Prosthesis, Elbow, Constrained, Cemented
FDA adverse event reports and recall history for prosthesis, elbow, constrained, cemented (product code JDC), from the public MAUDE corpus and the live FDA recall database.
- Total reports
- 3,106
- Deaths
- 13
- Injuries
- 2,888
- Malfunctions
- 199
Reports per year
Source: FDA adverse event reports by year received, 2012–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.
Recent recalls
- 2026-09-16Open, ClassifiedProcess change control
AXLE PIN ASSY LARGE ELBOW COBALT CHROME Catalog No. C-M112-2-500 Orthopaedic implant components manufactured from UHMWPE (polyethylene)
Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.
- 2023-03-16Open, ClassifiedEmployee error
TEMA Axle Small, REF 1590.15.010 (additional information is pending receipt of the firm's product label/labeling)
Due to the incorrect component, that is not cleared to be marketed in the U.S., being included in its elbow prosthesis.
- 2018-03-14TerminatedUnder Investigation by firm
Coonrad/Morrey Elbow Implant System humeral assembly 4IN XSML, Item Number: 32-8105-027-04 (UDI: (01)0088902427 4006(17)220731(10)63694912)
One lot of humeral stem implants do not include the humeral bushing components.
- 2016-09-08TerminatedProcess design
Coonrad/ Morrey Elbow Pin/ Bushing Replacement size Kit Extra Small. This product is sold sterile. The parts are placed into an inner cavity and are sealed using a nitrogen process. The inner cavity is then placed into an outer cavity and heat sealed with outer Tyvek lid on the outer cavity. The patient record label is placed on the Tyvek outerlid. The sealed cavities and package insert is placed into a folding carton, sealed and shrink wrapped. Total elbow prosthesis designed for use with bone cement
Zimmer Biomet is conducting a voluntary medical device recall of a single lot of the Coonrad/ Morrey Elbow Pin/ Bushing Replacement size Kit Extra Small because the lot was packaged with a missing humeral bushing. The missing bushing could result in a surgical delay or additional surgery if a replacement is not available.
- 2014-06-26TerminatedNonconforming Material/Component
DISCOVERY ELBOW prosthesis, 4 MM X 100 MM RIGHT HUMERAL FLANGED I BOND COAT Tl 6AL 4V ALLOY. Elbow joint replacement prostheses are intended for primary and revision joint arthroplasty for use in cemented applications.
Surface finish is different than specified. The implant has 30 grit blast on the exterior of the flange when the print calls for the glass bead blast. The patient may experience pain requiring surgical intervention.
- 2012-12-18TerminatedProcess change control
COONMAD/MORREY TOTAL ELBOW ELBOW COMPONENT REPLACEMENT SET BUSHINGS & PINS FOR USE W/SIZE REGULAR ELBOW ELBOW COMP REPLACEMENT-REG SET STERILE PROD CLASS: 4100 OPERATION:9000 QTY/PKG:1 Product Usage: The ulnar and humeral components are connected by placing the hollow, outer axis pin across the two components and securing It with the solid Internal axis pin. The Inner and outer pin connection Is articulation point of the placement elbow.
Zimmer is initiating a lot specific recall because affected lots could include an extra Small Inner Pin instead of the required Regular Inner Pin in the Coonrad/Morrey Elbow Replacement Set and the New Ulnar Revision Kit .
- 2012-12-18TerminatedProcess change control
Coonrad/Morrey New Ulnar Revision Kit for Use with Non-Interchangeable Regular Humerals STERILE Rx, Tivanium QTY/PKG:1 Product Usage: The ulnar and humeral components are connected by placing the hollow, outer axis pin across the two components and securing It with the solid Internal axis pin. The Inner and outer pin connection Is articulation point of the placement elbow.
Zimmer is initiating a lot specific recall because affected lots could include an extra Small Inner Pin instead of the required Regular Inner Pin in the Coonrad/Morrey Elbow Replacement Set and the New Ulnar Revision Kit .
- 2012-02-21TerminatedProcess change control
Discovery Elbow System Humeral Condyle Set- Hexalobular, REF 114700, Lot 797550 Alloy, Sterile, Biomet Orthopedics Warsaw IN 46581. Intended for primary and revision joint arthroplasty for use in cemented applications.
A Discovery Condyle Kit w/ Hexalobula was opened during an elbow joint replacement surgery, and was found to contain two male condyles.
- 2011-06-24TerminatedPackaging process control
Condyle Kit with Hexalobular, REF 114700 DISCOVERY ELBOW SYSTEM, HUMERAL CONDYLE SET - HEXALOBULAR, CO-CR-MO / YI-GAL-4V ALLOY, IF USED FOR REVISION SURGERY, INSTRUMENTS 414926 & 414923 MAY BE REQUIRED LOT 183300, 1 SET, BIOMET ORTHOPEDICS 56 EAST BELL DRIVE, P.O. BOX 587 WARSAW, IN 46581 USA EXPIRY DATE: 2021-03 Elbow joint replacement prosthesis intended for primary and revision joint arthroplasty for use in cemented applications.
The firm initiated this recall after becoming aware that a Discovery Condyle Kit with Hexalobula contained two male condyles instead of a male and female condyle.
- 2009-04-02TerminatedLabeling design
Zimmer Coonrad/Morrey Total Elbow, Interchangeable Humeral Assembly, extra small, 4 inch length, sterile, Zimmer Inc., Warsaw, IN; REF 32-8105-027-04. The device is utilized for a total elbow arthroplasty surgical procedure.
The surgical technique is inadequate for extra-small sized implants, resulting in the possibility of the provisional becoming stuck in the bone.