Photometric Method, Magnesium

FDA adverse event reports and recall history for photometric method, magnesium (product code JGJ), from the public MAUDE/AEMS corpus and the live FDA recall database.

Total reports
367
Deaths
0
Injuries
10
Malfunctions
356

Reports per year

12
13
14
15
16
17
18
19
20
21
22
23
24
25
26

Source: FDA adverse event reports by year received, 20122026. Reporting paused during the 2026 MAUDE-to-AEMS transition and has since resumed.

Recent recalls

  • 2024-12-26Open, ClassifiedNonconforming Material/Component

    Magnesium Flex reagent cartridge, 360 tests, Catalog Number: K3057; Siemens Material Number (SMN): 10445158;

    Siemens Healthcare Diagnostics is recalling two lots of their Dimension Vista Total Magnesium (MG) Flex Reagent Cartridges (lots 24057BB and 24064BC) due to the potential for falsely low measurements of magnesium in plasma. Because the error is intermittent, existing quality control procedures may not recognize that a patients results are incorrect, and the results may be reported out

  • 2022-09-26Open, ClassifiedUnder Investigation by firm

    Dimension Magnesium (MG) Flex reagent cartridge Siemens Material Number (SMN)/REF (Catalog Number):10444963/DF57

    Dimension Magnesium (MG) Flex reagent cartridge lots may exhibit imprecision for Quality Control (QC) and produce Abnormal Reaction flags. The issue is not always detected by QC and erroneous results may be produced in the absence of an Abnormal Reaction flag. Based on customer data, imprecision leading to erroneous, unflagged patient results with a bias of -15 to -59% may occur

  • 2015-11-25TerminatedNonconforming Material/Component

    Dimension Vista System (MG) ; Flex reagent cartridge Catalog number K3057, SMN 10445158. For the quantitative measurement of magnesium in human serum, plasma and urine on the Dimension Vista System.

    Erroneous low results on a small number of reagent wells. There is the potential for under-recovery on certain wells for quality control (QC) and patient sample results. The under-recovery ranged from -0.3 mg/dL [0.12 mmol/L] to -1.6 mg/dL [0.66 mmol/L]; equivalent to a maximum negative bias of -88%.

  • 2014-09-24TerminatedOther

    ABX PENTRA Magnesium RTU Reagent, Part No. A11A01646 Product Usage: ABX PENTRA Magnesium RTU reagent with associated calibrators and controls are for quantitative in vitro diagnostic determination of magnesium in human serum and plasma based on a photometric test using xylidyl blue. Magnesium measurements are used in the diagnosis and treatment of hypomagnesemnia (abnormally low plasma levels of magnesium) and hypermagnesemnia (abnormally high plasma levels of magnesium).

    Horiba Medical is recalling the ABX PENTRA Magnesium RTU Reagent because the on-board stability claims has been updated. Horiba has updated the ABX PENTRA Magnesium RTU package inserts by removing the 7-day on-board stability claim and replacing it with a 1-day on-board stability claim.

  • 2013-11-15TerminatedEquipment maintenance

    VITROS Chemistry Products Mg Slides packaged as 300 slides/pack catalog number 8255093 and packaged as 90 slides/pack catalog number 1921204 VITROS Mg Slides quantitatively measure magnesium (Mg) concentration in serum, plasma, and urine using VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System.

    Ortho Clinical Diagnostics is recalling Slide Dispense cartridges used with some VITROS Chemistry Product Slides due to slide dispense issues.

  • 2012-04-17TerminatedNonconforming Material/Component

    Dimension Vista(R) MG Flex(R) Reagent Cartridge The Dimension Vista(R) MG method is an in vitro diagnostic test for the quantitative measurement of magnesium in human serum, plasma and urine on the Dimension Vista(R) System

    Firm has confirmed a negative shift of approximately 0.2mg/dL [0.08 mmol/L] in patient and QC samples when using Vista Magnesium

  • 2004-01-16TerminatedOther

    Abbott Clinical Chemistry AEROSET c8000 Magnesium, List numbers 7D70-01, 7D70-20, 7D70-30.

    Complaints that instructions for use were not clear.

  • TerminatedUnder Investigation by firm

    Randox Mg R1 - Product Usage: For the quantitative in vitro determination of Magnesium in serum, plasma and urine Catalogue Number: MG3880

    Failing to meet the linearity performance claims quoted on the kit inserts-The linearity is reduced by approximately -30% for Serum/ Plasma and -50% for Urine

  • TerminatedUnder Investigation by firm

    Randox Rx Series Mg - Product Usage: For the quantitative in vitro determination of Magnesium in serum, plasma and urine Catalogue Number: MG8326

    Failing to meet the linearity performance claims quoted on the kit inserts-The linearity is reduced by approximately -30% for Serum/ Plasma and -50% for Urine

  • TerminatedUnder Investigation by firm

    AU/DxC AU Magnesium, Catalog/REF Number OSR6189, UDI 15099590010850

    Lipemic interference for the Magnesium serum application failed to meet the performance claim as defined in the IFU. Use may cause a maximum positive bias up to 30.38% in low magnesium patient samples. The impact is only to the clinical interpretation of magnesium results in the presence of lipemia.