Skip to main content
IEOClass 1

Processor, Tissue, Automated

FDA adverse event reports and recall history for processor, tissue, automated (product code IEO), from the public MAUDE corpus and the live FDA recall database.

Total reports
1,666
Deaths
0
Injuries
832
Malfunctions
803

Reports per year

12
13
14
15
16
17
18
19
20
21
22
23
24
25
26

Source: FDA adverse event reports by year received, 2012–2026. Reporting paused twice in 2026 during FDA reporting-system changes and has since resumed.

Recent recalls

  • 2025-04-28Open, ClassifiedRelease of Material/Component prior to receiving test results

    HistoCore PELORIS 3, Model/Catalog Number: 45.0005/45.7512.501 A11, Software Version: 3.4.0. The HistoCore PELORIS 3 Rapid Tissue Processor is a dual retort rapid tissue processor used to prepare tissue samples.

    There is a leakage issue associated with the tubing in the manifold of the instrument.

  • 2024-06-04Open, ClassifiedDevice Design

    Leica HistoCore PEGASUS Tissue Processor, REF 14048858005.

    Poorly processed and/or damaged biopsy tissue specimens on the device resulting from reagent levels exceeding the maximum fill level marks on reagent bottles or in the paraffin tanks.

  • 2023-12-20Open, ClassifiedOther

    Leica HistoCore PEGASUS (PROCESSOR, TISSUE, AUTOMATED)

    There is an issue regarding poorly processed and/or damaged biopsy tissue specimens on the HistoCore PEGASUS / HistoCore PEGASUS Plus resulting from an incorrect assignment of the carryover value set for created or edited protocols.

  • 2015-09-22TerminatedNonconforming Material/Component

    The Leica ASP6025 Tissue Processor. Pathology and Hematology Departments. The Leica ASP6025 Tissue Processor is an automated system used to process tissue specimens for examination through fixation, dehydration and infiltration.

    The ASP6025 Tissue Processor instrument is incorrectly getting into a state of bottle empty during retort filling, after passing the fill lever sensor 2 and prior to reaching level sensor 3, thus not completing a 5 litre fill.

  • 2015-09-17TerminatedNonconforming Material/Component

    The Leica ASP200S/ASP300S Tissue Processor. Pathology and Hematology Departments. The Leica ASP200S/ASP300S Tissue Processor is an automated system used to process tissue specimens for examination through fixation, dehydration and infiltration.

    The Leica ASP200S/ASP300S Tissue Processor have an incorrect internal wiring to the remote alarm plug, which can cause the remote alarm to not function correctly in the case of an instrument failure.

  • 2013-07-29TerminatedDevice Design

    Leica ASP6025 Vacuum Tissue Processor The Leica ASP6025 is a modular tissue processor for the following laboratory applications: fixation, dehydration, infiltration with intermedium, and the paraffin infiltration of histological tissue specimens.

    The Instruction for Use 1v9 Ref F and all previous versions specify incorrect dimensions for the Ready To Use (RTU) bottles from other suppliers than Leica. In case of a pressure failure of the device in combination with a bottle with incorrect dimensions, overflow can cause contamination of reagents. Additionally, the current instructions for use does not highlight the importance to check the filling levels of all reagent bottles (RTU and system bottles) prior to every process run. Without this important information, risk of tissue damage or loss is possible. Therefore the missing information will be distributed to all customers by a Field Service Notice.

  • 2012-04-27TerminatedSoftware design

    ASP 6025 Vacuum Tissue Processor; an automated modular tissue processor; Leica Biosystems Nussloch GmbH, Heidelberger Str. 17-19, D-69226 Nussloch, Germany; The intended use of this device is as an automated system used to process tissue specimens for examination through fixation, dehydration, and infiltration.

    There is a potential malfunction with the device in autorotation mode, which can lead to incorrect reagent handling of the device. The end result may lead to damage or to loss of tissue specimens.

  • 2011-06-20TerminatedNonconforming Material/Component

    Peloris II Rapid Tissue Processor; Leica Biosystems Melbourne Pty Ltd, Mount Waverly, 3149 VIC, Australia; catalog number 26.0005. This tissue processor prepares tissue samples for sectioning by transforming fixed samples into wax embedded samples. This is achieved by exposing the tissue samples to a sequence of reagents in the processing retorts.

    The processor potentially has a faulty impellor which could lead to reagent carry-over from one processing step to another,

  • 2010-05-07TerminatedOther

    Peloris Rapid Tissue Processor; Leica Biosystems Melbourne Pty Ltd, Mount Waverly, 3149 VIC, Australia; catalog number 26.0001 and 26.0005

    Confusion during manually changing reagents, general users running unvalidated protocols, and wax valve leakage have contributed to tissue damage during processing, leading to rebiopsies.

  • 2010-04-12TerminatedOther

    ThermoFisher Scientific Microm STP 420D automated tissue processor.

    The firm has received complaints regarding specimen integrity (higher rate of specimen fragmentation) as well as complaints that the chamber may stop rotating prematurely, resulting in sample deterioration.

Monitor processor, tissue, automated automatically

DeviceWatch watches FDA adverse events and recalls for the products you care about, summarizes clinical narratives with AI and alerts you when a safety signal emerges.