Skip to main content
IZOClass 1

Generator, High-Voltage, X-Ray, Diagnostic

FDA adverse event reports and recall history for generator, high-voltage, x-ray, diagnostic (product code IZO), from the public MAUDE corpus and the live FDA recall database.

Total reports
160
Deaths
8
Injuries
33
Malfunctions
95

Reports per year

07
08
09
10
11
12
13
14
17
18
19
20
22
23
26

Source: FDA adverse event reports by year received, 2007–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.

Recent recalls

  • 2011-03-04TerminatedRadiation Control for Health and Safety Act

    BuckyDiagnost general purpose x-ray system with Optimus Generator. General Purpose Diagnostic Radiography.

    Affected systems were determined not to prevent run-on of a selected technique factor if the technique selector and the exposure control were activated simultaneously.

  • 2010-06-25TerminatedProcess control

    Hydradjust IV DR Urological Table Sedecal Generator (p/n 750732 - 64W generator & p/n 750733 80W generator). Mallinckrodt, Inc. Cincinnati, OH 45237. Facilitates digital radiologic and/or fluoroscopic procedures requiring a beam of diagnostic quality radiation and a flat imaging table.

    Covidien received complaints of the Sedecal Generator failing and producing smoke at customer sites. The failure modes are located in different parts of the generator and will lead to the generator becoming inoperative.

  • 2008-10-27TerminatedComponent change control

    Monitor suspensions used with Advantx Legacy, RFX Classical, SFX-90, Model # 46-240838G2, manufactured after March 8, 1985 for Advantx Legacy, RFX and SFX systems. General-purpose radiographic system component used to suspend the system's image monitor.

    GE Healthcare has recently become aware of a potential risk of monitor dropping associated with the monitor suspension of the Advantx system that may impact patient safety. It has been reported that suspensions manufactured after March 8, 1995 did not have the required thread locking agent applied to the setscrew, allowing the setscrew to back out over time, which can lead to the monitor dropping

  • TerminatedUnder Investigation by firm

    Shimadzu X-ray High Voltage Generator Model No. UD150B-40 on SONIALVISION safire II/17 systems with 0.7/1.2JG326D X-Ray Tube Diagnostic x-ray high voltage generator.

    Shimadzu Medical Systems is recalling the Shimadzu X-ray High Voltage Generator due to the possibility of oil leaking from the x-ray tube housing under certain conditions.

Monitor generator, high-voltage, x-ray, diagnostic automatically

DeviceWatch watches FDA adverse events and recalls for the products you care about, summarizes clinical narratives with AI and alerts you when a safety signal emerges.