Skip to main content
JDQClass 2

Cerclage, Fixation

FDA adverse event reports and recall history for cerclage, fixation (product code JDQ), from the public MAUDE corpus and the live FDA recall database.

Total reports
2,629
Deaths
30
Injuries
1,259
Malfunctions
1,302

Reports per year

12
13
14
15
16
17
18
19
20
21
22
23
24
25
26

Source: FDA adverse event reports by year received, 2012–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.

Recent recalls

  • 2015-11-14TerminatedDevice Design

    Synthes Application Instrument for Sternal ZIPFIX. Indications for use include primary or secondary closure/repair of the sternum following sternotomy or fracture of the sternum to stabilize the sternum and promote fusion.

    Certain lots of the Application Instrument for Sternal ZIPFIX that were previously recalled on August 19, 2015 and subsequently repaired and returned to customers between August 26, 2015 and September 23, 2015. The screws on the repaired instruments have the potential to break or loosen. DePuy Synthes is requesting the return of the repaired device in exchange for a new device.

  • 2015-08-27TerminatedDevice Design

    Synthes Application Instrument for Sternal ZIPFIX; indications for use include primary or secondary closure/repair of the sternum following sternotomy or fracture of the sternum to stabilize the sternum and promote fusion

    The end cap may loosen and detach making the instrument non-functional. No injuries reported.

  • 2015-07-24TerminatedOther

    Medical Device Exchange 3.5mm Threaded Cerclage Pin; Pouch, Tyvek/Mylar, Sterile, Single Use Only; OEM Catalog number 298.838S; EXP (repackager), Synthes (OEM) Cerclage, fixation

    EXP did not register as a medical device establishment, list the devices being recalled, or establish its own quality system and instead relied on the fact that its vendors were registered and had their own quality systems.

  • 2015-07-09TerminatedLabeling Change Control

    Songer Spinal Cable System, Single Cable, Crimp, Bar, Stainless Steel, Sterile. The DePuy Spine Cable System can be utilized anywhere monofilament wire has been previously found to be indicated.

    The box end label is incorrectly labeled as Titanium instead of Stainless Steel on one lot of implantable Songer Spinal Cable System, Single Cable, Crimp, Bar, Stainless Steel, Sterile, distributed by DePuy Spine Inc.

  • 2014-07-17TerminatedEmployee error

    Tritium Sternal Plate System, Screw, 02.7mm X 12mm, 12 Pack, Sterile. The Pioneer Surgical Cable Plate System is used in the stabilization and fixation of fractures of the anterior chest wall including Sternal fixation following Sternotomy and Sternal reconstructive surgical procedures. The screws are packaged as a 12 pack using a double tray packaging configuration.

    Lack of Sterility Assurance

  • 2013-10-30TerminatedLabeling False and Misleading

    Synthes Application Instrument Sternal Zip Flex System intended for use in primary or secondary closure/repair of the sternum following sternotomy or fracture of the sternum to stabilize the sternum and promote fusion

    When using the 1st generation Zip Fix Application Instrument it is possible to cut the implant while the tensioning trigger is being squeezed which could lead to loosening of the implant.

  • 2010-08-10TerminatedProcess control

    CERCLAGE WIRE, 0.6 MM DIA. X 175 MM LENGTH, SST, REF 71163006, QTY: (6), STERILE R, CE 008, 2020/07, Smith & Nephew, Inc., Memphis, TN 38116, EC REP: Smith & Nephew Orthopaedics GmbH, Tuttlingen, Germany Usage: Bone fracture fixation

    Plastic trays which encase these products have burrs on the hinged packaging which can puncture the single barrier sterile pouch, compromising sterility.

  • 2010-08-10TerminatedProcess control

    CERCLAGE WIRE, 0.8 MM DIA. X 200 MM LENGTH, SST, REF 71163008, QTY: (6), STERILE R, CE 008, 2020/07, Smith & Nephew, Inc., Memphis, TN 38116, EC REP: Smith & Nephew Orthopaedics GmbH, Tuttlingen, Germany Usage: Bone fracture fixation

    Plastic trays which encase these products have burrs on the hinged packaging which can puncture the single barrier sterile pouch, compromising sterility.

  • 2008-11-09TerminatedLabeling mix-ups

    Biomet 2.0 MM Stainless Steel (SS) Crimp Sleeve, sterile, use w/2.0 MM S.S. cable # 350800, Biomet Orthopedics, Inc., Warsaw, IN; REF. 350805. Intended for general orthopedic repairs including long bone fractures, bone grafting, reinforcement of bone and reattachment of the greater trochanter.

    The outer package is properly labeled, but the inner package may be labeled as a femoral component.

  • 2008-08-22TerminatedEmployee error

    Osteo-Clage Cable System Package Assembly (01-0001-S). Product labeled in part, "2.0mm x 850 mm CoCr Cable / Sleeve... REF 01-0001-S LOT W59192... Acumed 5885 NW Cornelius Pass Road Hilsboro, OR 97124-9432 ".

    Cables of incorrect diameter were used in product assembly, resulting in the sleeve not being able to bind to the cable after crimping.

Monitor cerclage, fixation automatically

DeviceWatch watches FDA adverse events and recalls for the products you care about, summarizes clinical narratives with AI and alerts you when a safety signal emerges.