JOQClass 2

Generator, Pulse, Pacemaker, External Programmable (For Electrophysiological Studies Only)

FDA adverse event reports and recall history for generator, pulse, pacemaker, external programmable (for electrophysiological studies only) (product code JOQ), from the public MAUDE corpus and the live FDA recall database.

Total reports
149
Deaths
2
Injuries
46
Malfunctions
73

Reports per year

10
11
12
13
14
17
18
19
20
21
22
23
24
25
26

Source: FDA adverse event reports by year received, 2010–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.

Recent recalls

  • 2009-06-12TerminatedDevice Design

    EP-4 Computerized Cardiac Stimulator; SJM Part Numbers: 09-1527-0000 / 09-1527-0002; EP MedSystems, Inc. 575 Route 73 North, Units D, West Berlin, NJ 08091. Indicated for use during clinical electrophysiology procedures.

    Non-compliance with IEC Standard 60601-1 Clause 57.6 relative to fusing of AC input line. The neutral line is not fused. The hot line in fused. Clause 57.6 calls for both lines to be fused.

  • 2006-05-31TerminatedOther

    EP-4 Computerized Stimulator -- 2 channel and 4 channel. The product consists of a touchscreen portable PC controlling a two channel dedicated EP stimulator. The stimulator is capable of single channel, simultaneous and sequential stimulation at programmed output settings using built-in protocols, and includes an emergency stimulation function.

    Under certain circumstances the device may exhibit the following behaviors: 1)-Initiation of a rapid stimulation train without command, when the device is in use in conjuction with an electrosurgical unit or RF ablation unit-2)-Presence of a DC offset voltage on an output channel-3)-Inability to synchronize with a non-EP MedSystems recording system

  • 2005-04-19TerminatedOther

    Bloom DTU-215-B Programmable Stimulator

    Possibility that after a loss of power or brown-out, the device may transmit unintended stimulation to patients.

Monitor generator, pulse, pacemaker, external programmable (for electrophysiological studies only) automatically