supporting common 6 to 384 well microplates conforming to the ANS/SBS standards.
A firmware bug causes incorrect values to be displayed for specific measured luminescence values.
- 2015-01-15TerminatedEmployee error
Infinite 200 The Tecan Infinite 200 is a multifunctional microplate reader with injector option. The Infinite 200 provides high performance for the vast majority of todays microplate applications and research and is robotic compatible. The infinite 200 has been designed as a general purpose laboratory instrument for professional use, supporting common 6 to 384-well microplates conforming to the ANSI/SBS standards.
Incorrect lumi firmware version installed (E.027 instead of V2.00)
- 2015-01-15TerminatedEmployee error
Infinite 200PRO The Tecan Infinite 200 is a multifunctional microplate reader with injector option. The Infinite 200 provides high performance for the vast majority of todays microplate applications and research and is robotic compatible. The infinite 200 has been designed as a general purpose laboratory instrument for professional use, supporting common 6 to 384-well microplates conforming to the ANSI/SBS standards.
Incorrect lumi firmware version installed (E.027 instead of V2.00)
Monitor fluorometer, for clinical use automatically
DeviceWatch watches FDA adverse events and recalls for the products you care about, summarizes clinical narratives with AI and alerts you when a safety signal emerges.